Focal One C:\FocalOne\ErrorFO023.txt error file created on 26 Jul 2018 File updated on 24 Sep 2021 HOSPITAL N°PATIENT DATE J/M/A TIME H:M:S NUM ERROR MESSAGE --------- ---------- ---------- --------- ------- EDAP N° ---- 26/07/18 11:32:14 1.3.5101 Treatment cancelled at treatment data edition phase EDAP N° ---- 26/07/18 13:03:58 1.3.5101 Treatment cancelled at treatment data edition phase EDAP N° 0093 26/07/18 13:22:59 1.8.1302 Rectal wall position error Measured rectum wall position (29.9 mm) must be in the range 20.4 mm to 28.4 mm EDAP N° 0093 26/07/18 13:23:08 1.8.1302 Rectal wall position error Measured rectum wall position (29.5 mm) must be in the range 20.4 mm to 28.4 mm EDAP N° 0093 26/07/18 13:23:20 1.8.1302 Rectal wall position error Measured rectum wall position (16.6 mm) must be in the range 20.4 mm to 28.4 mm EDAP N° 0093 26/07/18 13:24:13 1.8.1302 Rectal wall position error Measured rectum wall position (21.5 mm) must be in the range 22.0 mm to 30.0 mm EDAP N° 0093 26/07/18 13:24:20 1.8.1302 Rectal wall position error Measured rectum wall position (21.7 mm) must be in the range 22.0 mm to 30.0 mm EDAP N° 0093 26/07/18 13:24:58 1.8.1303 Rectal wall position error Measured rectum wall position (28.3 mm) must be in the range 25.0 mm to 27.0 mm EDAP N° 0093 26/07/18 13:25:05 1.8.1303 Rectal wall position error Measured rectum wall position (27.5 mm) must be in the range 25.0 mm to 27.0 mm EDAP N° 0093 26/07/18 13:27:06 1.9.5104 Treatment cancelled at transversal localization phase EDAP N° 0093 26/07/18 13:45:44 1.6.5103 Treatment cancelled at treatment limits definition phase EDAP N° ---- 31/07/18 12:17:22 1.3.5101 Treatment cancelled at treatment data edition phase EDAP N° 0095 31/07/18 12:29:21 2.5.1202 The patient has moved EDAP N° 0095 31/07/18 12:29:23 2.5.1202 The patient has moved EDAP N° 0095 31/07/18 12:29:27 2.5.1202 The patient has moved EDAP N° 0095 31/07/18 12:34:56 1.8.1302 Rectal wall position error Measured rectum wall position (22.0 mm) must be in the range 22.0 mm to 30.0 mm EDAP N° 0095 31/07/18 12:35:22 1.8.1302 Rectal wall position error Measured rectum wall position (20.9 mm) must be in the range 21.7 mm to 29.7 mm EDAP N° 0095 31/07/18 12:35:59 1.8.1302 Rectal wall position error Measured rectum wall position (30.1 mm) must be in the range 22.0 mm to 30.0 mm EDAP N° 0095 31/07/18 12:37:51 1.9.5104 Treatment cancelled at transversal localization phase EDAP N° 0096 31/07/18 12:55:31 1.9.5104 Treatment cancelled at transversal localization phase EDAP N° 0096 31/07/18 13:13:43 1.9.5104 Treatment cancelled at transversal localization phase EDAP N° 0098 01/08/18 12:26:45 2.5.1202 El paciente se ha movido EDAP N° 0098 01/08/18 12:26:47 2.5.1202 El paciente se ha movido EDAP N° 0098 01/08/18 12:47:26 2.8.1413 La posición real está más allá de la posición solicitada en longitudinal axis EDAP N° 0098 01/08/18 12:47:26 2.8.1202 El paciente se ha movido EDAP N° 0098 01/08/18 12:48:12 1.9.5105 Tratamiento cancelado durante la fase de disparo EDAP N° 0098 01/08/18 12:53:48 1.6.5103 Tratamiento cancelado en la fase de definición de los límites del tratamiento CSM N° 0129 16/08/18 10:11:47 2.5.1202 El paciente se ha movido CSM N° 0129 16/08/18 11:00:45 2.8.1423 La posición real está más allá de la posición solicitada en antero-posterior axis EDAP N° 0099 16/08/18 13:03:04 2.5.1202 El paciente se ha movido EDAP N° 0099 16/08/18 13:33:50 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (24.5 mm) tiene que estar dentro del intervalo de 20.3 mm a 22.3 mm EDAP N° 0099 16/08/18 13:34:15 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.3 mm) tiene que estar dentro del intervalo de 20.3 mm a 22.3 mm EDAP N° 0099 16/08/18 13:34:53 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (21.0 mm) tiene que estar dentro del intervalo de 21.0 mm a 23.0 mm EDAP N° 0099 16/08/18 13:41:50 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.5 mm) tiene que estar dentro del intervalo de 22.6 mm a 24.6 mm EDAP N° ---- 16/08/18 13:46:04 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 16/08/18 13:46:17 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 16/08/18 13:46:26 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 16/08/18 13:46:28 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 16/08/18 13:46:29 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 16/08/18 13:46:35 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 16/08/18 13:46:38 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° 0100 16/08/18 14:28:09 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (24.0 mm) tiene que estar dentro del intervalo de 20.1 mm a 22.1 mm EDAP N° 0100 16/08/18 14:29:15 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.7 mm) tiene que estar dentro del intervalo de 20.1 mm a 22.1 mm EDAP N° 0100 16/08/18 14:29:33 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.3 mm) tiene que estar dentro del intervalo de 20.1 mm a 22.1 mm EDAP N° 0100 16/08/18 14:29:50 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.3 mm) tiene que estar dentro del intervalo de 20.1 mm a 22.1 mm EDAP N° 0100 16/08/18 14:30:18 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (20.8 mm) tiene que estar dentro del intervalo de 21.9 mm a 23.9 mm EDAP N° 0100 16/08/18 14:30:47 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.2 mm) tiene que estar dentro del intervalo de 23.3 mm a 25.3 mm EDAP N° 0100 16/08/18 14:31:12 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.9 mm) tiene que estar dentro del intervalo de 24.0 mm a 26.0 mm EDAP N° 0100 16/08/18 14:31:32 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.9 mm) tiene que estar dentro del intervalo de 24.4 mm a 26.4 mm EDAP N° 0100 16/08/18 14:35:54 1.6.5103 Tratamiento cancelado en la fase de definición de los límites del tratamiento EDAP N° 0100 16/08/18 14:36:15 1.5.5102 Tratamiento cancelado en la fase de adquisición EDAP N° ---- 16/08/18 14:45:30 1.3.5101 Tratamiento cancelado en la fase de edición de los datos del tratamiento EDAP N° 0101 16/08/18 15:27:39 1.6.5103 Treatment cancelled at treatment limits definition phase EDAP N° 0101 16/08/18 15:29:00 1.6.5103 Treatment cancelled at treatment limits definition phase EDAP N° 0101 16/08/18 15:36:43 2.5.1202 The patient has moved EDAP N° 0101 16/08/18 15:40:39 1.6.5103 Treatment cancelled at treatment limits definition phase EDAP N° 0101 16/08/18 15:44:33 1.6.5103 Tratamiento cancelado en la fase de definición de los límites del tratamiento CSM N° ---- 17/08/18 10:51:04 1.3.5101 Treatment cancelled at treatment data edition phase EDAP N° 0101 17/08/18 10:53:28 2.5.1202 The patient has moved EDAP N° 0101 17/08/18 10:54:44 2.6.1102 Probe temperature is too high : 20.0 EDAP N° 0101 17/08/18 10:54:51 2.6.1102 Probe temperature is too high : 20.0 EDAP N° 0101 17/08/18 10:55:29 2.6.1102 Probe temperature is too high : 20.1 EDAP N° 0101 17/08/18 10:55:30 1.6.5103 Treatment cancelled at treatment limits definition phase EDAP N° 0101 17/08/18 11:01:35 1.6.5103 Treatment cancelled at treatment limits definition phase EDAP N° 0101 17/08/18 11:35:32 1.6.5103 Treatment cancelled at treatment limits definition phase EDAP N° 0101 17/08/18 11:40:11 1.6.5103 Treatment cancelled at treatment limits definition phase CSM N° ---- 22/08/18 09:33:46 2.3.1901 Código Focalpak incorrecto CSM N° ---- 25/08/18 06:53:45 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/18 06:54:40 2.2.1604 El escáner no puede pasar a modo operativo EDAP N° 0102 27/08/18 15:24:34 2.5.1202 El paciente se ha movido EDAP N° 0102 27/08/18 15:24:35 2.5.1202 El paciente se ha movido EDAP N° 0102 27/08/18 15:54:03 1.6.5103 Tratamiento cancelado en la fase de definición de los límites del tratamiento CSM N° 0132 27/08/18 16:19:59 1.6.5103 Tratamiento cancelado en la fase de definición de los límites del tratamiento CSM N° 0132 27/08/18 16:36:22 1.6.5103 Tratamiento cancelado en la fase de definición de los límites del tratamiento EDAP N° ---- 27/08/18 17:18:42 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:52:39 2.3.1901 Código Focalpak incorrecto EDAP N° 0102 27/08/18 17:53:16 2.5.1604 El escáner no puede pasar a modo operativo EDAP N° 0102 27/08/18 17:53:38 2.5.1604 El escáner no puede pasar a modo operativo EDAP N° 0102 27/08/18 17:53:39 2.5.1604 El escáner no puede pasar a modo operativo EDAP N° 0102 27/08/18 17:53:39 2.5.1604 El escáner no puede pasar a modo operativo EDAP N° 0102 27/08/18 17:53:44 2.5.1604 El escáner no puede pasar a modo operativo EDAP N° 0102 27/08/18 17:53:47 1.5.5102 Tratamiento cancelado en la fase de adquisición EDAP N° ---- 27/08/18 17:57:28 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:57:36 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:57:38 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:57:38 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:57:38 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:57:38 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:57:39 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:57:39 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:57:39 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:57:39 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:57:40 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:57:40 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:57:40 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:57:40 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:57:40 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:57:41 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:57:41 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 27/08/18 17:57:42 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° 0102 27/08/18 18:14:20 1.6.5103 Tratamiento cancelado en la fase de definición de los límites del tratamiento CSM N° ---- 27/09/18 11:28:00 3.2.5002 Error de integridad en el MEP: WATCHDOG CSM N° ---- 27/09/18 11:28:47 2.2.1604 El escáner no puede pasar a modo operativo CSM N° ---- 27/09/18 11:29:58 2.2.1604 El escáner no puede pasar a modo operativo CSM N° ---- 27/09/18 11:32:19 3.2.5002 Error de integridad en el MEP: WATCHDOG CSM N° ---- 27/09/18 11:32:38 3.2.5002 Error de integridad en el MEP: WATCHDOG CSM N° ---- 27/09/18 11:33:11 3.2.5002 Error de integridad en el MEP: WATCHDOG CSM N° ---- 27/09/18 11:33:25 3.2.5002 Error de integridad en el MEP: WATCHDOG CSM N° 0133 27/09/18 12:21:49 2.5.1202 El paciente se ha movido CSM N° ---- 27/11/18 11:59:56 2.3.1901 Código Focalpak incorrecto CSM N° ---- 27/11/18 12:00:12 2.3.1901 Código Focalpak incorrecto CSM N° 0135 27/11/18 12:11:35 2.8.1202 El paciente se ha movido CSM N° 0135 27/11/18 12:11:35 2.8.1202 El paciente se ha movido CSM N° 0135 27/11/18 12:31:01 2.8.1401 No se puede comunicar con el sistema motor al enviar la instrucción: mover eje CSM N° 0135 27/11/18 12:41:45 2.8.1423 La posición real está más allá de la posición solicitada en antero-posterior axis CSM N° 0135 27/11/18 12:41:45 2.8.1202 El paciente se ha movido CSM N° 0135 27/11/18 12:49:20 1.9.5104 Tratamiento cancelado en la fase de localización transversal CSM N° 0135 27/11/18 12:49:28 2.7.1202 El paciente se ha movido CSM N° ---- 04/12/18 14:03:56 3.2.1503 La sonda no ha sido calibrada en esta máquina CSM N° ---- 04/12/18 14:04:06 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° ---- 04/12/18 15:04:46 2.2.1601 Cannot communicate with Ultrasound Scanner EDAP N° ---- 04/12/18 15:08:05 2.2.1601 Cannot communicate with Ultrasound Scanner EDAP N° ---- 04/12/18 15:08:07 2.2.1601 Cannot communicate with Ultrasound Scanner EDAP N° ---- 04/12/18 15:08:08 2.2.1601 Cannot communicate with Ultrasound Scanner EDAP N° ---- 04/12/18 15:08:14 2.2.1601 Cannot communicate with Ultrasound Scanner EDAP N° ---- 04/12/18 15:08:15 2.2.1601 Cannot communicate with Ultrasound Scanner EDAP N° 0103 04/12/18 15:21:56 2.5.1202 The patient has moved EDAP N° 0103 04/12/18 15:37:02 1.9.5105 Treatment cancelled during firing phase CSM N° 0136 12/03/19 09:09:08 2.8.1202 El paciente se ha movido CSM N° 0136 12/03/19 09:09:08 2.8.1202 El paciente se ha movido CSM N° 0136 12/03/19 09:24:58 2.7.1202 El paciente se ha movido CSM N° 0136 12/03/19 09:52:50 2.7.1202 El paciente se ha movido CSM N° 0136 12/03/19 09:56:06 1.9.5104 Tratamiento cancelado en la fase de localización transversal CSM N° 0136 12/03/19 09:56:11 2.7.1202 El paciente se ha movido EDAP N° 0105 22/03/19 11:56:54 2.5.1102 Probe temperature is too high : 20.4 EDAP N° 0105 22/03/19 12:10:41 2.7.1202 The patient has moved EDAP N° 0106 22/03/19 14:05:03 2.8.1202 The patient has moved EDAP N° 0106 22/03/19 14:05:03 2.8.1202 The patient has moved CSM N° ---- 26/03/19 07:25:54 2.2.1423 La posición real está más allá de la posición solicitada en antero-posterior axis CSM N° ---- 26/03/19 07:27:44 2.2.1423 La posición real está más allá de la posición solicitada en antero-posterior axis CSM N° ---- 26/03/19 07:27:47 2.2.1423 La posición real está más allá de la posición solicitada en antero-posterior axis CSM N° 0137 26/03/19 09:20:18 2.8.1202 El paciente se ha movido CSM N° 0137 26/03/19 09:20:19 2.8.1202 El paciente se ha movido CSM N° 0137 26/03/19 09:21:27 2.7.1202 El paciente se ha movido CSM N° 0137 26/03/19 09:28:17 2.8.1423 La posición real está más allá de la posición solicitada en antero-posterior axis CSM N° 0137 26/03/19 09:43:46 2.8.1010 El MEP no ha recibido el final de la trayectoria. Se ha parado el disparo. CSM N° 0137 26/03/19 09:44:33 1.9.5104 Tratamiento cancelado en la fase de localización transversal CSM N° 0138 26/03/19 11:05:49 2.5.1102 Temperatura de la sonda excesiva : 24.4 CSM N° 0138 26/03/19 11:18:50 2.7.1202 El paciente se ha movido CSM N° 0138 26/03/19 11:39:45 2.8.1202 El paciente se ha movido CSM N° 0138 26/03/19 11:40:27 2.8.1202 El paciente se ha movido CSM N° 0138 26/03/19 11:46:55 2.8.1202 El paciente se ha movido CSM N° 0138 26/03/19 11:46:55 2.8.1202 El paciente se ha movido CSM N° 0138 26/03/19 12:30:11 1.9.5104 Tratamiento cancelado en la fase de localización transversal EDAP N° 0107 11/04/19 12:13:42 2.5.1202 The patient has moved EDAP N° 0107 11/04/19 12:18:35 1.9.5104 Treatment cancelled at transversal localization phase EDAP N° 0107 11/04/19 12:21:58 1.8.1302 Rectal wall position error Measured rectum wall position (16.8 mm) must be in the range 19.4 mm to 27.4 mm EDAP N° 0107 11/04/19 12:22:29 1.8.1302 Rectal wall position error Measured rectum wall position (18.8 mm) must be in the range 21.4 mm to 29.4 mm EDAP N° 0107 11/04/19 12:23:19 1.9.5105 Treatment cancelled during firing phase EDAP N° 0107 11/04/19 12:27:48 1.8.1302 Rectal wall position error Measured rectum wall position (19.2 mm) must be in the range 20.4 mm to 28.4 mm EDAP N° 0108 11/04/19 13:47:52 1.9.5104 Treatment cancelled at transversal localization phase EDAP N° ---- 11/04/19 16:10:54 1.3.5101 Treatment cancelled at treatment data edition phase EDAP N° 0109 11/04/19 16:14:32 2.5.1107 Status error on Cooling System: Fan 1 failure EDAP N° 0109 11/04/19 16:14:40 2.5.1107 Status error on Cooling System: Fan 1 failure EDAP N° 0109 11/04/19 16:14:45 2.5.1107 Status error on Cooling System: Fan 1 failure EDAP N° 0109 11/04/19 16:14:47 2.5.1107 Status error on Cooling System: Fan 1 failure EDAP N° 0109 11/04/19 16:14:49 2.5.1107 Status error on Cooling System: Fan 1 failure EDAP N° 0109 11/04/19 16:14:51 2.5.1107 Status error on Cooling System: Fan 1 failure EDAP N° 0109 11/04/19 16:18:36 2.5.1202 The patient has moved EDAP N° 0109 11/04/19 16:18:38 2.5.1202 The patient has moved EDAP N° 0109 11/04/19 16:24:29 1.8.1302 Rectal wall position error Measured rectum wall position (30.5 mm) must be in the range 19.9 mm to 27.9 mm EDAP N° 0109 11/04/19 16:24:50 1.8.1302 Rectal wall position error Measured rectum wall position (29.3 mm) must be in the range 19.9 mm to 27.9 mm EDAP N° 0109 11/04/19 16:24:58 1.8.1302 Rectal wall position error Measured rectum wall position (28.5 mm) must be in the range 19.9 mm to 27.9 mm EDAP N° 0109 11/04/19 16:26:16 1.8.1302 Rectal wall position error Measured rectum wall position (36.8 mm) must be in the range 19.2 mm to 27.2 mm EDAP N° 0111 11/04/19 17:12:22 1.8.1303 Rectal wall position error Measured rectum wall position (23.9 mm) must be in the range 24.4 mm to 26.4 mm EDAP N° 0111 11/04/19 17:12:42 1.8.1303 Rectal wall position error Measured rectum wall position (23.9 mm) must be in the range 24.7 mm to 26.7 mm EDAP N° 0111 11/04/19 17:13:10 1.8.1303 Rectal wall position error Measured rectum wall position (23.0 mm) must be in the range 24.7 mm to 26.7 mm EDAP N° 0111 11/04/19 17:13:38 1.8.1303 Rectal wall position error Measured rectum wall position (24.8 mm) must be in the range 25.0 mm to 27.0 mm EDAP N° 0111 11/04/19 17:19:41 1.8.1303 Rectal wall position error Measured rectum wall position (24.2 mm) must be in the range 25.0 mm to 27.0 mm EDAP N° 0111 11/04/19 17:21:01 1.8.1303 Rectal wall position error Measured rectum wall position (25.6 mm) must be in the range 23.0 mm to 25.0 mm EDAP N° 0111 11/04/19 17:21:29 1.8.1303 Rectal wall position error Measured rectum wall position (24.4 mm) must be in the range 21.3 mm to 23.3 mm EDAP N° 0111 11/04/19 17:21:59 1.8.1303 Rectal wall position error Measured rectum wall position (22.9 mm) must be in the range 23.9 mm to 25.9 mm EDAP N° 0111 11/04/19 17:22:18 1.8.1303 Rectal wall position error Measured rectum wall position (24.2 mm) must be in the range 24.7 mm to 26.7 mm EDAP N° 0111 11/04/19 17:22:26 1.8.1303 Rectal wall position error Measured rectum wall position (24.4 mm) must be in the range 24.7 mm to 26.7 mm EDAP N° 0111 11/04/19 17:22:40 1.8.1303 Rectal wall position error Measured rectum wall position (25.0 mm) must be in the range 25.0 mm to 27.0 mm EDAP N° 0111 11/04/19 17:23:53 1.8.1303 Rectal wall position error Measured rectum wall position (25.3 mm) must be in the range 23.2 mm to 25.2 mm EDAP N° 0111 11/04/19 17:31:51 1.8.1303 Rectal wall position error Measured rectum wall position (24.4 mm) must be in the range 24.4 mm to 26.4 mm EDAP N° 0111 11/04/19 17:32:15 1.8.1303 Rectal wall position error Measured rectum wall position (24.5 mm) must be in the range 24.8 mm to 26.8 mm EDAP N° 0111 11/04/19 17:33:43 1.8.1303 Rectal wall position error Measured rectum wall position (25.8 mm) must be in the range 22.7 mm to 24.7 mm EDAP N° 0111 11/04/19 17:34:05 1.8.1303 Rectal wall position error Measured rectum wall position (24.2 mm) must be in the range 20.9 mm to 22.9 mm EDAP N° 0111 11/04/19 17:34:52 1.8.1303 Rectal wall position error Measured rectum wall position (22.7 mm) must be in the range 22.9 mm to 24.9 mm EDAP N° 0111 11/04/19 17:36:25 1.8.1303 Rectal wall position error Measured rectum wall position (23.6 mm) must be in the range 23.7 mm to 25.7 mm EDAP N° 0111 11/04/19 17:37:43 1.8.1301 Rectal wall not detected EDAP N° 0111 11/04/19 17:40:15 1.8.1303 Rectal wall position error Measured rectum wall position (24.4 mm) must be in the range 24.9 mm to 26.9 mm EDAP N° 0111 11/04/19 17:56:04 1.8.1303 Rectal wall position error Measured rectum wall position (25.2 mm) must be in the range 22.6 mm to 24.6 mm EDAP N° 0111 11/04/19 17:57:47 1.8.1303 Rectal wall position error Measured rectum wall position (24.4 mm) must be in the range 24.5 mm to 26.5 mm EDAP N° 0111 11/04/19 18:07:54 1.8.1303 Rectal wall position error Measured rectum wall position (25.0 mm) must be in the range 22.9 mm to 24.9 mm EDAP N° 0111 11/04/19 18:08:31 1.8.1301 Rectal wall not detected EDAP N° 0111 11/04/19 18:09:53 1.8.1301 Rectal wall not detected EDAP N° 0111 11/04/19 18:10:27 1.8.1301 Rectal wall not detected EDAP N° 0111 11/04/19 18:13:31 1.8.1303 Rectal wall position error Measured rectum wall position (23.5 mm) must be in the range 23.8 mm to 25.8 mm EDAP N° 0111 11/04/19 18:19:01 1.9.5104 Treatment cancelled at transversal localization phase CSM N° 0139 16/04/19 10:40:36 1.9.5104 Tratamiento cancelado en la fase de localización transversal CSM N° 0140 24/04/19 09:42:54 2.7.1202 El paciente se ha movido CSM N° 0140 24/04/19 10:14:15 1.9.5104 Tratamiento cancelado en la fase de localización transversal CSM N° ---- 16/05/19 10:02:32 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 16/05/19 10:03:25 2.2.1604 El escáner no puede pasar a modo operativo CSM N° ---- 16/05/19 10:05:44 2.2.1502 No se puede comunicar con la sonda CSM N° 0141 16/05/19 11:28:45 2.6.1202 El paciente se ha movido CSM N° 0141 16/05/19 12:48:42 1.9.5105 Tratamiento cancelado durante la fase de disparo CSM N° ---- 13/06/19 11:36:17 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 13/06/19 11:36:25 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 13/06/19 11:36:29 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 13/06/19 11:36:33 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° 0142 13/06/19 12:12:26 1.9.5105 Tratamiento cancelado durante la fase de disparo CSM N° 0142 13/06/19 12:16:16 2.5.1202 El paciente se ha movido CSM N° 0142 13/06/19 12:16:40 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0142 13/06/19 12:16:43 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0142 13/06/19 12:16:45 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0142 13/06/19 12:16:47 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° 0142 13/06/19 12:16:50 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0142 13/06/19 12:16:56 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° 0142 13/06/19 12:16:59 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0142 13/06/19 12:17:05 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0142 13/06/19 12:17:08 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0142 13/06/19 12:17:21 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0142 13/06/19 12:17:25 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° 0142 13/06/19 12:17:28 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0142 13/06/19 12:17:34 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° ---- 06/08/19 08:30:58 3.2.1503 La sonda no ha sido calibrada en esta máquina CSM N° ---- 06/08/19 08:35:13 3.2.1503 La sonda no ha sido calibrada en esta máquina CSM N° ---- 06/08/19 08:39:46 3.2.1503 La sonda no ha sido calibrada en esta máquina CSM N° ---- 06/08/19 08:40:00 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° 0144 06/08/19 09:16:30 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:16:34 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:16:36 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:16:37 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:16:39 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:16:40 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:16:51 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:16:53 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:16:57 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:17:00 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:17:01 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:17:03 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:17:11 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:17:13 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:17:26 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:17:28 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:17:30 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:17:32 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° 0144 06/08/19 09:17:36 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:17:39 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:17:41 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:17:50 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:17:53 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° 0144 06/08/19 09:17:58 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:02 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:24 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:26 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:28 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:30 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:32 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:34 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:36 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:37 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:38 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:39 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:41 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:42 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:44 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:45 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:47 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:48 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:49 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:50 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:51 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:52 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:53 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:54 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:56 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:57 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:58 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:18:59 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:19:00 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:19:01 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:19:04 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:19:05 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:19:06 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:19:10 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:19:12 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:19:21 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:19:26 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:19:29 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:19:29 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:19:31 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:19:32 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:19:33 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:19:49 2.5.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0144 06/08/19 09:21:09 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° ---- 06/08/19 09:27:11 2.2.1451 Error al inicializar la comunicación en el motor: 1 2 3 CSM N° 0144 06/08/19 09:46:46 2.8.1006 Error de potencia en demasiados canales: en canal(es) 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, CSM N° ---- 09/08/19 10:37:02 3.2.1503 La sonda no ha sido calibrada en esta máquina CSM N° ---- 09/08/19 11:25:53 2.0.5004 Impossible to open COM Port for MEP CSM N° ---- 09/08/19 11:26:06 2.0.5004 Impossible to open COM Port for MEP CSM N° ---- 09/08/19 11:27:48 2.0.5004 Impossible to open COM Port for MEP EDAP N° 0112 09/08/19 11:37:06 2.5.1202 The patient has moved EDAP N° ---- 09/08/19 12:35:10 2.0.5004 Impossible to open COM Port for MEP EDAP N° ---- 09/08/19 12:35:21 2.0.5004 Impossible to open COM Port for MEP EDAP N° 0113 09/08/19 12:46:47 1.6.5103 Tratamiento cancelado en la fase de definición de los límites del tratamiento CSM N° ---- 09/08/19 13:36:41 1.3.5101 Treatment cancelled at treatment data edition phase CSM N° 0145 13/08/19 09:05:47 2.6.1202 El paciente se ha movido CSM N° 0145 13/08/19 09:09:57 2.7.1202 El paciente se ha movido CSM N° 0145 13/08/19 09:10:35 2.7.1202 El paciente se ha movido CSM N° 0145 13/08/19 09:13:33 2.8.1202 El paciente se ha movido CSM N° 0145 13/08/19 09:19:32 2.8.1202 El paciente se ha movido CSM N° 0145 13/08/19 09:19:32 2.8.1202 El paciente se ha movido CSM N° 0145 13/08/19 09:20:08 2.8.1202 El paciente se ha movido CSM N° 0145 13/08/19 09:39:26 2.8.1003 Potencia medida insuficiente: menos del 90 % de lo solicitado (89.7 %). CSM N° 0145 13/08/19 09:39:39 2.8.1003 Potencia medida insuficiente: menos del 90 % de lo solicitado (89.9 %). CSM N° 0145 13/08/19 09:39:45 2.8.1003 Potencia medida insuficiente: menos del 90 % de lo solicitado (89.3 %). CSM N° 0145 13/08/19 09:44:26 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (27.7 mm) tiene que estar dentro del intervalo de 19.0 mm a 27.0 mm CSM N° 0145 13/08/19 09:47:53 2.8.1003 Potencia medida insuficiente: menos del 90 % de lo solicitado (88.7 %). CSM N° 0145 13/08/19 09:48:22 2.8.1003 Potencia medida insuficiente: menos del 90 % de lo solicitado (88.0 %). CSM N° 0145 13/08/19 09:51:30 2.8.1003 Potencia medida insuficiente: menos del 90 % de lo solicitado (87.4 %). CSM N° 0145 13/08/19 09:51:49 2.8.1003 Potencia medida insuficiente: menos del 90 % de lo solicitado (89.4 %). CSM N° 0145 13/08/19 09:51:57 2.8.1002 Potencia medida insuficiente: más de 16W por debajo de lo solicitado (-16.2 W). CSM N° 0145 13/08/19 09:52:10 2.8.1002 Potencia medida insuficiente: más de 16W por debajo de lo solicitado (-17.1 W). CSM N° 0145 13/08/19 09:54:58 2.8.1003 Potencia medida insuficiente: menos del 90 % de lo solicitado (88.8 %). CSM N° 0145 13/08/19 09:55:05 2.8.1003 Potencia medida insuficiente: menos del 90 % de lo solicitado (88.0 %). EDAP N° 0114 21/08/19 09:36:35 2.5.1202 The patient has moved EDAP N° 0114 21/08/19 09:36:36 2.5.1202 The patient has moved EDAP N° 0114 21/08/19 09:40:05 2.8.1006 Power error on too many channels: on channel(s) 5, 6, 8, 9, 12, 13, EDAP N° 0114 21/08/19 09:40:18 2.8.1003 Measured power too low: less than 90% of requested (88.7 %). EDAP N° 0114 21/08/19 09:40:31 2.8.1006 Power error on too many channels: on channel(s) 5, 6, 8, 9, EDAP N° 0114 21/08/19 09:41:11 2.8.1006 Power error on too many channels: on channel(s) 5, 6, 8, 9, EDAP N° 0114 21/08/19 09:41:29 2.8.1006 Power error on too many channels: on channel(s) 5, 6, 8, 9, EDAP N° 0114 21/08/19 09:41:42 2.8.1003 Measured power too low: less than 90% of requested (89.7 %). EDAP N° 0114 21/08/19 09:41:52 2.8.1006 Power error on too many channels: on channel(s) 5, 6, 7, 8, 9, 12, 13, 15, EDAP N° 0115 21/08/19 10:22:33 2.8.1004 Measured power too high: more than 110% of requested (112.0%). EDAP N° 0115 21/08/19 10:24:12 2.8.1004 Measured power too high: more than 110% of requested (113.0%). EDAP N° 0115 21/08/19 10:24:44 2.8.1004 Measured power too high: more than 110% of requested (111.2%). EDAP N° 0115 21/08/19 10:26:19 2.8.1004 Measured power too high: more than 110% of requested (114.8%). EDAP N° 0115 21/08/19 10:29:11 2.8.1004 Measured power too high: more than 110% of requested (112.7%). EDAP N° 0115 21/08/19 10:30:26 2.8.1001 Measured power too high: more than 16W over requested (19.7 W). EDAP N° 0115 21/08/19 10:30:38 2.8.1001 Measured power too high: more than 16W over requested (17.5 W). EDAP N° 0115 21/08/19 10:30:44 2.8.1006 Power error on too many channels: on channel(s) 7, 11, 12, 13, EDAP N° 0115 21/08/19 10:30:49 2.8.1006 Power error on too many channels: on channel(s) 9, 11, 12, 13, EDAP N° 0115 21/08/19 10:31:30 2.8.1004 Measured power too high: more than 110% of requested (110.3%). EDAP N° 0115 21/08/19 10:31:36 2.8.1004 Measured power too high: more than 110% of requested (110.8%). EDAP N° 0115 21/08/19 10:31:42 2.8.1004 Measured power too high: more than 110% of requested (114.3%). EDAP N° 0115 21/08/19 10:33:03 2.8.1001 Measured power too high: more than 16W over requested (21.6 W). EDAP N° 0115 21/08/19 10:33:47 2.8.1001 Measured power too high: more than 16W over requested (17.9 W). EDAP N° 0115 21/08/19 10:34:01 2.8.1001 Measured power too high: more than 16W over requested (18.4 W). EDAP N° 0115 21/08/19 10:34:41 2.8.1004 Measured power too high: more than 110% of requested (111.8%). EDAP N° 0115 21/08/19 10:35:14 2.8.1004 Measured power too high: more than 110% of requested (112.2%). EDAP N° 0115 21/08/19 10:35:27 1.9.5104 Treatment cancelled at transversal localization phase EDAP N° 0116 21/08/19 11:17:59 1.5.5102 Treatment cancelled at acquisition phase EDAP N° 0118 28/08/19 12:51:41 2.5.1202 The patient has moved EDAP N° 0118 28/08/19 12:54:54 2.8.1004 Measured power too high: more than 110% of requested (110.5%). EDAP N° 0118 28/08/19 12:55:49 2.8.1004 Measured power too high: more than 110% of requested (112.7%). EDAP N° 0118 28/08/19 12:55:53 2.8.1004 Measured power too high: more than 110% of requested (113.1%). EDAP N° 0118 28/08/19 12:55:56 2.8.1004 Measured power too high: more than 110% of requested (112.6%). EDAP N° 0118 28/08/19 12:56:00 2.8.1004 Measured power too high: more than 110% of requested (112.5%). EDAP N° 0118 28/08/19 12:56:04 2.8.1004 Measured power too high: more than 110% of requested (113.4%). EDAP N° 0118 28/08/19 12:56:11 2.8.1004 Measured power too high: more than 110% of requested (114.8%). EDAP N° 0118 28/08/19 12:56:15 2.8.1004 Measured power too high: more than 110% of requested (114.0%). EDAP N° 0118 28/08/19 12:56:19 2.8.1004 Measured power too high: more than 110% of requested (113.2%). EDAP N° 0118 28/08/19 12:56:26 2.8.1004 Measured power too high: more than 110% of requested (115.4%). EDAP N° 0118 28/08/19 12:56:31 2.8.1004 Measured power too high: more than 110% of requested (115.2%). EDAP N° 0118 28/08/19 12:56:35 2.8.1004 Measured power too high: more than 110% of requested (114.4%). EDAP N° 0118 28/08/19 12:56:39 2.8.1006 Power error on too many channels: on channel(s) 1, 2, 3, 13, EDAP N° 0118 28/08/19 12:56:43 2.8.1006 Power error on too many channels: on channel(s) 1, 2, 3, 13, EDAP N° 0118 28/08/19 12:56:47 2.8.1006 Power error on too many channels: on channel(s) 1, 2, 3, 13, EDAP N° 0118 28/08/19 12:57:15 2.8.1006 Power error on too many channels: on channel(s) 1, 2, 3, 13, EDAP N° 0118 28/08/19 12:57:19 2.8.1004 Measured power too high: more than 110% of requested (115.5%). EDAP N° 0118 28/08/19 12:57:27 2.8.1004 Measured power too high: more than 110% of requested (116.8%). EDAP N° 0118 28/08/19 12:57:33 2.8.1006 Power error on too many channels: on channel(s) 1, 2, 3, 4, 11, 13, EDAP N° 0118 28/08/19 12:57:37 2.8.1004 Measured power too high: more than 110% of requested (115.3%). EDAP N° 0118 28/08/19 12:57:41 2.8.1004 Measured power too high: more than 110% of requested (114.9%). EDAP N° 0118 28/08/19 12:57:45 2.8.1004 Measured power too high: more than 110% of requested (114.9%). EDAP N° 0118 28/08/19 12:58:52 1.5.5102 Treatment cancelled at acquisition phase EDAP N° 0119 28/08/19 13:16:49 2.8.1004 Measured power too high: more than 110% of requested (110.3%). EDAP N° 0119 28/08/19 13:16:56 2.8.1004 Measured power too high: more than 110% of requested (112.2%). EDAP N° 0119 28/08/19 13:17:01 2.8.1004 Measured power too high: more than 110% of requested (110.8%). EDAP N° 0119 28/08/19 13:17:12 2.8.1004 Measured power too high: more than 110% of requested (113.5%). EDAP N° 0119 28/08/19 13:17:48 2.8.1006 Power error on too many channels: on channel(s) 7, 13, 14, 15, CSM N° ---- 03/09/19 07:55:57 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 03/09/19 07:57:38 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 03/09/19 07:58:32 2.2.1604 El escáner no puede pasar a modo operativo CSM N° ---- 03/09/19 08:00:56 2.2.1604 El escáner no puede pasar a modo operativo CSM N° ---- 03/09/19 08:01:09 2.2.1604 El escáner no puede pasar a modo operativo CSM N° 0147 22/10/19 13:46:31 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (27.0 mm) tiene que estar dentro del intervalo de 19.0 mm a 27.0 mm CSM N° 0147 22/10/19 13:49:53 1.9.5105 Tratamiento cancelado durante la fase de disparo CSM N° 0148 30/10/19 11:34:40 1.6.5103 Tratamiento cancelado en la fase de definición de los límites del tratamiento CSM N° ---- 06/11/19 10:38:07 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 06/11/19 10:41:33 2.2.1502 No se puede comunicar con la sonda CSM N° 0149 06/11/19 12:00:49 2.5.1102 Temperatura de la sonda excesiva : 28.6 CSM N° 0149 06/11/19 13:02:19 1.9.5105 Tratamiento cancelado durante la fase de disparo CSM N° 0149 06/11/19 13:02:26 2.7.1202 El paciente se ha movido CSM N° 0150 13/11/19 12:42:27 1.9.5104 Tratamiento cancelado en la fase de localización transversal CSM N° ---- 26/11/19 09:04:14 3.2.1503 La sonda no ha sido calibrada en esta máquina CSM N° ---- 26/11/19 09:05:22 3.2.1503 La sonda no ha sido calibrada en esta máquina CSM N° ---- 26/11/19 09:09:22 3.2.1503 La sonda no ha sido calibrada en esta máquina CSM N° ---- 26/11/19 09:09:35 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 26/11/19 09:10:03 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 26/11/19 09:10:10 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 26/11/19 09:15:10 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 26/11/19 09:15:18 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 26/11/19 09:15:22 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° 0151 26/11/19 09:46:38 2.5.1202 El paciente se ha movido CSM N° 0153 08/01/20 09:35:04 2.5.1202 El paciente se ha movido CSM N° 0153 08/01/20 09:42:09 2.8.1202 El paciente se ha movido CSM N° ---- 22/01/20 09:42:09 3.2.1503 La sonda no ha sido calibrada en esta máquina CSM N° 0154 22/01/20 11:01:39 2.5.1202 El paciente se ha movido CSM N° 0154 22/01/20 11:01:52 2.5.1102 Temperatura de la sonda excesiva : 19.8 CSM N° 0154 22/01/20 11:01:55 2.5.1102 Temperatura de la sonda excesiva : 19.6 CSM N° 0154 22/01/20 11:36:57 2.8.1202 El paciente se ha movido CSM N° 0154 22/01/20 11:36:57 2.8.1202 El paciente se ha movido EDAP N° ---- 19/02/20 12:26:35 1.3.5101 Tratamiento cancelado en la fase de edición de los datos del tratamiento EDAP N° ---- 19/02/20 16:45:58 3.2.1503 La sonda no ha sido calibrada en esta máquina EDAP N° ---- 19/02/20 16:48:56 1.3.5101 Tratamiento cancelado en la fase de edición de los datos del tratamiento EDAP N° 0120 20/02/20 10:16:27 1.8.1302 Rectal wall position error Measured rectum wall position (30.0 mm) must be in the range 20.3 mm to 28.3 mm EDAP N° 0120 20/02/20 10:17:05 1.8.1302 Rectal wall position error Measured rectum wall position (27.4 mm) must be in the range 18.5 mm to 26.5 mm EDAP N° 0120 20/02/20 10:19:41 1.8.1302 Rectal wall position error Measured rectum wall position (29.7 mm) must be in the range 21.5 mm to 29.5 mm EDAP N° 0120 20/02/20 10:19:57 2.8.1421 Probe movement is impossible. The motor has reached the mechanical stop on Transversal Axis EDAP N° 0120 20/02/20 10:20:22 2.8.1421 Probe movement is impossible. The motor has reached the mechanical stop on Transversal Axis EDAP N° 0120 20/02/20 10:20:36 2.8.1421 Probe movement is impossible. The motor has reached the mechanical stop on Transversal Axis EDAP N° 0120 20/02/20 10:20:39 2.8.1421 Probe movement is impossible. The motor has reached the mechanical stop on Transversal Axis EDAP N° 0120 20/02/20 10:20:42 2.8.1421 Probe movement is impossible. The motor has reached the mechanical stop on Transversal Axis EDAP N° 0120 20/02/20 10:21:04 1.8.1302 Rectal wall position error Measured rectum wall position (17.1 mm) must be in the range 17.4 mm to 25.4 mm EDAP N° 0120 20/02/20 10:21:33 1.8.1302 Rectal wall position error Measured rectum wall position (19.6 mm) must be in the range 21.2 mm to 29.2 mm EDAP N° 0121 20/02/20 12:26:55 1.6.5103 Treatment cancelled at treatment limits definition phase EDAP N° ---- 20/02/20 13:23:52 1.3.5101 Treatment cancelled at treatment data edition phase EDAP N° 0122 20/02/20 18:18:36 1.8.1302 Rectal wall position error Measured rectum wall position (17.9 mm) must be in the range 21.1 mm to 29.1 mm EDAP N° 0122 20/02/20 18:20:55 1.8.1302 Rectal wall position error Measured rectum wall position (20.1 mm) must be in the range 20.6 mm to 28.6 mm EDAP N° 0122 20/02/20 18:26:26 1.9.5104 Treatment cancelled at transversal localization phase EDAP N° 0123 20/02/20 20:00:13 1.5.5102 Treatment cancelled at acquisition phase EDAP N° 0124 02/03/20 10:19:28 2.5.1202 The patient has moved EDAP N° 0125 02/03/20 11:35:08 1.9.5104 Treatment cancelled at transversal localization phase EDAP N° 0126 02/03/20 11:52:36 1.9.5104 Treatment cancelled at transversal localization phase EDAP N° ---- 02/03/20 14:04:02 2.3.1901 Incorrect Focalpak code EDAP N° 0127 02/03/20 14:10:58 2.8.1202 The patient has moved EDAP N° 0127 02/03/20 14:10:58 2.8.1202 The patient has moved EDAP N° 0127 02/03/20 14:11:11 2.8.1202 The patient has moved EDAP N° 0127 02/03/20 14:11:15 2.8.1202 The patient has moved EDAP N° 0127 02/03/20 14:11:17 2.8.1202 The patient has moved EDAP N° 0127 02/03/20 14:11:25 2.8.1202 The patient has moved EDAP N° 0127 02/03/20 14:11:27 2.8.1202 The patient has moved EDAP N° 0127 02/03/20 14:11:32 2.8.1202 The patient has moved EDAP N° 0127 02/03/20 14:11:36 2.8.1202 The patient has moved EDAP N° 0127 02/03/20 14:11:41 2.8.1202 The patient has moved EDAP N° 0127 02/03/20 14:11:42 2.8.1202 The patient has moved EDAP N° 0127 02/03/20 14:11:47 2.7.1202 The patient has moved EDAP N° 0128 02/03/20 14:16:47 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:16:48 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:16:49 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:16:50 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:16:50 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:16:51 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:16:51 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:16:52 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:16:52 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:17:00 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:17:50 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:17:51 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:17:51 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:17:52 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:17:56 1.5.5102 Treatment cancelled at acquisition phase EDAP N° 0128 02/03/20 14:18:12 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:18:25 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:18:27 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:18:37 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:18:39 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:18:40 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:18:40 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:18:41 2.5.1202 The patient has moved EDAP N° 0128 02/03/20 14:18:44 1.5.5102 Treatment cancelled at acquisition phase EDAP N° 0129 02/03/20 14:22:01 2.5.1202 The patient has moved EDAP N° 0129 02/03/20 14:22:02 2.5.1202 The patient has moved EDAP N° 0129 02/03/20 14:22:03 2.5.1202 The patient has moved EDAP N° 0129 02/03/20 14:22:06 1.5.5102 Treatment cancelled at acquisition phase EDAP N° 0130 02/03/20 14:24:49 2.5.1202 The patient has moved EDAP N° 0130 02/03/20 14:24:53 1.5.5102 Treatment cancelled at acquisition phase EDAP N° 0131 02/03/20 14:31:30 2.5.1202 The patient has moved EDAP N° 0131 02/03/20 14:38:54 1.9.5104 Treatment cancelled at transversal localization phase EDAP N° 0132 02/03/20 14:49:27 1.9.5105 Treatment cancelled during firing phase EDAP N° 0133 02/03/20 14:55:10 1.8.1302 Rectal wall position error Measured rectum wall position (28.0 mm) must be in the range 18.7 mm to 26.7 mm EDAP N° 0133 02/03/20 14:55:32 1.8.1301 Rectal wall not detected EDAP N° 0133 02/03/20 14:55:47 1.8.1302 Rectal wall position error Measured rectum wall position (29.5 mm) must be in the range 20.9 mm to 28.9 mm EDAP N° 0133 02/03/20 14:56:05 1.8.1302 Rectal wall position error Measured rectum wall position (20.7 mm) must be in the range 22.0 mm to 30.0 mm EDAP N° 0133 02/03/20 14:56:18 1.8.1302 Rectal wall position error Measured rectum wall position (20.6 mm) must be in the range 22.0 mm to 30.0 mm EDAP N° 0133 02/03/20 14:56:28 1.8.1302 Rectal wall position error Measured rectum wall position (20.7 mm) must be in the range 22.0 mm to 30.0 mm EDAP N° 0133 02/03/20 14:56:58 1.8.1302 Rectal wall position error Measured rectum wall position (20.8 mm) must be in the range 22.0 mm to 30.0 mm EDAP N° 0133 02/03/20 14:57:51 1.8.1302 Rectal wall position error Measured rectum wall position (19.3 mm) must be in the range 20.9 mm to 28.9 mm EDAP N° 0133 02/03/20 14:59:34 1.9.5104 Treatment cancelled at transversal localization phase CSM N° 0155 28/07/20 09:55:48 2.5.1202 El paciente se ha movido CSM N° 0155 28/07/20 10:22:31 2.8.1423 La posición real está más allá de la posición solicitada en antero-posterior axis EDAP N° ---- 20/08/20 13:52:56 1.3.5101 Treatment cancelled at treatment data edition phase EDAP N° 0134 20/08/20 14:04:49 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:04:51 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:05:05 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:05:30 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:05:31 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:05:32 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:06:06 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:06:07 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:06:07 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:06:08 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:06:08 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:06:10 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:06:11 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:06:11 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:06:12 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:06:12 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:06:13 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:07:04 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:07:04 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:07:17 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:07:18 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:07:20 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:07:20 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:07:20 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:07:21 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:08:16 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:08:16 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:08:19 1.5.5102 Treatment cancelled at acquisition phase EDAP N° 0134 20/08/20 14:08:37 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:08:38 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:08:39 2.5.1202 The patient has moved EDAP N° 0134 20/08/20 14:08:41 1.5.5102 Treatment cancelled at acquisition phase CSM N° 0156 27/08/20 12:30:11 2.7.1202 El paciente se ha movido CSM N° 0156 27/08/20 13:02:54 1.5.5102 Tratamiento cancelado en la fase de adquisición EDAP N° ---- 15/09/20 10:09:01 2.2.1502 No se puede comunicar con la sonda EDAP N° ---- 15/09/20 10:10:03 2.2.1604 El escáner no puede pasar a modo operativo EDAP N° ---- 15/09/20 10:13:44 2.2.1502 No se puede comunicar con la sonda EDAP N° ---- 15/09/20 10:23:58 1.3.5101 Tratamiento cancelado en la fase de edición de los datos del tratamiento EDAP N° ---- 15/09/20 12:25:29 1.3.5101 Tratamiento cancelado en la fase de edición de los datos del tratamiento EDAP N° ---- 15/09/20 12:32:54 2.0.1009 Imposible abrir puerto COM del amplificador EDAP N° ---- 15/09/20 12:53:12 2.0.1009 Imposible abrir puerto COM del amplificador EDAP N° ---- 15/09/20 13:21:08 1.3.5101 Tratamiento cancelado en la fase de edición de los datos del tratamiento EDAP N° ---- 13/10/20 09:53:11 2.0.5004 Imposible abrir puerto COM del MEP EDAP N° ---- 13/10/20 09:53:35 2.0.5004 Imposible abrir puerto COM del MEP EDAP N° ---- 13/10/20 10:18:54 2.0.5004 Imposible abrir puerto COM del MEP EDAP N° ---- 13/10/20 10:19:32 2.0.5004 Imposible abrir puerto COM del MEP EDAP N° ---- 13/10/20 10:28:35 2.0.1009 probe number , Imposible abrir puerto COM del amplificador EDAP N° ---- 13/10/20 10:34:36 1.3.5101 Tratamiento cancelado en la fase de edición de los datos del tratamiento EDAP N° ---- 13/10/20 10:36:34 1.3.5101 Tratamiento cancelado en la fase de edición de los datos del tratamiento CSM N° ---- 03/11/20 13:30:14 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 03/11/20 13:34:10 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 03/11/20 14:42:14 2.3.1901 Código Focalpak incorrecto CSM N° ---- 03/11/20 14:42:20 1.3.5101 Tratamiento cancelado en la fase de edición de los datos del tratamiento CSM N° 0157 03/11/20 15:11:38 2.8.1202 El paciente se ha movido CSM N° 0157 03/11/20 15:11:38 2.8.1202 El paciente se ha movido CSM N° 0157 03/11/20 15:12:33 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (25.7 mm) tiene que estar dentro del intervalo de 23.3 mm a 25.3 mm CSM N° 0157 03/11/20 15:12:49 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (25.9 mm) tiene que estar dentro del intervalo de 23.3 mm a 25.3 mm CSM N° 0157 03/11/20 15:16:11 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0157 03/11/20 15:16:17 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0158 04/11/20 09:43:52 2.5.1202 El paciente se ha movido CSM N° 0158 04/11/20 09:44:10 2.5.1202 El paciente se ha movido CSM N° 0158 04/11/20 10:16:06 2.5.1202 El paciente se ha movido CSM N° 0158 04/11/20 10:31:18 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (17.1 mm) tiene que estar dentro del intervalo de 18.8 mm a 26.8 mm CSM N° 0158 04/11/20 10:31:35 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (17.3 mm) tiene que estar dentro del intervalo de 18.8 mm a 26.8 mm CSM N° 0158 04/11/20 10:34:55 2.5.1202 El paciente se ha movido CSM N° 0158 04/11/20 10:34:59 2.5.1202 El paciente se ha movido CSM N° 0158 04/11/20 10:35:00 2.5.1202 El paciente se ha movido CSM N° 0158 04/11/20 10:41:04 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (21.4 mm) tiene que estar dentro del intervalo de 21.7 mm a 23.7 mm CSM N° 0158 04/11/20 10:41:18 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.3 mm) tiene que estar dentro del intervalo de 23.0 mm a 25.0 mm CSM N° 0158 04/11/20 10:42:18 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.5 mm) tiene que estar dentro del intervalo de 20.1 mm a 22.1 mm CSM N° 0158 04/11/20 10:43:33 2.8.1202 El paciente se ha movido CSM N° 0158 04/11/20 10:47:46 2.8.1202 El paciente se ha movido CSM N° 0158 04/11/20 10:47:46 2.8.1202 El paciente se ha movido CSM N° 0158 04/11/20 10:48:41 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0158 04/11/20 10:48:47 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0158 04/11/20 10:48:53 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0158 04/11/20 10:54:46 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (20.8 mm) tiene que estar dentro del intervalo de 21.3 mm a 23.3 mm CSM N° 0158 04/11/20 10:55:09 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.2 mm) tiene que estar dentro del intervalo de 22.4 mm a 24.4 mm CSM N° 0158 04/11/20 10:55:36 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.3 mm) tiene que estar dentro del intervalo de 22.4 mm a 24.4 mm CSM N° 0158 04/11/20 10:55:58 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.2 mm) tiene que estar dentro del intervalo de 22.4 mm a 24.4 mm CSM N° 0158 04/11/20 10:56:20 2.8.1202 El paciente se ha movido CSM N° 0158 04/11/20 10:56:20 2.8.1202 El paciente se ha movido CSM N° 0158 04/11/20 10:56:24 2.8.1202 El paciente se ha movido CSM N° 0158 04/11/20 10:56:30 2.8.1202 El paciente se ha movido CSM N° 0158 04/11/20 10:56:31 2.8.1202 El paciente se ha movido CSM N° 0158 04/11/20 10:58:11 2.8.1202 El paciente se ha movido CSM N° 0158 04/11/20 10:58:12 2.8.1202 El paciente se ha movido CSM N° 0158 04/11/20 11:00:12 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (25.9 mm) tiene que estar dentro del intervalo de 17.4 mm a 25.4 mm CSM N° 0158 04/11/20 11:00:18 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (25.7 mm) tiene que estar dentro del intervalo de 17.4 mm a 25.4 mm CSM N° 0158 04/11/20 11:01:37 2.8.1202 El paciente se ha movido CSM N° 0158 04/11/20 11:01:37 2.8.1202 El paciente se ha movido CSM N° 0158 04/11/20 11:14:46 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (21.4 mm) tiene que estar dentro del intervalo de 21.6 mm a 29.6 mm CSM N° 0158 04/11/20 11:19:48 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° ---- 11/11/20 09:24:54 2.2.1410 Superado el tiempo de comunicación con longitudinal axis CSM N° 0159 11/11/20 11:22:23 2.5.1202 El paciente se ha movido CSM N° 0159 11/11/20 11:29:21 2.8.1202 El paciente se ha movido CSM N° 0159 11/11/20 11:29:21 2.8.1202 El paciente se ha movido CSM N° 0159 11/11/20 11:30:08 2.8.1202 El paciente se ha movido CSM N° 0159 11/11/20 11:30:08 2.8.1202 El paciente se ha movido CSM N° 0160 18/11/20 13:15:28 2.7.1202 El paciente se ha movido CSM N° 0160 18/11/20 13:24:36 2.8.1202 El paciente se ha movido CSM N° 0160 18/11/20 13:24:36 2.8.1202 El paciente se ha movido CSM N° 0160 18/11/20 13:27:41 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:28:08 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:28:33 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:28:40 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:28:42 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.4 mm) tiene que estar dentro del intervalo de 24.7 mm a 26.7 mm CSM N° 0160 18/11/20 13:28:45 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:28:47 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:28:56 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:28:57 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:28:59 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:29:00 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:29:02 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:29:04 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:29:09 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:29:11 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:29:12 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:39:49 2.7.1202 El paciente se ha movido CSM N° 0160 18/11/20 13:40:07 2.7.1202 El paciente se ha movido CSM N° 0160 18/11/20 13:44:57 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.4 mm) tiene que estar dentro del intervalo de 20.3 mm a 22.3 mm CSM N° 0160 18/11/20 13:46:37 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.2 mm) tiene que estar dentro del intervalo de 20.1 mm a 22.1 mm CSM N° 0160 18/11/20 13:48:44 2.7.1202 El paciente se ha movido CSM N° 0160 18/11/20 13:49:55 2.8.1202 El paciente se ha movido CSM N° 0160 18/11/20 13:49:55 2.8.1202 El paciente se ha movido CSM N° 0160 18/11/20 13:50:35 2.8.1202 El paciente se ha movido CSM N° 0160 18/11/20 13:50:35 2.8.1202 El paciente se ha movido CSM N° 0160 18/11/20 13:50:39 2.8.1202 El paciente se ha movido CSM N° 0160 18/11/20 13:50:54 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:50:59 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:51:02 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:52:10 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:52:16 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0160 18/11/20 13:53:54 2.5.1202 El paciente se ha movido CSM N° 0160 18/11/20 13:54:02 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° 0160 18/11/20 13:54:06 2.5.1202 El paciente se ha movido CSM N° 0160 18/11/20 13:58:44 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (16.4 mm) tiene que estar dentro del intervalo de 17.4 mm a 25.4 mm CSM N° 0161 10/12/20 14:59:12 2.5.1202 El paciente se ha movido CSM N° 0161 10/12/20 14:59:15 2.5.1202 El paciente se ha movido CSM N° 0161 10/12/20 14:59:16 2.5.1202 El paciente se ha movido CSM N° 0161 10/12/20 15:00:08 2.5.1202 El paciente se ha movido CSM N° 0161 10/12/20 15:00:10 2.5.1202 El paciente se ha movido CSM N° 0161 10/12/20 15:01:53 2.5.1202 El paciente se ha movido CSM N° 0161 10/12/20 15:28:16 2.8.1202 El paciente se ha movido CSM N° 0161 10/12/20 15:28:16 2.8.1202 El paciente se ha movido CSM N° 0162 02/01/21 13:11:28 2.8.1202 El paciente se ha movido CSM N° 0162 02/01/21 13:11:28 2.8.1202 El paciente se ha movido CSM N° 0162 02/01/21 13:15:41 2.7.1202 El paciente se ha movido CSM N° 0162 02/01/21 13:16:25 2.8.1202 El paciente se ha movido CSM N° 0162 02/01/21 13:16:25 2.8.1202 El paciente se ha movido CSM N° 0162 02/01/21 13:24:04 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (19.2 mm) tiene que estar dentro del intervalo de 21.0 mm a 29.0 mm CSM N° 0162 02/01/21 13:34:57 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (18.3 mm) tiene que estar dentro del intervalo de 18.8 mm a 26.8 mm CSM N° 0162 02/01/21 13:35:08 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (19.6 mm) tiene que estar dentro del intervalo de 20.1 mm a 28.1 mm CSM N° 0162 02/01/21 13:35:19 2.8.1202 El paciente se ha movido CSM N° 0162 02/01/21 13:38:56 2.8.1439 No se detectó movimiento en rotary axis CSM N° 0162 02/01/21 13:39:00 2.8.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:39:05 2.8.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:39:18 2.9.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:39:27 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:39:42 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:39:45 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:39:54 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° 0162 02/01/21 13:39:58 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:40:03 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° 0162 02/01/21 13:40:31 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:40:40 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:41:06 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:41:15 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° 0162 02/01/21 13:42:08 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:42:21 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:42:26 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:42:27 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:42:27 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:42:32 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:42:35 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:42:50 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:42:52 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:42:53 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:43:41 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:43:46 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° 0162 02/01/21 13:43:53 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:44:20 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° 0162 02/01/21 13:44:31 2.5.1437 Centrado no realizado con éxito en rotary axis CSM N° 0162 02/01/21 13:45:04 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° ---- 02/01/21 13:47:26 2.3.1901 Código Focalpak incorrecto CSM N° ---- 02/01/21 13:47:53 2.3.1901 Código Focalpak incorrecto CSM N° ---- 02/01/21 13:48:00 1.3.5101 Tratamiento cancelado en la fase de edición de los datos del tratamiento CSM N° ---- 02/01/21 13:50:45 2.3.1902 Código Focalpak ya utilizado CSM N° 0163 02/01/21 14:12:51 1.9.5105 Tratamiento cancelado durante la fase de disparo CSM N° ---- 09/01/21 13:07:55 2.3.1902 Código Focalpak ya utilizado CSM N° 0164 09/01/21 13:13:49 2.5.1202 El paciente se ha movido CSM N° 0164 09/01/21 13:29:34 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.5 %). CSM N° 0164 09/01/21 13:30:11 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, CSM N° 0164 09/01/21 13:32:01 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.9 %). CSM N° 0164 09/01/21 13:32:08 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, CSM N° 0164 09/01/21 13:35:36 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.8 %). CSM N° 0164 09/01/21 13:35:42 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.9 %). CSM N° 0164 09/01/21 13:35:48 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.9 %). CSM N° 0164 09/01/21 13:35:50 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, CSM N° 0164 09/01/21 13:37:36 1.9.5105 Tratamiento cancelado durante la fase de disparo CSM N° 0164 09/01/21 13:37:48 2.7.1202 El paciente se ha movido CSM N° ---- 23/01/21 11:45:08 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 23/01/21 11:45:21 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 23/01/21 11:46:13 2.2.1604 El escáner no puede pasar a modo operativo CSM N° ---- 23/01/21 11:46:34 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 23/01/21 11:47:18 2.2.1604 El escáner no puede pasar a modo operativo CSM N° ---- 23/01/21 12:17:50 2.2.1451 Error al inicializar la comunicación en el motor: 1 2 3 CSM N° 0165 23/01/21 13:34:19 2.5.1202 El paciente se ha movido CSM N° 0165 23/01/21 13:34:21 2.5.1202 El paciente se ha movido CSM N° 0165 23/01/21 13:36:21 2.6.1202 El paciente se ha movido CSM N° 0165 23/01/21 13:36:25 2.6.1202 El paciente se ha movido CSM N° 0165 23/01/21 13:37:45 2.6.1202 El paciente se ha movido CSM N° 0165 23/01/21 13:37:47 2.6.1202 El paciente se ha movido CSM N° 0165 23/01/21 13:58:26 2.8.1202 El paciente se ha movido CSM N° 0165 23/01/21 13:58:26 2.8.1202 El paciente se ha movido CSM N° 0165 23/01/21 14:03:26 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.6 mm) tiene que estar dentro del intervalo de 21.5 mm a 23.5 mm CSM N° 0165 23/01/21 14:07:23 2.7.1202 El paciente se ha movido CSM N° 0165 23/01/21 14:15:03 2.8.1202 El paciente se ha movido CSM N° 0165 23/01/21 14:15:03 2.8.1202 El paciente se ha movido CSM N° 0165 23/01/21 14:15:35 2.8.1202 El paciente se ha movido CSM N° 0165 23/01/21 14:15:35 2.8.1202 El paciente se ha movido CSM N° 0165 23/01/21 14:20:23 2.8.1202 El paciente se ha movido CSM N° 0165 23/01/21 14:20:23 2.8.1202 El paciente se ha movido CSM N° 0165 23/01/21 14:35:34 2.8.1202 El paciente se ha movido CSM N° 0165 23/01/21 14:35:34 2.8.1202 El paciente se ha movido CSM N° 0165 23/01/21 14:35:37 2.8.1202 El paciente se ha movido CSM N° 0165 23/01/21 14:35:38 2.8.1202 El paciente se ha movido CSM N° 0165 23/01/21 14:35:50 2.8.1441 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje vertical CSM N° 0165 23/01/21 14:36:08 2.7.1202 El paciente se ha movido CSM N° 0166 06/02/21 13:04:53 2.5.1202 El paciente se ha movido CSM N° 0166 06/02/21 13:05:08 2.5.1102 Temperatura de la sonda excesiva : 30.6 CSM N° 0166 06/02/21 13:05:16 2.5.1102 Temperatura de la sonda excesiva : 30.6 CSM N° 0166 06/02/21 13:05:18 2.5.1102 Temperatura de la sonda excesiva : 30.6 CSM N° 0166 06/02/21 13:05:27 2.5.1102 Temperatura de la sonda excesiva : 30.7 CSM N° 0166 06/02/21 13:05:29 2.5.1102 Temperatura de la sonda excesiva : 30.8 CSM N° 0166 06/02/21 13:05:32 2.5.1102 Temperatura de la sonda excesiva : 30.8 CSM N° 0166 06/02/21 13:06:03 2.5.1102 Temperatura de la sonda excesiva : 30.9 CSM N° 0166 06/02/21 13:06:06 2.5.1102 Temperatura de la sonda excesiva : 31.0 CSM N° 0166 06/02/21 13:06:08 2.5.1102 Temperatura de la sonda excesiva : 31.0 CSM N° 0166 06/02/21 13:06:12 2.5.1102 Temperatura de la sonda excesiva : 31.0 CSM N° 0166 06/02/21 13:06:20 2.5.1102 Temperatura de la sonda excesiva : 31.1 CSM N° 0166 06/02/21 13:06:22 2.5.1102 Temperatura de la sonda excesiva : 31.2 CSM N° 0166 06/02/21 13:06:34 2.5.1102 Temperatura de la sonda excesiva : 31.3 CSM N° 0166 06/02/21 13:06:35 2.5.1102 Temperatura de la sonda excesiva : 31.3 CSM N° 0166 06/02/21 13:06:44 2.5.1102 Temperatura de la sonda excesiva : 31.3 CSM N° 0166 06/02/21 13:07:27 2.5.1102 Temperatura de la sonda excesiva : 31.8 CSM N° 0166 06/02/21 13:07:29 2.5.1102 Temperatura de la sonda excesiva : 31.8 CSM N° 0166 06/02/21 13:08:51 2.5.1102 Temperatura de la sonda excesiva : 30.1 CSM N° 0166 06/02/21 13:08:55 2.5.1102 Temperatura de la sonda excesiva : 30.0 CSM N° 0166 06/02/21 13:09:00 2.5.1102 Temperatura de la sonda excesiva : 30.2 CSM N° 0166 06/02/21 13:09:02 2.5.1102 Temperatura de la sonda excesiva : 30.3 CSM N° 0166 06/02/21 13:10:01 2.5.1102 Temperatura de la sonda excesiva : 29.8 CSM N° 0166 06/02/21 13:10:04 2.5.1102 Temperatura de la sonda excesiva : 29.8 CSM N° 0166 06/02/21 13:10:09 2.5.1102 Temperatura de la sonda excesiva : 29.8 CSM N° 0166 06/02/21 13:10:19 2.5.1102 Temperatura de la sonda excesiva : 30.0 CSM N° 0166 06/02/21 13:10:29 2.5.1102 Temperatura de la sonda excesiva : 30.3 CSM N° 0166 06/02/21 13:11:04 2.5.1102 Temperatura de la sonda excesiva : 29.5 CSM N° 0166 06/02/21 13:11:11 2.5.1102 Temperatura de la sonda excesiva : 29.6 CSM N° 0166 06/02/21 13:11:19 2.5.1102 Temperatura de la sonda excesiva : 29.7 CSM N° 0166 06/02/21 13:11:21 2.5.1102 Temperatura de la sonda excesiva : 29.7 CSM N° 0166 06/02/21 13:11:50 2.5.1102 Temperatura de la sonda excesiva : 29.7 CSM N° 0166 06/02/21 13:11:51 2.5.1102 Temperatura de la sonda excesiva : 29.3 CSM N° 0166 06/02/21 13:11:53 2.5.1102 Temperatura de la sonda excesiva : 28.4 CSM N° 0166 06/02/21 13:12:02 2.5.1102 Temperatura de la sonda excesiva : 25.6 CSM N° 0166 06/02/21 13:12:13 2.5.1102 Temperatura de la sonda excesiva : 22.1 CSM N° 0166 06/02/21 13:12:15 2.5.1102 Temperatura de la sonda excesiva : 21.0 CSM N° 0166 06/02/21 13:12:22 2.5.1102 Temperatura de la sonda excesiva : 19.6 CSM N° 0166 06/02/21 13:12:24 2.5.1102 Temperatura de la sonda excesiva : 19.8 CSM N° 0166 06/02/21 13:12:28 2.5.1102 Temperatura de la sonda excesiva : 19.7 CSM N° 0166 06/02/21 13:37:27 2.8.1423 La posición real está más allá de la posición solicitada en antero-posterior axis CSM N° 0166 06/02/21 13:59:59 2.7.1202 El paciente se ha movido CSM N° 0166 06/02/21 14:00:07 2.7.1202 El paciente se ha movido CSM N° ---- 06/02/21 14:14:08 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 06/02/21 14:31:31 2.2.1502 No se puede comunicar con la sonda CSM N° 0167 06/02/21 15:02:59 2.5.1202 El paciente se ha movido CSM N° 0167 06/02/21 15:06:12 2.6.1202 El paciente se ha movido CSM N° 0167 06/02/21 15:06:20 2.6.1202 El paciente se ha movido CSM N° 0167 06/02/21 15:08:43 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (18.5 mm) tiene que estar dentro del intervalo de 20.0 mm a 28.0 mm CSM N° 0167 06/02/21 15:08:56 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.1 mm) tiene que estar dentro del intervalo de 23.0 mm a 25.0 mm CSM N° 0167 06/02/21 15:09:12 2.8.1202 El paciente se ha movido CSM N° 0167 06/02/21 15:09:34 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (21.5 mm) tiene que estar dentro del intervalo de 23.0 mm a 25.0 mm CSM N° 0167 06/02/21 15:09:39 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (21.3 mm) tiene que estar dentro del intervalo de 23.0 mm a 25.0 mm CSM N° 0167 06/02/21 15:10:12 2.8.1202 El paciente se ha movido CSM N° 0167 06/02/21 15:10:15 2.8.1202 El paciente se ha movido CSM N° 0167 06/02/21 15:13:05 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (18.6 mm) tiene que estar dentro del intervalo de 20.5 mm a 28.5 mm CSM N° 0167 06/02/21 15:13:17 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (18.4 mm) tiene que estar dentro del intervalo de 20.5 mm a 28.5 mm CSM N° 0167 06/02/21 15:15:26 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (16.6 mm) tiene que estar dentro del intervalo de 19.5 mm a 27.5 mm CSM N° 0167 06/02/21 15:16:01 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (18.6 mm) tiene que estar dentro del intervalo de 20.0 mm a 28.0 mm CSM N° 0167 06/02/21 15:18:36 2.5.1202 El paciente se ha movido CSM N° 0167 06/02/21 15:19:15 2.5.1102 Temperatura de la sonda excesiva : 20.4 CSM N° 0167 06/02/21 15:19:38 2.5.1102 Temperatura de la sonda excesiva : 21.1 CSM N° 0167 06/02/21 15:19:54 2.5.1202 El paciente se ha movido CSM N° 0167 06/02/21 15:19:56 2.5.1102 Temperatura de la sonda excesiva : 20.0 CSM N° 0167 06/02/21 15:20:30 2.5.1102 Temperatura de la sonda excesiva : 20.5 CSM N° 0167 06/02/21 15:20:52 2.5.1102 Temperatura de la sonda excesiva : 21.6 CSM N° 0167 06/02/21 15:20:53 2.5.1102 Temperatura de la sonda excesiva : 21.8 CSM N° 0167 06/02/21 15:20:55 2.5.1102 Temperatura de la sonda excesiva : 21.8 CSM N° 0167 06/02/21 15:20:56 2.5.1102 Temperatura de la sonda excesiva : 21.9 CSM N° 0167 06/02/21 15:22:00 2.5.1102 Temperatura de la sonda excesiva : 20.6 CSM N° 0167 06/02/21 15:22:17 2.5.1102 Temperatura de la sonda excesiva : 21.4 CSM N° 0167 06/02/21 15:22:18 2.5.1102 Temperatura de la sonda excesiva : 21.3 CSM N° 0167 06/02/21 15:23:08 2.5.1102 Temperatura de la sonda excesiva : 22.6 CSM N° 0167 06/02/21 15:23:09 2.5.1102 Temperatura de la sonda excesiva : 22.7 CSM N° 0167 06/02/21 15:23:11 2.5.1102 Temperatura de la sonda excesiva : 22.8 CSM N° 0167 06/02/21 15:23:12 2.5.1102 Temperatura de la sonda excesiva : 22.8 CSM N° 0167 06/02/21 15:23:13 2.5.1102 Temperatura de la sonda excesiva : 22.8 CSM N° 0167 06/02/21 15:24:50 2.5.1102 Temperatura de la sonda excesiva : 22.0 CSM N° 0167 06/02/21 15:27:11 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° 0167 06/02/21 15:27:14 2.5.1102 Temperatura de la sonda excesiva : 20.2 CSM N° 0168 13/03/21 13:13:46 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (69.5 %). CSM N° 0168 13/03/21 13:13:52 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, CSM N° 0168 13/03/21 13:13:55 2.8.1002 probe number UV138-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (78.5 W). CSM N° 0168 13/03/21 13:13:58 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 8, 9, 10, 11, 12, 13, 14, 15, 16, CSM N° 0168 13/03/21 13:14:01 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.5 %). CSM N° 0168 13/03/21 13:14:06 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.8 %). CSM N° 0168 13/03/21 13:14:14 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.7 %). CSM N° 0168 13/03/21 13:14:18 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.8 %). CSM N° 0168 13/03/21 13:14:22 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.7 %). CSM N° 0168 13/03/21 13:14:26 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.0 %). CSM N° 0168 13/03/21 13:14:32 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.1 %). CSM N° 0168 13/03/21 13:14:35 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.1 %). CSM N° 0168 13/03/21 13:14:40 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.3 %). CSM N° 0168 13/03/21 13:14:49 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.0 %). CSM N° 0168 13/03/21 13:14:50 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 8, 9, 10, 11, 12, 15, CSM N° 0168 13/03/21 13:14:54 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.9 %). CSM N° 0168 13/03/21 13:15:00 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.0 %). CSM N° 0168 13/03/21 13:15:02 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 8, 9, 10, 11, 12, 15, CSM N° 0168 13/03/21 13:15:10 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (73.3 %). CSM N° 0168 13/03/21 13:15:30 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0168 13/03/21 13:15:33 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (70.5 %). CSM N° 0168 13/03/21 13:15:37 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0168 13/03/21 13:15:52 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (70.6 %). CSM N° 0168 13/03/21 13:15:54 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0168 13/03/21 13:15:59 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.6 %). CSM N° 0168 13/03/21 13:16:00 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 8, 10, 12, 15, CSM N° 0168 13/03/21 13:16:11 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.7 %). CSM N° 0168 13/03/21 13:16:12 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 8, 10, 11, 12, 15, CSM N° 0168 13/03/21 13:16:19 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.8 %). CSM N° 0168 13/03/21 13:16:26 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.8 %). CSM N° 0168 13/03/21 13:16:31 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (90.0 %). CSM N° 0168 13/03/21 13:16:35 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.9 %). CSM N° 0168 13/03/21 13:16:40 2.8.1002 probe number UV138-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (83.3 W). CSM N° 0168 13/03/21 13:16:41 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 8, 9, 10, 11, 12, 15, CSM N° 0168 13/03/21 13:16:46 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (75.7 %). CSM N° 0168 13/03/21 13:16:47 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0168 13/03/21 13:16:54 2.8.1002 probe number UV138-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (72.2 W). CSM N° 0168 13/03/21 13:16:55 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0168 13/03/21 13:16:59 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (75.5 %). CSM N° 0168 13/03/21 13:17:00 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0168 13/03/21 13:17:05 2.8.1002 probe number UV138-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (83.3 W). CSM N° 0168 13/03/21 13:17:12 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0168 13/03/21 13:17:16 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.3 %). CSM N° 0168 13/03/21 13:17:17 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 8, 9, 11, 12, 13, 15, CSM N° 0168 13/03/21 13:17:22 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.9 %). CSM N° 0168 13/03/21 13:17:31 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (82.9 %). CSM N° 0168 13/03/21 13:17:33 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 8, 9, 10, 11, 12, 13, 15, CSM N° 0168 13/03/21 13:17:42 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.6 %). CSM N° 0168 13/03/21 13:17:42 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 8, 9, 10, 12, 13, 15, CSM N° 0168 13/03/21 13:17:51 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.0 %). CSM N° 0168 13/03/21 13:17:59 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (77.0 %). CSM N° 0168 13/03/21 13:18:00 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0168 13/03/21 13:18:03 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (74.1 %). CSM N° 0168 13/03/21 13:18:04 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0168 13/03/21 13:18:10 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (74.8 %). CSM N° 0168 13/03/21 13:18:18 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0168 13/03/21 13:19:29 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° 0169 13/03/21 13:51:31 2.6.1108 Sonda terapéutica no conectada. CSM N° 0169 13/03/21 13:51:37 2.6.1108 Sonda terapéutica no conectada.: Impossible to read temperature CSM N° 0169 13/03/21 13:52:08 2.6.1108 Sonda terapéutica no conectada.: Impossible to read temperature CSM N° 0169 13/03/21 13:52:35 2.6.1108 Sonda terapéutica no conectada. CSM N° 0169 13/03/21 13:52:36 2.6.1202 El paciente se ha movido CSM N° 0169 13/03/21 13:52:37 2.6.1108 Sonda terapéutica no conectada.: Impossible to read temperature CSM N° 0169 13/03/21 13:52:38 2.6.1108 Sonda terapéutica no conectada.: Impossible to read temperature CSM N° 0169 13/03/21 13:52:38 2.6.1108 Sonda terapéutica no conectada. CSM N° 0169 13/03/21 13:52:40 2.6.1108 Sonda terapéutica no conectada.: Impossible to read temperature CSM N° 0169 13/03/21 13:53:06 2.6.1108 Sonda terapéutica no conectada. CSM N° 0169 13/03/21 13:53:07 2.6.1108 Sonda terapéutica no conectada.: Impossible to read temperature CSM N° 0169 13/03/21 13:53:08 2.6.1108 Sonda terapéutica no conectada.: Impossible to read temperature CSM N° 0169 13/03/21 13:53:09 2.6.1108 Sonda terapéutica no conectada. CSM N° 0169 13/03/21 13:53:26 2.6.1101 Temperatura de la sonda insuficiente : 12.2 CSM N° 0169 13/03/21 13:53:27 2.6.1101 Temperatura de la sonda insuficiente : 12.2 CSM N° 0169 13/03/21 13:53:28 2.6.1101 Temperatura de la sonda insuficiente : 12.2 CSM N° 0169 13/03/21 13:57:27 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (24.5 mm) tiene que estar dentro del intervalo de 22.3 mm a 24.3 mm CSM N° 0169 13/03/21 13:58:06 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.0 %). CSM N° 0169 13/03/21 13:58:09 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0169 13/03/21 13:58:20 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (78.8 W). CSM N° 0169 13/03/21 13:58:22 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0169 13/03/21 13:58:25 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.9 %). CSM N° 0169 13/03/21 13:58:26 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 10, 11, 12, 13, 14, 15, CSM N° 0169 13/03/21 13:58:35 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (87.5 W). CSM N° 0169 13/03/21 13:58:42 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.7 %). CSM N° 0169 13/03/21 13:58:46 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.6 W). CSM N° 0169 13/03/21 13:58:52 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.9 W). CSM N° 0169 13/03/21 13:58:57 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.3 %). CSM N° 0169 13/03/21 13:58:58 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 9, 10, 11, 12, 13, 14, 15, CSM N° 0169 13/03/21 13:59:01 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.1 %). CSM N° 0169 13/03/21 13:59:02 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 10, 11, 12, 13, 14, 15, CSM N° 0169 13/03/21 13:59:05 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (77.6 %). CSM N° 0169 13/03/21 13:59:07 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0169 13/03/21 13:59:10 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (77.3 %). CSM N° 0169 13/03/21 13:59:13 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0169 13/03/21 13:59:15 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (77.0 %). CSM N° 0169 13/03/21 13:59:16 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0169 13/03/21 13:59:18 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (75.3 %). CSM N° 0169 13/03/21 13:59:19 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0169 13/03/21 13:59:21 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (74.6 %). CSM N° 0169 13/03/21 13:59:22 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0169 13/03/21 13:59:25 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (74.2 %). CSM N° 0169 13/03/21 13:59:26 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0169 13/03/21 13:59:29 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (74.5 %). CSM N° 0169 13/03/21 13:59:32 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0169 13/03/21 13:59:35 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (75.6 %). CSM N° 0169 13/03/21 13:59:36 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0169 13/03/21 13:59:38 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (75.3 %). CSM N° 0169 13/03/21 13:59:39 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0169 13/03/21 13:59:42 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (74.7 %). CSM N° 0169 13/03/21 13:59:43 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0169 13/03/21 13:59:45 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (74.8 %). CSM N° 0169 13/03/21 13:59:46 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0169 13/03/21 13:59:49 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (75.1 %). CSM N° 0169 13/03/21 13:59:50 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0169 13/03/21 13:59:52 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (75.1 %). CSM N° 0169 13/03/21 13:59:53 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 14, CSM N° 0169 13/03/21 13:59:55 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (75.3 %). CSM N° 0169 13/03/21 13:59:56 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 13:59:59 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (75.2 %). CSM N° 0169 13/03/21 14:00:00 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:00:02 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (75.2 %). CSM N° 0169 13/03/21 14:00:03 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0169 13/03/21 14:00:05 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (75.3 %). CSM N° 0169 13/03/21 14:00:12 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0169 13/03/21 14:00:15 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (77.5 %). CSM N° 0169 13/03/21 14:00:16 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:00:20 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.2 %). CSM N° 0169 13/03/21 14:00:21 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 15, CSM N° 0169 13/03/21 14:00:24 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.2 %). CSM N° 0169 13/03/21 14:00:24 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 15, CSM N° 0169 13/03/21 14:00:28 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.3 W). CSM N° 0169 13/03/21 14:00:34 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.3 W). CSM N° 0169 13/03/21 14:00:39 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.7 %). CSM N° 0169 13/03/21 14:00:43 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.3 W). CSM N° 0169 13/03/21 14:00:48 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.4 W). CSM N° 0169 13/03/21 14:00:55 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.5 %). CSM N° 0169 13/03/21 14:01:00 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.0 W). CSM N° 0169 13/03/21 14:01:04 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.0 W). CSM N° 0169 13/03/21 14:01:10 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.4 %). CSM N° 0169 13/03/21 14:01:12 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:01:14 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.1 %). CSM N° 0169 13/03/21 14:01:15 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:01:20 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (77.6 %). CSM N° 0169 13/03/21 14:01:21 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:01:25 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (77.2 %). CSM N° 0169 13/03/21 14:01:26 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:01:29 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.7 %). CSM N° 0169 13/03/21 14:01:33 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:01:38 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (82.1 %). CSM N° 0169 13/03/21 14:01:39 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:01:42 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.3 W). CSM N° 0169 13/03/21 14:01:46 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.3 W). CSM N° 0169 13/03/21 14:01:51 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.1 %). CSM N° 0169 13/03/21 14:01:57 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.1 W). CSM N° 0169 13/03/21 14:02:57 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (78.1 %). CSM N° 0169 13/03/21 14:02:59 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:03:07 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (77.4 %). CSM N° 0169 13/03/21 14:03:08 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:03:11 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.5 %). CSM N° 0169 13/03/21 14:03:13 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 15, CSM N° 0169 13/03/21 14:03:16 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.7 %). CSM N° 0169 13/03/21 14:03:17 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 15, CSM N° 0169 13/03/21 14:03:22 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.1 W). CSM N° 0169 13/03/21 14:03:26 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.8 W). CSM N° 0169 13/03/21 14:03:31 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.1 %). CSM N° 0169 13/03/21 14:03:41 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.4 %). CSM N° 0169 13/03/21 14:03:49 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.2 %). CSM N° 0169 13/03/21 14:03:59 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.8 %). CSM N° 0169 13/03/21 14:04:08 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.7 %). CSM N° 0169 13/03/21 14:04:19 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.5 %). CSM N° 0169 13/03/21 14:04:25 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.4 %). CSM N° 0169 13/03/21 14:04:41 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.4 %). CSM N° 0169 13/03/21 14:04:48 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.6 %). CSM N° 0169 13/03/21 14:04:56 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.7 %). CSM N° 0169 13/03/21 14:05:03 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.9 W). CSM N° 0169 13/03/21 14:05:07 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.4 W). CSM N° 0169 13/03/21 14:06:45 2.8.1102 Temperatura de la sonda excesiva : 103.9 CSM N° 0169 13/03/21 14:06:48 2.8.1102 Temperatura de la sonda excesiva : 106.4 CSM N° 0169 13/03/21 14:06:49 2.8.1102 Temperatura de la sonda excesiva : 107.4 CSM N° 0169 13/03/21 14:06:53 2.8.1102 Temperatura de la sonda excesiva : 104.1 CSM N° 0169 13/03/21 14:06:54 2.8.1102 Temperatura de la sonda excesiva : 103.7 CSM N° 0169 13/03/21 14:06:56 2.8.1102 Temperatura de la sonda excesiva : 101.4 CSM N° 0169 13/03/21 14:07:31 2.8.1102 Temperatura de la sonda excesiva : 24.9 CSM N° 0169 13/03/21 14:07:37 2.8.1102 Temperatura de la sonda excesiva : 21.6 CSM N° 0169 13/03/21 14:07:46 2.8.1102 Temperatura de la sonda excesiva : 8.8 CSM N° 0169 13/03/21 14:07:59 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (79.4 %). CSM N° 0169 13/03/21 14:08:01 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 15, CSM N° 0169 13/03/21 14:08:04 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (78.7 %). CSM N° 0169 13/03/21 14:08:04 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 15, CSM N° 0169 13/03/21 14:08:10 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (75.5 W). CSM N° 0169 13/03/21 14:08:11 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0169 13/03/21 14:08:16 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (76.2 %). CSM N° 0169 13/03/21 14:08:17 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0169 13/03/21 14:08:21 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (75.3 W). CSM N° 0169 13/03/21 14:08:21 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0169 13/03/21 14:08:25 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (79.0 %). CSM N° 0169 13/03/21 14:08:26 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 15, CSM N° 0169 13/03/21 14:08:29 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (79.3 %). CSM N° 0169 13/03/21 14:08:30 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 15, CSM N° 0169 13/03/21 14:08:59 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (76.6 W). CSM N° 0169 13/03/21 14:09:00 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0169 13/03/21 14:09:04 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (79.8 %). CSM N° 0169 13/03/21 14:09:09 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 15, CSM N° 0169 13/03/21 14:09:12 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.7 %). CSM N° 0169 13/03/21 14:09:13 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 15, CSM N° 0169 13/03/21 14:09:18 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.9 W). CSM N° 0169 13/03/21 14:09:23 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.3 W). CSM N° 0169 13/03/21 14:09:30 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.3 %). CSM N° 0169 13/03/21 14:09:37 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.8 %). CSM N° 0169 13/03/21 14:09:49 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.0 %). CSM N° 0169 13/03/21 14:09:58 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.5 %). CSM N° 0169 13/03/21 14:10:06 2.8.1102 Temperatura de la sonda excesiva : 8.6 CSM N° 0169 13/03/21 14:10:09 2.8.1102 Temperatura de la sonda excesiva : 8.9 CSM N° 0169 13/03/21 14:10:19 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.8 %). CSM N° 0169 13/03/21 14:10:31 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.1 %). CSM N° 0169 13/03/21 14:10:39 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.9 %). CSM N° 0169 13/03/21 14:10:48 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.7 %). CSM N° 0169 13/03/21 14:10:52 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.0 W). CSM N° 0169 13/03/21 14:10:56 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (84.9 W). CSM N° 0169 13/03/21 14:11:00 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (79.6 %). CSM N° 0169 13/03/21 14:11:01 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 15, CSM N° 0169 13/03/21 14:11:04 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (79.4 %). CSM N° 0169 13/03/21 14:11:05 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 15, CSM N° 0169 13/03/21 14:11:12 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (76.6 %). CSM N° 0169 13/03/21 14:11:13 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0169 13/03/21 14:11:19 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (76.2 W). CSM N° 0169 13/03/21 14:11:21 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0169 13/03/21 14:11:29 2.8.1102 Temperatura de la sonda excesiva : 12.9 CSM N° 0169 13/03/21 14:12:38 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (76.7 %). CSM N° 0169 13/03/21 14:12:40 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0169 13/03/21 14:12:45 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.3 %). CSM N° 0169 13/03/21 14:12:47 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 15, CSM N° 0169 13/03/21 14:12:50 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.2 W). CSM N° 0169 13/03/21 14:12:54 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (84.8 W). CSM N° 0169 13/03/21 14:12:59 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.7 %). CSM N° 0169 13/03/21 14:13:07 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.4 %). CSM N° 0169 13/03/21 14:13:16 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.4 %). CSM N° 0169 13/03/21 14:13:26 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.7 %). CSM N° 0169 13/03/21 14:13:32 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (79.3 %). CSM N° 0169 13/03/21 14:13:34 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 15, CSM N° 0169 13/03/21 14:13:37 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (75.9 %). CSM N° 0169 13/03/21 14:13:39 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0169 13/03/21 14:14:33 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (76.5 %). CSM N° 0169 13/03/21 14:14:34 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0169 13/03/21 14:14:37 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.0 W). CSM N° 0169 13/03/21 14:14:41 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (84.6 W). CSM N° 0169 13/03/21 14:14:48 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.0 %). CSM N° 0169 13/03/21 14:14:56 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.4 %). CSM N° 0169 13/03/21 14:15:03 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.1 %). CSM N° 0169 13/03/21 14:15:13 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.9 %). CSM N° 0169 13/03/21 14:15:19 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.9 %). CSM N° 0169 13/03/21 14:15:26 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.1 %). CSM N° 0169 13/03/21 14:15:33 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.9 %). CSM N° 0169 13/03/21 14:15:40 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.0 %). CSM N° 0169 13/03/21 14:15:46 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.6 %). CSM N° 0169 13/03/21 14:15:54 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.2 %). CSM N° 0169 13/03/21 14:16:02 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.4 %). CSM N° 0169 13/03/21 14:16:11 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.4 %). CSM N° 0169 13/03/21 14:16:18 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.4 %). CSM N° 0169 13/03/21 14:16:26 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.3 %). CSM N° 0169 13/03/21 14:16:30 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (84.7 W). CSM N° 0169 13/03/21 14:16:34 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (79.2 %). CSM N° 0169 13/03/21 14:16:38 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 15, CSM N° 0169 13/03/21 14:16:43 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (76.3 %). CSM N° 0169 13/03/21 14:16:44 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0169 13/03/21 14:16:47 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (76.3 %). CSM N° 0169 13/03/21 14:16:48 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0169 13/03/21 14:16:52 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (76.6 %). CSM N° 0169 13/03/21 14:16:55 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:17:00 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.2 %). CSM N° 0169 13/03/21 14:17:01 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 15, CSM N° 0169 13/03/21 14:17:05 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.2 %). CSM N° 0169 13/03/21 14:17:06 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 15, CSM N° 0169 13/03/21 14:17:09 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (84.9 W). CSM N° 0169 13/03/21 14:17:16 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.1 %). CSM N° 0169 13/03/21 14:17:24 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.6 %). CSM N° 0169 13/03/21 14:17:33 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.6 %). CSM N° 0169 13/03/21 14:17:41 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.2 %). CSM N° 0169 13/03/21 14:17:50 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.3 %). CSM N° 0169 13/03/21 14:17:57 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.5 %). CSM N° 0169 13/03/21 14:18:04 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.1 %). CSM N° 0169 13/03/21 14:18:11 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.2 %). CSM N° 0169 13/03/21 14:18:17 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.2 %). CSM N° 0169 13/03/21 14:18:23 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.0 %). CSM N° 0169 13/03/21 14:18:31 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.4 %). CSM N° 0169 13/03/21 14:18:39 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.4 %). CSM N° 0169 13/03/21 14:18:47 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.5 %). CSM N° 0169 13/03/21 14:18:56 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.8 %). CSM N° 0169 13/03/21 14:19:00 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.2 W). CSM N° 0169 13/03/21 14:19:04 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.3 W). CSM N° 0169 13/03/21 14:19:12 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.7 %). CSM N° 0169 13/03/21 14:19:13 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:19:16 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.7 %). CSM N° 0169 13/03/21 14:19:18 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:19:21 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.9 %). CSM N° 0169 13/03/21 14:19:23 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:19:26 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.8 %). CSM N° 0169 13/03/21 14:19:27 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:19:30 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.6 W). CSM N° 0169 13/03/21 14:19:34 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.6 W). CSM N° 0169 13/03/21 14:19:41 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.8 %). CSM N° 0169 13/03/21 14:19:49 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.1 %). CSM N° 0169 13/03/21 14:19:58 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.1 %). CSM N° 0169 13/03/21 14:20:07 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.4 %). CSM N° 0169 13/03/21 14:20:12 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.5 %). CSM N° 0169 13/03/21 14:20:18 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.8 %). CSM N° 0169 13/03/21 14:20:30 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.6 %). CSM N° 0169 13/03/21 14:21:27 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (75.9 %). CSM N° 0169 13/03/21 14:21:28 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:21:30 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (79.8 %). CSM N° 0169 13/03/21 14:21:32 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:21:35 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (79.9 %). CSM N° 0169 13/03/21 14:21:36 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:21:39 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.3 W). CSM N° 0169 13/03/21 14:21:43 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.1 W). CSM N° 0169 13/03/21 14:21:48 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.2 %). CSM N° 0169 13/03/21 14:21:57 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.2 %). CSM N° 0169 13/03/21 14:22:05 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.0 %). CSM N° 0169 13/03/21 14:22:14 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.2 %). CSM N° 0169 13/03/21 14:22:36 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.7 %). CSM N° 0169 13/03/21 14:22:47 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.1 %). CSM N° 0169 13/03/21 14:22:55 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.1 %). CSM N° 0169 13/03/21 14:23:01 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.2 W). CSM N° 0169 13/03/21 14:23:05 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.2 W). CSM N° 0169 13/03/21 14:23:06 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 7, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:23:11 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.4 %). CSM N° 0169 13/03/21 14:23:12 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:23:15 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.3 %). CSM N° 0169 13/03/21 14:23:16 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, CSM N° 0169 13/03/21 14:23:19 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (77.8 %). CSM N° 0169 13/03/21 14:23:19 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:23:24 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (77.6 W). CSM N° 0169 13/03/21 14:23:25 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0169 13/03/21 14:23:29 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (78.1 %). CSM N° 0169 13/03/21 14:23:30 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0169 13/03/21 14:23:33 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (78.2 %). CSM N° 0169 13/03/21 14:23:33 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0169 13/03/21 14:23:36 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (78.9 %). CSM N° 0169 13/03/21 14:23:37 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0169 13/03/21 14:23:40 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (78.9 %). CSM N° 0169 13/03/21 14:23:41 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0169 13/03/21 14:23:44 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (79.3 %). CSM N° 0169 13/03/21 14:23:44 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0169 13/03/21 14:23:52 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.1 %). CSM N° 0169 13/03/21 14:23:53 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 10, 11, CSM N° 0169 13/03/21 14:23:59 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.3 %). CSM N° 0169 13/03/21 14:24:06 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.7 %). CSM N° 0169 13/03/21 14:24:18 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.4 %). CSM N° 0169 13/03/21 14:27:48 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (74.6 W). CSM N° 0169 13/03/21 14:27:51 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:27:55 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.1 %). CSM N° 0169 13/03/21 14:27:56 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:28:01 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.7 W). CSM N° 0169 13/03/21 14:28:21 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (77.3 %). CSM N° 0169 13/03/21 14:28:23 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:28:27 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.5 %). CSM N° 0169 13/03/21 14:28:29 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:28:32 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (82.3 %). CSM N° 0169 13/03/21 14:28:33 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:28:36 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.8 W). CSM N° 0169 13/03/21 14:28:40 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (82.3 %). CSM N° 0169 13/03/21 14:28:41 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:28:44 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (79.0 %). CSM N° 0169 13/03/21 14:28:45 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:28:49 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (79.5 %). CSM N° 0169 13/03/21 14:28:51 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0169 13/03/21 14:28:58 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.5 %). CSM N° 0169 13/03/21 14:29:00 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 11, CSM N° 0169 13/03/21 14:29:03 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.4 %). CSM N° 0169 13/03/21 14:29:06 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 11, CSM N° 0169 13/03/21 14:29:09 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.7 %). CSM N° 0169 13/03/21 14:29:10 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 11, CSM N° 0169 13/03/21 14:29:15 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.9 %). CSM N° 0169 13/03/21 14:29:16 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 11, CSM N° 0169 13/03/21 14:29:21 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.9 %). CSM N° 0169 13/03/21 14:29:22 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 11, 12, CSM N° 0169 13/03/21 14:29:25 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (80.0 W). CSM N° 0169 13/03/21 14:29:27 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0169 13/03/21 14:29:31 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.5 %). CSM N° 0169 13/03/21 14:29:32 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 11, CSM N° 0169 13/03/21 14:29:36 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.4 %). CSM N° 0169 13/03/21 14:29:37 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 11, CSM N° 0169 13/03/21 14:29:42 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (80.9 W). CSM N° 0169 13/03/21 14:29:42 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 11, 12, CSM N° 0169 13/03/21 14:29:46 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.7 %). CSM N° 0169 13/03/21 14:29:46 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 11, CSM N° 0169 13/03/21 14:29:49 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.8 %). CSM N° 0169 13/03/21 14:29:51 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, CSM N° 0169 13/03/21 14:29:56 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.4 %). CSM N° 0169 13/03/21 14:30:04 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.3 %). CSM N° 0169 13/03/21 14:30:08 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.4 %). CSM N° 0169 13/03/21 14:30:12 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (82.1 %). CSM N° 0169 13/03/21 14:30:13 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 11, CSM N° 0169 13/03/21 14:30:17 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (82.1 %). CSM N° 0169 13/03/21 14:30:18 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 11, CSM N° 0169 13/03/21 14:30:21 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.4 %). CSM N° 0169 13/03/21 14:30:22 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, CSM N° 0169 13/03/21 14:30:25 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.4 %). CSM N° 0169 13/03/21 14:30:30 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.7 %). CSM N° 0169 13/03/21 14:30:33 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, CSM N° 0169 13/03/21 14:30:36 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.8 %). CSM N° 0169 13/03/21 14:30:43 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.8 %). CSM N° 0169 13/03/21 14:30:46 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.4 %). CSM N° 0169 13/03/21 14:30:54 2.8.1002 probe number UV135-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (82.7 W). CSM N° 0169 13/03/21 14:30:54 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 11, CSM N° 0169 13/03/21 14:30:59 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.4 %). CSM N° 0169 13/03/21 14:31:20 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (82.8 %). CSM N° 0169 13/03/21 14:31:23 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 11, CSM N° 0169 13/03/21 14:31:26 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.2 %). CSM N° 0169 13/03/21 14:31:30 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.2 %). CSM N° 0169 13/03/21 14:31:34 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.0 %). CSM N° 0169 13/03/21 14:31:38 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.3 %). CSM N° 0169 13/03/21 14:31:42 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.2 %). CSM N° 0169 13/03/21 14:31:46 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.4 %). CSM N° 0169 13/03/21 14:31:50 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.6 %). CSM N° 0169 13/03/21 14:31:56 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.7 %). CSM N° 0169 13/03/21 14:31:57 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, CSM N° 0169 13/03/21 14:32:01 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (82.6 %). CSM N° 0169 13/03/21 14:32:02 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, CSM N° 0169 13/03/21 14:32:04 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.1 %). CSM N° 0169 13/03/21 14:32:05 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, CSM N° 0169 13/03/21 14:32:08 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.4 %). CSM N° 0169 13/03/21 14:32:12 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.8 %). CSM N° 0169 13/03/21 14:32:15 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.1 %). CSM N° 0169 13/03/21 14:32:20 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.2 %). CSM N° 0169 13/03/21 14:32:24 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.3 %). CSM N° 0169 13/03/21 14:32:28 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.5 %). CSM N° 0169 13/03/21 14:32:32 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.8 %). CSM N° 0169 13/03/21 14:32:35 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.6 %). CSM N° 0169 13/03/21 14:33:32 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.6 %). CSM N° 0169 13/03/21 14:33:34 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, CSM N° 0169 13/03/21 14:33:42 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.5 %). CSM N° 0169 13/03/21 14:33:43 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, CSM N° 0169 13/03/21 14:33:46 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.6 %). CSM N° 0169 13/03/21 14:33:52 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.8 %). CSM N° 0169 13/03/21 14:33:59 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.2 %). CSM N° 0169 13/03/21 14:34:03 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.0 %). CSM N° 0169 13/03/21 14:34:06 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.4 %). CSM N° 0169 13/03/21 14:34:11 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.8 %). CSM N° 0169 13/03/21 14:34:15 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.3 %). CSM N° 0169 13/03/21 14:34:19 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.7 %). CSM N° 0169 13/03/21 14:34:23 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.8 %). CSM N° 0169 13/03/21 14:34:24 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, CSM N° 0169 13/03/21 14:34:29 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.7 %). CSM N° 0169 13/03/21 14:34:34 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.0 %). CSM N° 0169 13/03/21 14:34:40 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.8 %). CSM N° 0169 13/03/21 14:34:45 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.3 %). CSM N° 0169 13/03/21 14:34:50 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.7 %). CSM N° 0169 13/03/21 14:34:54 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.8 %). CSM N° 0169 13/03/21 14:34:59 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.1 %). CSM N° 0169 13/03/21 14:35:04 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.7 %). CSM N° 0169 13/03/21 14:35:09 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.6 %). CSM N° 0169 13/03/21 14:35:19 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.1 %). CSM N° 0169 13/03/21 14:35:20 2.8.1006 probe number UV135-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0169 13/03/21 14:35:33 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.4 %). CSM N° 0169 13/03/21 14:35:44 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.7 %). CSM N° 0169 13/03/21 14:45:05 2.8.1102 Temperatura de la sonda excesiva : 59.2 CSM N° 0169 13/03/21 14:45:22 2.8.1102 Temperatura de la sonda excesiva : 14.1 CSM N° 0169 13/03/21 14:45:26 2.8.1102 Temperatura de la sonda excesiva : 14.1 CSM N° 0169 13/03/21 14:48:24 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.4 mm) tiene que estar dentro del intervalo de 21.4 mm a 23.4 mm CSM N° 0169 13/03/21 14:48:41 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (24.5 mm) tiene que estar dentro del intervalo de 21.4 mm a 23.4 mm CSM N° 0169 13/03/21 14:52:06 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (25.6 mm) tiene que estar dentro del intervalo de 23.5 mm a 25.5 mm CSM N° 0169 13/03/21 14:56:14 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (24.8 mm) tiene que estar dentro del intervalo de 25.0 mm a 27.0 mm CSM N° 0169 13/03/21 14:58:56 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.7 %). CSM N° 0169 13/03/21 14:59:08 2.8.1003 probe number UV135-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.8 %). CSM N° 0169 13/03/21 15:03:35 2.8.1102 Temperatura de la sonda excesiva : 10.2 CSM N° 0169 13/03/21 15:08:04 2.8.1102 Temperatura de la sonda excesiva : 10.5 EDAP N° ---- 16/03/21 13:34:52 2.2.1401 No se puede comunicar con el sistema motor EDAP N° ---- 16/03/21 13:35:51 2.2.1401 No se puede comunicar con el sistema motor EDAP N° ---- 16/03/21 13:36:49 2.2.1401 No se puede comunicar con el sistema motor EDAP N° ---- 16/03/21 13:37:18 2.2.1401 No se puede comunicar con el sistema motor EDAP N° ---- 16/03/21 13:37:32 2.2.1401 No se puede comunicar con el sistema motor EDAP N° 0136 16/03/21 13:55:37 2.5.1202 El paciente se ha movido EDAP N° 0136 16/03/21 13:55:41 2.5.1202 El paciente se ha movido EDAP N° 0136 16/03/21 13:55:41 2.5.1202 El paciente se ha movido EDAP N° 0136 16/03/21 14:03:37 2.5.1202 El paciente se ha movido EDAP N° 0136 16/03/21 14:03:38 2.5.1202 El paciente se ha movido EDAP N° 0136 16/03/21 14:03:38 2.5.1202 El paciente se ha movido EDAP N° 0136 16/03/21 14:03:39 2.5.1202 El paciente se ha movido EDAP N° 0136 16/03/21 14:14:15 2.8.1003 probe number UV138-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (64.4 %). EDAP N° 0136 16/03/21 14:15:06 2.8.1006 probe number UV138-----------, Error de potencia en demasiados canales: en canal(es) 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, EDAP N° ---- 16/03/21 15:19:27 2.2.1502 No se puede comunicar con la sonda EDAP N° ---- 16/03/21 15:19:36 2.2.1502 No se puede comunicar con la sonda EDAP N° ---- 16/03/21 15:19:40 2.2.1502 No se puede comunicar con la sonda EDAP N° ---- 16/03/21 15:20:07 2.2.1502 No se puede comunicar con la sonda EDAP N° ---- 16/03/21 15:20:11 2.2.1502 No se puede comunicar con la sonda EDAP N° ---- 16/03/21 15:33:55 2.2.1502 No se puede comunicar con la sonda EDAP N° ---- 16/03/21 15:34:02 2.2.1502 No se puede comunicar con la sonda EDAP N° ---- 16/03/21 15:34:06 2.2.1502 No se puede comunicar con la sonda EDAP N° ---- 16/03/21 15:34:34 2.2.1502 No se puede comunicar con la sonda EDAP N° ---- 16/03/21 15:34:38 2.2.1502 No se puede comunicar con la sonda EDAP N° ---- 16/03/21 15:34:42 2.2.1502 No se puede comunicar con la sonda EDAP N° ---- 16/03/21 15:34:46 2.2.1502 No se puede comunicar con la sonda EDAP N° ---- 16/03/21 15:34:49 2.2.1502 No se puede comunicar con la sonda EDAP N° ---- 16/03/21 15:34:53 2.2.1502 No se puede comunicar con la sonda EDAP N° ---- 16/03/21 15:36:01 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:14:52 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:15:01 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:15:46 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:15:51 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:16:08 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:16:12 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:16:20 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:17:15 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:17:42 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:17:45 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:18:42 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:18:46 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:18:55 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:19:24 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:19:57 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:20:01 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:21:18 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:21:22 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:21:32 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:22:17 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:22:21 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:23:33 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:23:37 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:24:23 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:24:27 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:25:01 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:28:57 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:29:25 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:30:15 2.2.1604 El escáner no puede pasar a modo operativo CSM N° ---- 20/03/21 11:30:43 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:33:53 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:34:08 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 20/03/21 11:35:49 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:37:19 2.2.1604 El escáner no puede pasar a modo operativo CSM N° ---- 20/03/21 11:40:57 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:41:05 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:41:14 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:41:19 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:41:24 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:41:30 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:41:37 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:41:52 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:42:05 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:42:12 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:42:49 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:42:54 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:43:35 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:43:40 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:43:54 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:43:58 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:44:06 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:45:11 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:45:17 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:45:22 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:45:28 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:45:34 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:45:40 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:46:29 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:46:34 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:47:16 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:47:36 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:52:07 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:52:58 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:53:28 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:54:39 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:55:05 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:59:51 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 11:59:59 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 12:00:07 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 12:00:35 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 12:12:26 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 12:12:41 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 12:13:11 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 12:13:38 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 12:13:44 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 12:13:50 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 12:14:13 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 12:23:49 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 12:23:54 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 12:37:45 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 20/03/21 12:38:55 2.2.1502 No se puede comunicar con la sonda EDAP N° 0138 30/03/21 15:31:54 2.5.1202 El paciente se ha movido CSM N° ---- 31/03/21 07:59:46 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 31/03/21 08:06:44 2.2.1604 El escáner no puede pasar a modo operativo CSM N° ---- 31/03/21 08:07:43 2.2.1502 No se puede comunicar con la sonda CSM N° 0170 31/03/21 10:07:35 2.5.1202 El paciente se ha movido CSM N° 0170 31/03/21 10:07:49 2.5.1202 El paciente se ha movido CSM N° 0170 31/03/21 10:07:51 2.5.1202 El paciente se ha movido CSM N° 0170 31/03/21 10:08:43 2.6.1202 El paciente se ha movido CSM N° 0170 31/03/21 10:08:46 2.6.1202 El paciente se ha movido CSM N° ---- 31/03/21 10:34:11 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 31/03/21 10:48:35 2.2.1604 El escáner no puede pasar a modo operativo CSM N° 0171 31/03/21 12:25:11 2.8.1202 El paciente se ha movido CSM N° 0171 31/03/21 12:25:11 2.8.1202 El paciente se ha movido CSM N° ---- 06/04/21 10:19:00 2.3.1901 Código Focalpak incorrecto CSM N° ---- 06/04/21 10:25:21 2.3.1902 Código Focalpak ya utilizado CSM N° ---- 06/04/21 10:30:41 1.3.5101 Tratamiento cancelado en la fase de edición de los datos del tratamiento CSM N° 0173 06/04/21 10:39:28 2.5.1202 El paciente se ha movido CSM N° 0173 06/04/21 10:59:23 1.9.5105 Tratamiento cancelado durante la fase de disparo CSM N° ---- 06/04/21 11:59:27 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 06/04/21 11:59:55 2.2.1604 El escáner no puede pasar a modo operativo CSM N° ---- 06/04/21 12:17:27 2.2.1502 No se puede comunicar con la sonda CSM N° 0174 06/04/21 12:47:17 2.5.1102 Temperatura de la sonda excesiva : 24.6 CSM N° 0174 06/04/21 13:15:49 1.9.5105 Tratamiento cancelado durante la fase de disparo CSM N° 0174 06/04/21 13:38:20 1.9.5104 Tratamiento cancelado en la fase de localización transversal CSM N° ---- 13/04/21 14:53:50 2.2.1423 La posición real está más allá de la posición solicitada en antero-posterior axis CSM N° ---- 13/04/21 14:57:15 2.2.1423 La posición real está más allá de la posición solicitada en antero-posterior axis CSM N° ---- 13/04/21 14:57:16 2.2.1423 La posición real está más allá de la posición solicitada en antero-posterior axis CSM N° 0175 13/04/21 16:08:26 2.5.1202 El paciente se ha movido CSM N° 0175 13/04/21 16:14:26 2.8.1202 El paciente se ha movido CSM N° 0175 13/04/21 16:34:29 1.9.5105 Tratamiento cancelado durante la fase de disparo CSM N° 0176 08/05/21 14:10:57 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (27.6 mm) tiene que estar dentro del intervalo de 17.9 mm a 25.9 mm CSM N° 0176 08/05/21 14:11:14 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (27.5 mm) tiene que estar dentro del intervalo de 17.9 mm a 25.9 mm CSM N° 0176 08/05/21 14:43:03 1.9.5105 Tratamiento cancelado durante la fase de disparo CSM N° 0177 25/05/21 11:48:05 2.7.1202 El paciente se ha movido CSM N° 0177 25/05/21 11:50:52 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (24.5 mm) tiene que estar dentro del intervalo de 25.0 mm a 27.0 mm CSM N° 0177 25/05/21 11:51:06 2.8.1202 El paciente se ha movido CSM N° 0177 25/05/21 11:51:09 2.8.1202 El paciente se ha movido CSM N° 0177 25/05/21 11:51:09 2.8.1202 El paciente se ha movido CSM N° 0177 25/05/21 11:51:10 2.8.1202 El paciente se ha movido CSM N° 0177 25/05/21 11:51:12 2.8.1202 El paciente se ha movido CSM N° 0177 25/05/21 11:51:15 2.8.1202 El paciente se ha movido CSM N° 0177 25/05/21 11:51:19 2.8.1202 El paciente se ha movido CSM N° 0177 25/05/21 11:51:23 2.8.1202 El paciente se ha movido CSM N° 0177 25/05/21 11:51:27 2.8.1202 El paciente se ha movido CSM N° 0177 25/05/21 11:51:33 2.8.1202 El paciente se ha movido CSM N° 0177 25/05/21 11:51:38 2.8.1202 El paciente se ha movido CSM N° 0177 25/05/21 12:01:57 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.2 mm) tiene que estar dentro del intervalo de 20.9 mm a 22.9 mm CSM N° 0177 25/05/21 12:02:17 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.4 mm) tiene que estar dentro del intervalo de 20.9 mm a 22.9 mm CSM N° 0177 25/05/21 12:05:31 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.7 mm) tiene que estar dentro del intervalo de 21.0 mm a 23.0 mm CSM N° 0177 25/05/21 12:05:44 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.8 mm) tiene que estar dentro del intervalo de 21.0 mm a 23.0 mm CSM N° 0177 25/05/21 12:06:26 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (24.5 mm) tiene que estar dentro del intervalo de 25.0 mm a 27.0 mm CSM N° 0177 25/05/21 12:06:31 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (24.6 mm) tiene que estar dentro del intervalo de 25.0 mm a 27.0 mm CSM N° 0177 25/05/21 12:07:23 2.8.1202 El paciente se ha movido CSM N° 0177 25/05/21 12:07:23 2.8.1202 El paciente se ha movido CSM N° 0177 25/05/21 12:11:26 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.2 mm) tiene que estar dentro del intervalo de 23.5 mm a 25.5 mm CSM N° 0177 25/05/21 12:13:32 2.8.1202 El paciente se ha movido CSM N° 0177 25/05/21 12:13:50 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0177 25/05/21 12:14:00 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0177 25/05/21 12:14:03 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0177 25/05/21 12:14:04 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0177 25/05/21 12:15:18 2.5.1202 El paciente se ha movido CSM N° 0177 25/05/21 12:35:42 2.7.1202 El paciente se ha movido CSM N° 0177 25/05/21 12:38:05 1.9.5105 Tratamiento cancelado durante la fase de disparo CSM N° 0178 15/06/21 10:34:04 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:35:54 2.5.1202 El paciente se ha movido CSM N° 0178 15/06/21 10:36:05 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:36:16 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:36:18 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:36:37 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:36:49 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:37:06 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:37:09 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:37:58 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:38:09 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:38:17 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:38:21 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:38:35 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:39:02 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:39:30 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:39:36 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:39:40 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:40:00 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:40:12 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:40:17 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:40:19 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:40:32 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:40:41 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:40:43 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:41:18 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:41:20 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:41:31 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:41:33 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:41:48 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:42:04 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:42:29 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:42:30 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:42:32 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:42:55 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:43:04 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:43:06 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:43:09 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:43:11 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° 0178 15/06/21 10:43:21 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:43:24 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:44:31 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:44:38 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:45:09 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° 0178 15/06/21 10:45:19 2.5.1101 Temperatura de la sonda insuficiente : 0.0 CSM N° 0178 15/06/21 10:45:31 1.5.5102 Tratamiento cancelado en la fase de adquisición CSM N° ---- 15/06/21 10:47:13 2.2.1420 Superado el tiempo de comunicación con antero-posterior axis CSM N° ---- 15/06/21 10:52:37 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 15/06/21 10:53:23 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 15/06/21 10:54:27 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 15/06/21 10:54:28 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 15/06/21 10:54:30 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 15/06/21 10:58:12 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 15/06/21 10:58:25 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° 0179 15/06/21 11:02:45 2.5.1102 Temperatura de la sonda excesiva : 20.6 CSM N° 0179 15/06/21 11:02:51 2.5.1102 Temperatura de la sonda excesiva : 20.4 CSM N° 0179 15/06/21 11:02:55 2.5.1102 Temperatura de la sonda excesiva : 20.3 CSM N° 0179 15/06/21 11:03:05 2.5.1102 Temperatura de la sonda excesiva : 19.9 CSM N° 0179 15/06/21 11:03:07 2.5.1102 Temperatura de la sonda excesiva : 19.8 CSM N° 0179 15/06/21 11:09:29 1.8.1301 Pared rectal no detectada CSM N° 0179 15/06/21 11:09:42 1.8.1301 Pared rectal no detectada CSM N° 0179 15/06/21 11:10:50 2.8.1202 El paciente se ha movido CSM N° 0179 15/06/21 11:14:52 1.8.1301 Pared rectal no detectada CSM N° 0179 15/06/21 11:28:24 1.9.5105 Tratamiento cancelado durante la fase de disparo CSM N° ---- 15/06/21 13:00:38 2.3.1901 Código Focalpak incorrecto CSM N° 0180 15/06/21 13:01:22 2.5.1102 Temperatura de la sonda excesiva : 26.2 CSM N° 0180 15/06/21 13:35:05 2.5.1202 El paciente se ha movido CSM N° 0180 15/06/21 13:35:23 2.5.1202 El paciente se ha movido CSM N° 0180 15/06/21 13:41:01 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (69.3 W). CSM N° 0180 15/06/21 13:41:07 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 14, 15, CSM N° 0180 15/06/21 13:41:09 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (70.6 W). CSM N° 0180 15/06/21 13:41:15 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 14, 15, CSM N° 0180 15/06/21 13:41:18 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (73.3 W). CSM N° 0180 15/06/21 13:41:22 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 14, 15, CSM N° 0180 15/06/21 13:41:24 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (75.8 W). CSM N° 0180 15/06/21 13:41:27 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 14, 15, CSM N° 0180 15/06/21 13:41:30 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (76.8 W). CSM N° 0180 15/06/21 13:41:33 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0180 15/06/21 13:42:08 2.5.1202 El paciente se ha movido CSM N° 0180 15/06/21 13:46:11 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.1 %). CSM N° 0180 15/06/21 13:46:15 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 7, 8, 12, 14, 15, 16, CSM N° 0180 15/06/21 13:46:26 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.0 %). CSM N° 0180 15/06/21 13:46:27 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 13, 14, 15, 16, CSM N° 0180 15/06/21 13:46:37 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.2 %). CSM N° 0180 15/06/21 13:46:39 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 13, 14, 15, 16, CSM N° 0180 15/06/21 13:46:49 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.3 %). CSM N° 0180 15/06/21 13:46:57 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 13, 14, 15, 16, CSM N° 0180 15/06/21 13:47:05 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.0 %). CSM N° 0180 15/06/21 13:47:06 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 13, 14, 15, 16, CSM N° 0180 15/06/21 13:47:11 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.0 %). CSM N° 0180 15/06/21 13:47:12 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 13, 14, 15, 16, CSM N° 0180 15/06/21 13:47:17 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.7 %). CSM N° 0180 15/06/21 13:47:18 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 13, 14, 15, 16, CSM N° 0180 15/06/21 13:47:24 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.1 %). CSM N° 0180 15/06/21 13:47:28 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, 7, 8, 10, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:47:33 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.7 %). CSM N° 0180 15/06/21 13:47:35 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 5, 6, 7, 8, 10, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:47:41 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.5 %). CSM N° 0180 15/06/21 13:47:42 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 5, 6, 7, 8, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:47:47 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.3 %). CSM N° 0180 15/06/21 13:47:48 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 5, 6, 7, 8, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:47:53 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.1 %). CSM N° 0180 15/06/21 13:47:54 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 5, 6, 8, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:47:59 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.9 %). CSM N° 0180 15/06/21 13:48:01 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 5, 6, 7, 8, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:48:06 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.1 %). CSM N° 0180 15/06/21 13:48:07 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 6, 8, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:48:12 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.2 %). CSM N° 0180 15/06/21 13:48:17 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 5, 6, 8, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:48:22 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.6 %). CSM N° 0180 15/06/21 13:48:26 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 5, 6, 7, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:48:31 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.8 %). CSM N° 0180 15/06/21 13:48:32 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:48:37 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.7 %). CSM N° 0180 15/06/21 13:48:39 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 5, 6, 7, 8, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:48:47 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.7 %). CSM N° 0180 15/06/21 13:48:48 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, 7, 8, 10, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:48:52 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.4 %). CSM N° 0180 15/06/21 13:48:56 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 5, 6, 7, 8, 10, 13, 14, 15, CSM N° 0180 15/06/21 13:49:00 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.4 %). CSM N° 0180 15/06/21 13:49:01 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 5, 6, 7, 8, 10, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:49:15 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.6 %). CSM N° 0180 15/06/21 13:49:17 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 5, 6, 7, 8, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:49:24 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.0 %). CSM N° 0180 15/06/21 13:49:25 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 5, 6, 7, 8, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:49:29 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (82.7 W). CSM N° 0180 15/06/21 13:49:30 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 13, 14, 15, CSM N° 0180 15/06/21 13:49:34 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (81.7 W). CSM N° 0180 15/06/21 13:49:35 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 13, 14, 15, CSM N° 0180 15/06/21 13:49:39 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.6 %). CSM N° 0180 15/06/21 13:49:40 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:49:46 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.7 %). CSM N° 0180 15/06/21 13:49:47 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:49:58 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.9 %). CSM N° 0180 15/06/21 13:49:59 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:50:10 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.3 mm) tiene que estar dentro del intervalo de 20.4 mm a 22.4 mm CSM N° 0180 15/06/21 13:50:21 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.6 %). CSM N° 0180 15/06/21 13:50:22 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 10, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:50:29 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.8 %). CSM N° 0180 15/06/21 13:50:30 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 10, 13, 14, 15, CSM N° 0180 15/06/21 13:51:01 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.8 %). CSM N° 0180 15/06/21 13:51:02 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 13, 14, 15, CSM N° 0180 15/06/21 13:52:00 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (21.0 mm) tiene que estar dentro del intervalo de 21.4 mm a 29.4 mm CSM N° 0180 15/06/21 13:52:46 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (21.3 mm) tiene que estar dentro del intervalo de 21.4 mm a 29.4 mm CSM N° 0180 15/06/21 13:52:53 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (21.0 mm) tiene que estar dentro del intervalo de 21.4 mm a 29.4 mm CSM N° 0180 15/06/21 13:53:37 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.0 %). CSM N° 0180 15/06/21 13:53:43 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:53:49 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.3 %). CSM N° 0180 15/06/21 13:53:49 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:53:53 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.7 %). CSM N° 0180 15/06/21 13:53:54 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 5, 6, 7, 8, 9, 10, 13, 14, 15, 16, CSM N° 0180 15/06/21 13:53:59 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.7 %). CSM N° 0180 15/06/21 13:54:00 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 10, 13, 14, 15, CSM N° 0180 15/06/21 13:54:41 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.3 mm) tiene que estar dentro del intervalo de 20.0 mm a 22.0 mm CSM N° 0180 15/06/21 13:54:50 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.7 mm) tiene que estar dentro del intervalo de 20.5 mm a 22.5 mm CSM N° 0180 15/06/21 13:55:20 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.9 %). CSM N° 0180 15/06/21 13:55:22 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 10, 13, 14, 15, CSM N° 0180 15/06/21 13:56:00 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.1 %). CSM N° 0180 15/06/21 13:56:02 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 13:56:09 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.2 %). CSM N° 0180 15/06/21 13:56:10 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 13:56:23 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.6 mm) tiene que estar dentro del intervalo de 24.7 mm a 26.7 mm CSM N° 0180 15/06/21 13:56:40 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.0 %). CSM N° 0180 15/06/21 13:56:42 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, 15, CSM N° 0180 15/06/21 13:56:45 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.9 %). CSM N° 0180 15/06/21 13:56:46 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 13:56:56 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.2 %). CSM N° 0180 15/06/21 13:56:57 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 13, 14, 15, CSM N° 0180 15/06/21 13:57:03 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.6 %). CSM N° 0180 15/06/21 13:57:04 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 13, 14, CSM N° 0180 15/06/21 13:57:12 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.1 %). CSM N° 0180 15/06/21 13:57:15 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 13:57:19 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.2 %). CSM N° 0180 15/06/21 13:57:20 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, 15, CSM N° 0180 15/06/21 13:57:25 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.7 %). CSM N° 0180 15/06/21 13:57:26 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 13:57:31 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.9 %). CSM N° 0180 15/06/21 13:57:31 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 13:57:43 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.8 mm) tiene que estar dentro del intervalo de 20.2 mm a 22.2 mm CSM N° 0180 15/06/21 13:57:56 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.6 %). CSM N° 0180 15/06/21 13:57:59 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 13:58:30 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.7 %). CSM N° 0180 15/06/21 13:58:32 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 13:58:42 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.4 %). CSM N° 0180 15/06/21 13:58:43 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 13:58:48 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.1 %). CSM N° 0180 15/06/21 13:58:49 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 13:58:56 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.2 %). CSM N° 0180 15/06/21 13:58:58 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, 15, CSM N° 0180 15/06/21 13:59:02 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.4 %). CSM N° 0180 15/06/21 13:59:03 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, 15, CSM N° 0180 15/06/21 13:59:16 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.1 %). CSM N° 0180 15/06/21 13:59:17 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, 15, CSM N° 0180 15/06/21 13:59:21 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.7 %). CSM N° 0180 15/06/21 13:59:22 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 13:59:32 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.6 %). CSM N° 0180 15/06/21 13:59:33 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, 15, CSM N° 0180 15/06/21 14:02:15 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.9 %). CSM N° 0180 15/06/21 14:02:17 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, 15, 16, CSM N° 0180 15/06/21 14:02:20 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.5 W). CSM N° 0180 15/06/21 14:02:21 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 13, 14, 15, CSM N° 0180 15/06/21 14:02:30 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.3 %). CSM N° 0180 15/06/21 14:02:31 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, 15, CSM N° 0180 15/06/21 14:02:38 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.2 %). CSM N° 0180 15/06/21 14:02:39 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, 15, CSM N° 0180 15/06/21 14:02:45 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.7 %). CSM N° 0180 15/06/21 14:02:47 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:04:16 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.6 %). CSM N° 0180 15/06/21 14:04:19 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:04:26 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.2 %). CSM N° 0180 15/06/21 14:04:27 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, 15, CSM N° 0180 15/06/21 14:04:35 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.9 %). CSM N° 0180 15/06/21 14:04:36 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:04:48 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.8 %). CSM N° 0180 15/06/21 14:04:49 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:04:53 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.6 %). CSM N° 0180 15/06/21 14:04:55 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:05:00 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.2 %). CSM N° 0180 15/06/21 14:05:00 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 14:05:09 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.3 %). CSM N° 0180 15/06/21 14:05:10 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:05:15 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.9 %). CSM N° 0180 15/06/21 14:05:16 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 14:05:27 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.9 %). CSM N° 0180 15/06/21 14:05:28 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0180 15/06/21 14:05:34 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.3 %). CSM N° 0180 15/06/21 14:05:35 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 14:05:43 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.5 %). CSM N° 0180 15/06/21 14:05:44 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 14:05:56 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.9 %). CSM N° 0180 15/06/21 14:05:57 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 10, 13, 14, CSM N° 0180 15/06/21 14:06:36 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.9 %). CSM N° 0180 15/06/21 14:06:40 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:06:49 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.6 %). CSM N° 0180 15/06/21 14:06:51 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, 15, CSM N° 0180 15/06/21 14:06:58 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.6 %). CSM N° 0180 15/06/21 14:06:59 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:07:04 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.8 %). CSM N° 0180 15/06/21 14:07:05 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 13, 15, CSM N° 0180 15/06/21 14:07:15 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.6 %). CSM N° 0180 15/06/21 14:07:16 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, CSM N° 0180 15/06/21 14:07:20 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (84.9 W). CSM N° 0180 15/06/21 14:07:21 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0180 15/06/21 14:07:28 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.6 %). CSM N° 0180 15/06/21 14:07:29 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 15, CSM N° 0180 15/06/21 14:08:38 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (78.6 %). CSM N° 0180 15/06/21 14:08:41 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 15, CSM N° 0180 15/06/21 14:08:52 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.2 %). CSM N° 0180 15/06/21 14:08:53 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 15, 16, CSM N° 0180 15/06/21 14:09:02 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.0 mm) tiene que estar dentro del intervalo de 20.7 mm a 22.7 mm CSM N° 0180 15/06/21 14:09:08 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.2 %). CSM N° 0180 15/06/21 14:09:09 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0180 15/06/21 14:09:14 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.8 %). CSM N° 0180 15/06/21 14:09:15 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:09:24 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.0 %). CSM N° 0180 15/06/21 14:09:25 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 14:09:35 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.2 %). CSM N° 0180 15/06/21 14:09:36 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 14:09:45 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.5 %). CSM N° 0180 15/06/21 14:09:46 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:10:41 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.3 %). CSM N° 0180 15/06/21 14:10:42 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 14:10:46 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.1 %). CSM N° 0180 15/06/21 14:10:48 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:10:50 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.2 %). CSM N° 0180 15/06/21 14:10:52 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:10:55 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.8 %). CSM N° 0180 15/06/21 14:10:56 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0180 15/06/21 14:10:59 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (79.7 W). CSM N° 0180 15/06/21 14:11:00 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0180 15/06/21 14:11:11 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.3 %). CSM N° 0180 15/06/21 14:11:13 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0180 15/06/21 14:11:16 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.6 %). CSM N° 0180 15/06/21 14:11:17 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 13, CSM N° 0180 15/06/21 14:11:22 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.8 %). CSM N° 0180 15/06/21 14:11:23 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 13, CSM N° 0180 15/06/21 14:11:26 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.9 %). CSM N° 0180 15/06/21 14:11:28 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 13, CSM N° 0180 15/06/21 14:11:37 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.9 %). CSM N° 0180 15/06/21 14:11:38 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 13, CSM N° 0180 15/06/21 14:11:41 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.4 %). CSM N° 0180 15/06/21 14:11:42 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 13, CSM N° 0180 15/06/21 14:11:46 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.3 %). CSM N° 0180 15/06/21 14:11:47 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:11:52 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.4 %). CSM N° 0180 15/06/21 14:11:53 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:12:49 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.8 %). CSM N° 0180 15/06/21 14:12:51 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:12:55 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.3 %). CSM N° 0180 15/06/21 14:12:58 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:13:00 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.3 %). CSM N° 0180 15/06/21 14:13:02 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, CSM N° 0180 15/06/21 14:13:06 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.9 %). CSM N° 0180 15/06/21 14:13:07 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 14:13:18 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.6 %). CSM N° 0180 15/06/21 14:13:19 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, CSM N° 0180 15/06/21 14:13:23 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (82.3 %). CSM N° 0180 15/06/21 14:13:24 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 13, CSM N° 0180 15/06/21 14:14:15 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.5 %). CSM N° 0180 15/06/21 14:14:17 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 13, CSM N° 0180 15/06/21 14:14:27 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.9 %). CSM N° 0180 15/06/21 14:14:29 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, 15, 16, CSM N° 0180 15/06/21 14:14:36 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.3 W). CSM N° 0180 15/06/21 14:14:38 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:14:41 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.7 %). CSM N° 0180 15/06/21 14:14:42 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 14:14:46 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.0 %). CSM N° 0180 15/06/21 14:14:46 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:14:52 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.3 %). CSM N° 0180 15/06/21 14:14:52 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:15:53 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.4 %). CSM N° 0180 15/06/21 14:15:55 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:15:59 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.6 %). CSM N° 0180 15/06/21 14:16:06 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:16:09 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.8 %). CSM N° 0180 15/06/21 14:16:11 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, CSM N° 0180 15/06/21 14:16:14 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (87.5 W). CSM N° 0180 15/06/21 14:16:16 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 14:16:27 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.9 %). CSM N° 0180 15/06/21 14:16:28 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, CSM N° 0180 15/06/21 14:16:32 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.0 %). CSM N° 0180 15/06/21 14:16:32 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 13, CSM N° 0180 15/06/21 14:16:36 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (81.9 W). CSM N° 0180 15/06/21 14:16:37 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 13, CSM N° 0180 15/06/21 14:16:45 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.5 %). CSM N° 0180 15/06/21 14:16:46 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, CSM N° 0180 15/06/21 14:16:56 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.6 W). CSM N° 0180 15/06/21 14:16:57 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 14:17:00 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.0 %). CSM N° 0180 15/06/21 14:17:02 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 13, CSM N° 0180 15/06/21 14:17:06 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.9 %). CSM N° 0180 15/06/21 14:17:07 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 14:17:12 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.9 %). CSM N° 0180 15/06/21 14:17:13 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:17:21 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.6 %). CSM N° 0180 15/06/21 14:17:22 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:18:01 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.9 %). CSM N° 0180 15/06/21 14:18:02 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:18:11 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.8 %). CSM N° 0180 15/06/21 14:18:13 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:18:16 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.9 %). CSM N° 0180 15/06/21 14:18:17 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 13, 14, CSM N° 0180 15/06/21 14:18:21 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.9 %). CSM N° 0180 15/06/21 14:18:22 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 14:18:26 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.3 W). CSM N° 0180 15/06/21 14:18:27 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 14:18:37 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.8 %). CSM N° 0180 15/06/21 14:18:38 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 13, CSM N° 0180 15/06/21 14:18:42 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (81.1 W). CSM N° 0180 15/06/21 14:18:43 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 13, CSM N° 0180 15/06/21 14:24:03 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.0 %). CSM N° 0180 15/06/21 14:24:05 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 7, 8, 9, 10, 12, 13, 14, 15, 16, CSM N° 0180 15/06/21 14:24:11 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.7 %). CSM N° 0180 15/06/21 14:24:14 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 7, 8, 9, 10, 12, 13, 14, CSM N° 0180 15/06/21 14:24:21 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.6 %). CSM N° 0180 15/06/21 14:24:22 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 13, 14, CSM N° 0180 15/06/21 14:24:29 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.9 %). CSM N° 0180 15/06/21 14:24:31 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:24:37 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.3 %). CSM N° 0180 15/06/21 14:24:40 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:24:43 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.5 %). CSM N° 0180 15/06/21 14:24:44 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:24:47 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.7 %). CSM N° 0180 15/06/21 14:24:48 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:24:51 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.5 %). CSM N° 0180 15/06/21 14:24:53 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:24:56 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.6 %). CSM N° 0180 15/06/21 14:24:57 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:25:00 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.6 %). CSM N° 0180 15/06/21 14:25:01 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:25:04 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.2 %). CSM N° 0180 15/06/21 14:25:06 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:25:09 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.4 %). CSM N° 0180 15/06/21 14:25:10 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:25:14 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.2 %). CSM N° 0180 15/06/21 14:25:15 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:25:18 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (82.5 %). CSM N° 0180 15/06/21 14:25:19 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:25:22 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (82.6 %). CSM N° 0180 15/06/21 14:25:23 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0180 15/06/21 14:25:26 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.8 %). CSM N° 0180 15/06/21 14:25:28 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0180 15/06/21 14:25:31 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.3 %). CSM N° 0180 15/06/21 14:25:34 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0180 15/06/21 14:25:43 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.0 %). CSM N° 0180 15/06/21 14:25:45 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0180 15/06/21 14:25:50 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.2 %). CSM N° 0180 15/06/21 14:25:51 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0180 15/06/21 14:25:57 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.0 %). CSM N° 0180 15/06/21 14:25:58 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0180 15/06/21 14:26:02 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.1 %). CSM N° 0180 15/06/21 14:26:03 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0180 15/06/21 14:26:07 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.5 %). CSM N° 0180 15/06/21 14:26:08 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 9, 10, 12, 13, CSM N° 0180 15/06/21 14:26:14 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.4 %). CSM N° 0180 15/06/21 14:26:16 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 9, 11, 12, 13, CSM N° 0180 15/06/21 14:26:22 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.6 %). CSM N° 0180 15/06/21 14:26:29 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.5 %). CSM N° 0180 15/06/21 14:26:30 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0180 15/06/21 14:26:37 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.3 %). CSM N° 0180 15/06/21 14:26:38 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0180 15/06/21 14:26:44 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.1 %). CSM N° 0180 15/06/21 14:26:46 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0180 15/06/21 14:26:50 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.1 %). CSM N° 0180 15/06/21 14:26:59 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0180 15/06/21 14:27:02 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.2 W). CSM N° 0180 15/06/21 14:27:04 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0180 15/06/21 14:27:09 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.2 %). CSM N° 0180 15/06/21 14:27:12 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0180 15/06/21 14:27:18 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.1 %). CSM N° 0180 15/06/21 14:27:21 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0180 15/06/21 14:27:28 2.8.1002 probe number UV140-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.5 W). CSM N° 0180 15/06/21 14:27:30 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0180 15/06/21 14:27:34 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.1 %). CSM N° 0180 15/06/21 14:27:36 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 8, 9, 10, 13, CSM N° 0180 15/06/21 14:27:40 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.5 %). CSM N° 0180 15/06/21 14:27:41 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 8, 10, 11, 13, CSM N° 0180 15/06/21 14:27:53 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.9 %). CSM N° 0180 15/06/21 14:27:54 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 13, CSM N° 0180 15/06/21 14:28:01 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.9 %). CSM N° 0180 15/06/21 14:28:03 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 13, CSM N° 0180 15/06/21 14:30:09 2.8.1003 probe number UV140-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.9 %). CSM N° 0180 15/06/21 14:30:15 2.8.1006 probe number UV140-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 13, EDAP N° ---- 18/06/21 11:15:01 3.2.1503 La sonda no ha sido calibrada en esta máquina EDAP N° 0140 18/06/21 12:02:03 2.5.1202 El paciente se ha movido EDAP N° 0140 18/06/21 12:05:41 2.8.1003 probe number ----------------, Potencia medida insuficiente: menos del 90 % de lo solicitado (70.8 %). EDAP N° 0140 18/06/21 12:05:48 2.8.1006 probe number ----------------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, EDAP N° 0140 18/06/21 12:05:51 2.8.1003 probe number ----------------, Potencia medida insuficiente: menos del 90 % de lo solicitado (76.8 %). EDAP N° 0140 18/06/21 12:06:08 2.8.1006 probe number ----------------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, EDAP N° 0140 18/06/21 12:06:10 2.8.1003 probe number ----------------, Potencia medida insuficiente: menos del 90 % de lo solicitado (77.1 %). EDAP N° 0140 18/06/21 12:06:14 2.8.1006 probe number ----------------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, EDAP N° 0140 18/06/21 12:06:18 2.8.1014 probe number ----------------, Potencia medida insuficiente: más de 1,6W por debajo de lo solicitado (-3.3 W). EDAP N° 0140 18/06/21 12:06:26 2.8.1006 probe number ----------------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, EDAP N° 0140 18/06/21 12:06:31 2.8.1003 probe number ----------------, Potencia medida insuficiente: menos del 90 % de lo solicitado (78.6 %). EDAP N° 0140 18/06/21 12:06:33 2.8.1006 probe number ----------------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, EDAP N° 0140 18/06/21 12:06:37 2.8.1003 probe number ----------------, Potencia medida insuficiente: menos del 90 % de lo solicitado (77.5 %). EDAP N° 0140 18/06/21 12:06:39 2.8.1006 probe number ----------------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, EDAP N° 0140 18/06/21 12:06:43 2.8.1003 probe number ----------------, Potencia medida insuficiente: menos del 90 % de lo solicitado (78.3 %). EDAP N° 0140 18/06/21 12:06:44 2.8.1006 probe number ----------------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, EDAP N° 0140 18/06/21 12:06:48 2.8.1003 probe number ----------------, Potencia medida insuficiente: menos del 90 % de lo solicitado (78.9 %). EDAP N° 0140 18/06/21 12:06:49 2.8.1006 probe number ----------------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, EDAP N° 0140 18/06/21 12:06:53 2.8.1003 probe number ----------------, Potencia medida insuficiente: menos del 90 % de lo solicitado (80.1 %). EDAP N° 0140 18/06/21 12:06:56 2.8.1006 probe number ----------------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, EDAP N° 0140 18/06/21 12:07:00 2.8.1003 probe number ----------------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.2 %). EDAP N° 0140 18/06/21 12:07:01 2.8.1006 probe number ----------------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, 7, 8, 9, 11, EDAP N° 0140 18/06/21 12:07:05 2.8.1003 probe number ----------------, Potencia medida insuficiente: menos del 90 % de lo solicitado (82.4 %). EDAP N° 0140 18/06/21 12:07:06 2.8.1006 probe number ----------------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, 7, 8, 10, EDAP N° 0140 18/06/21 12:07:09 2.8.1003 probe number ----------------, Potencia medida insuficiente: menos del 90 % de lo solicitado (82.8 %). EDAP N° 0140 18/06/21 12:07:11 2.8.1006 probe number ----------------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, EDAP N° 0140 18/06/21 12:10:09 2.8.1003 probe number ----------------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.4 %). EDAP N° 0140 18/06/21 12:10:13 2.8.1006 probe number ----------------, Error de potencia en demasiados canales: en canal(es) 4, 6, 7, 8, 12, EDAP N° 0140 18/06/21 12:16:53 2.8.1015 probe number ----------------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.6 W). EDAP N° 0140 18/06/21 12:19:18 2.8.1015 probe number ----------------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.7 W). EDAP N° 0140 18/06/21 12:19:37 2.8.1015 probe number ----------------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.7 W). EDAP N° 0140 18/06/21 12:19:52 2.8.1015 probe number ----------------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.8 W). EDAP N° 0140 18/06/21 12:20:06 2.8.1015 probe number ----------------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.8 W). EDAP N° 0140 18/06/21 12:20:19 2.8.1015 probe number ----------------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.7 W). EDAP N° 0140 18/06/21 12:20:47 2.8.1015 probe number ----------------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.6 W). EDAP N° 0140 18/06/21 12:23:17 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (111.8%). EDAP N° 0140 18/06/21 12:23:23 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (112.2%). EDAP N° 0140 18/06/21 12:24:03 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.2%). EDAP N° 0140 18/06/21 12:24:09 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.0%). EDAP N° 0140 18/06/21 12:24:31 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (112.7%). EDAP N° 0140 18/06/21 12:24:36 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (113.3%). EDAP N° 0140 18/06/21 12:24:54 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.2%). EDAP N° 0140 18/06/21 12:25:01 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). EDAP N° 0140 18/06/21 12:25:08 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.5%). EDAP N° 0140 18/06/21 12:25:13 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.3%). EDAP N° 0140 18/06/21 12:25:19 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.3%). EDAP N° 0140 18/06/21 12:25:25 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). EDAP N° 0140 18/06/21 12:25:30 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). EDAP N° 0140 18/06/21 12:25:36 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.2%). EDAP N° 0140 18/06/21 12:26:03 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.3%). EDAP N° 0140 18/06/21 12:26:10 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (111.4%). EDAP N° 0140 18/06/21 12:26:26 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). EDAP N° 0140 18/06/21 12:26:37 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.7%). EDAP N° 0140 18/06/21 12:26:46 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.7%). EDAP N° 0140 18/06/21 12:27:11 2.8.1015 probe number ----------------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.7 W). EDAP N° 0140 18/06/21 12:27:18 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.3%). EDAP N° 0140 18/06/21 12:27:25 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.4%). EDAP N° 0140 18/06/21 12:27:59 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.9%). EDAP N° 0140 18/06/21 12:28:11 2.8.1015 probe number ----------------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.7 W). EDAP N° 0140 18/06/21 12:28:23 2.8.1015 probe number ----------------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.7 W). EDAP N° 0140 18/06/21 12:28:42 2.8.1015 probe number ----------------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.8 W). EDAP N° 0140 18/06/21 12:29:13 2.8.1015 probe number ----------------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.7 W). EDAP N° 0140 18/06/21 12:29:26 2.8.1015 probe number ----------------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.9 W). EDAP N° 0140 18/06/21 12:29:41 2.8.1015 probe number ----------------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.8 W). EDAP N° 0140 18/06/21 12:29:59 2.8.1015 probe number ----------------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.7 W). EDAP N° 0140 18/06/21 12:30:09 2.8.1015 probe number ----------------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.7 W). EDAP N° 0140 18/06/21 12:31:49 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (112.2%). EDAP N° 0140 18/06/21 12:32:00 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (112.2%). EDAP N° 0140 18/06/21 12:32:11 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (111.9%). EDAP N° 0140 18/06/21 12:32:15 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (113.0%). EDAP N° 0140 18/06/21 12:32:29 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.4%). EDAP N° 0140 18/06/21 12:32:36 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (111.1%). EDAP N° 0140 18/06/21 12:32:44 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.7%). EDAP N° 0140 18/06/21 12:32:51 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (111.7%). EDAP N° 0140 18/06/21 12:32:57 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (112.1%). EDAP N° 0140 18/06/21 12:33:03 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (111.4%). EDAP N° 0140 18/06/21 12:33:10 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (111.2%). EDAP N° 0140 18/06/21 12:33:14 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.0%). EDAP N° 0140 18/06/21 12:33:18 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (110.2%). EDAP N° 0140 18/06/21 12:33:22 2.8.1004 probe number ----------------, Potencia medida excesiva: más del 110 % de lo solicitado (115.5%). EDAP N° 0141 18/06/21 13:22:55 2.8.1003 probe number ----------------, Potencia medida insuficiente: menos del 90 % de lo solicitado (79.5 %). EDAP N° 0141 18/06/21 13:48:40 2.8.1006 probe number ----------------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, EDAP N° 0142 23/07/21 22:58:05 2.5.1202 El paciente se ha movido EDAP N° 0142 23/07/21 23:02:49 1.5.5102 Tratamiento cancelado en la fase de adquisición EDAP N° 0143 23/07/21 23:10:53 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.3 %). EDAP N° 0143 23/07/21 23:11:20 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, EDAP N° 0143 23/07/21 23:11:23 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.7 %). EDAP N° 0143 23/07/21 23:11:27 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, EDAP N° 0143 23/07/21 23:11:30 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.2 %). EDAP N° 0143 23/07/21 23:11:33 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, EDAP N° 0143 23/07/21 23:11:36 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.8 %). EDAP N° 0143 23/07/21 23:11:53 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, EDAP N° 0143 23/07/21 23:11:57 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.9 %). EDAP N° 0143 23/07/21 23:12:21 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, EDAP N° 0143 23/07/21 23:12:24 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.3 %). EDAP N° 0143 23/07/21 23:13:34 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, EDAP N° 0143 23/07/21 23:13:37 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.2 %). EDAP N° 0143 23/07/21 23:13:38 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, EDAP N° 0143 23/07/21 23:13:41 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.4 %). EDAP N° 0143 23/07/21 23:13:43 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, EDAP N° 0143 23/07/21 23:13:46 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.7 %). EDAP N° 0143 23/07/21 23:15:06 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, EDAP N° 0143 23/07/21 23:15:09 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.6 %). EDAP N° 0143 23/07/21 23:15:12 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, EDAP N° 0143 23/07/21 23:15:16 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.3 %). EDAP N° 0143 23/07/21 23:15:17 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, EDAP N° 0143 23/07/21 23:15:21 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.8 %). EDAP N° 0143 23/07/21 23:15:39 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, EDAP N° 0143 23/07/21 23:16:06 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.4 %). EDAP N° 0143 23/07/21 23:16:10 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, EDAP N° 0143 23/07/21 23:17:49 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.5 %). EDAP N° 0143 23/07/21 23:18:00 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, EDAP N° 0143 23/07/21 23:18:04 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.9 %). EDAP N° 0143 23/07/21 23:18:05 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, EDAP N° 0143 23/07/21 23:18:10 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.8 %). EDAP N° 0143 23/07/21 23:18:12 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, EDAP N° 0143 23/07/21 23:18:15 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.9 %). EDAP N° 0143 23/07/21 23:18:19 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, EDAP N° 0143 23/07/21 23:18:23 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.6 %). EDAP N° 0143 23/07/21 23:18:25 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, EDAP N° 0143 23/07/21 23:18:29 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.9 %). EDAP N° 0143 23/07/21 23:19:11 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, EDAP N° 0143 23/07/21 23:20:51 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.6 %). EDAP N° 0143 23/07/21 23:20:53 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, EDAP N° 0143 23/07/21 23:26:21 1.9.5104 Tratamiento cancelado en la fase de localización transversal EDAP N° 0144 24/07/21 00:09:37 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.7 %). EDAP N° 0144 24/07/21 00:09:49 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (90.0 %). EDAP N° 0144 24/07/21 00:10:02 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.5 %). EDAP N° 0144 24/07/21 00:10:33 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.4 %). EDAP N° 0144 24/07/21 00:10:40 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.1 %). EDAP N° 0144 24/07/21 00:10:44 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.1 %). EDAP N° 0144 24/07/21 00:10:48 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.7 %). EDAP N° 0144 24/07/21 00:11:33 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.6 %). EDAP N° 0144 24/07/21 00:14:31 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.2%). EDAP N° 0144 24/07/21 00:14:35 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:16:38 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.6 %). EDAP N° 0144 24/07/21 00:16:50 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.8 %). EDAP N° 0144 24/07/21 00:16:59 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.5 %). EDAP N° 0144 24/07/21 00:17:01 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, EDAP N° 0144 24/07/21 00:17:05 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.8 %). EDAP N° 0144 24/07/21 00:17:24 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 7, EDAP N° 0144 24/07/21 00:17:28 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.4 %). EDAP N° 0144 24/07/21 00:17:30 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, EDAP N° 0144 24/07/21 00:22:56 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.7 %). EDAP N° 0144 24/07/21 00:25:31 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.0%). EDAP N° 0144 24/07/21 00:26:12 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:26:26 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.4%). EDAP N° 0144 24/07/21 00:26:28 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:26:34 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.3%). EDAP N° 0144 24/07/21 00:26:39 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:27:58 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.2%). EDAP N° 0144 24/07/21 00:28:05 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:28:11 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.5%). EDAP N° 0144 24/07/21 00:28:13 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:28:16 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.8%). EDAP N° 0144 24/07/21 00:28:20 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:49:39 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (82.6 %). EDAP N° 0144 24/07/21 00:49:44 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, EDAP N° 0144 24/07/21 00:51:23 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.1%). EDAP N° 0144 24/07/21 00:51:27 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 7, 8, EDAP N° 0144 24/07/21 00:51:41 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.3%). EDAP N° 0144 24/07/21 00:52:22 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:52:30 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.2%). EDAP N° 0144 24/07/21 00:52:33 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:52:36 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). EDAP N° 0144 24/07/21 00:52:43 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:52:46 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.4%). EDAP N° 0144 24/07/21 00:52:48 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:52:50 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.4%). EDAP N° 0144 24/07/21 00:52:51 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:52:54 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.3%). EDAP N° 0144 24/07/21 00:52:55 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:52:59 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.9%). EDAP N° 0144 24/07/21 00:53:00 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:53:03 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.3%). EDAP N° 0144 24/07/21 00:53:05 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:53:08 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (116.9 W). EDAP N° 0144 24/07/21 00:53:23 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:53:27 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (114.2 W). EDAP N° 0144 24/07/21 00:53:31 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:53:34 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.5%). EDAP N° 0144 24/07/21 00:53:37 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:53:41 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.3%). EDAP N° 0144 24/07/21 00:53:42 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:53:45 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (115.1%). EDAP N° 0144 24/07/21 00:53:51 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:53:54 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.3%). EDAP N° 0144 24/07/21 00:56:01 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:56:09 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.3%). EDAP N° 0144 24/07/21 00:56:12 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:56:15 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.9%). EDAP N° 0144 24/07/21 00:56:17 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:56:20 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.0%). EDAP N° 0144 24/07/21 00:56:22 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:56:25 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.9%). EDAP N° 0144 24/07/21 00:56:27 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:56:30 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.5%). EDAP N° 0144 24/07/21 00:56:31 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:56:34 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.0%). EDAP N° 0144 24/07/21 00:56:35 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:56:38 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (116.5%). EDAP N° 0144 24/07/21 00:56:40 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:56:43 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (117.0%). EDAP N° 0144 24/07/21 00:56:46 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:56:49 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.3%). EDAP N° 0144 24/07/21 00:56:50 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:56:53 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.1%). EDAP N° 0144 24/07/21 00:56:55 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, EDAP N° 0144 24/07/21 00:56:57 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.8%). EDAP N° 0144 24/07/21 00:56:59 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:57:02 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (115.2%). EDAP N° 0144 24/07/21 00:57:03 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:57:06 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (115.4%). EDAP N° 0144 24/07/21 00:57:07 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:57:10 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (115.3%). EDAP N° 0144 24/07/21 00:57:12 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:57:14 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (115.7%). EDAP N° 0144 24/07/21 00:57:16 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:57:19 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (116.0%). EDAP N° 0144 24/07/21 00:57:20 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:57:23 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (116.3%). EDAP N° 0144 24/07/21 00:57:24 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:57:27 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (116.6%). EDAP N° 0144 24/07/21 00:57:28 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:57:31 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (116.9%). EDAP N° 0144 24/07/21 00:57:33 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, EDAP N° 0144 24/07/21 00:57:36 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (119.6 W). EDAP N° 0144 24/07/21 00:58:15 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, CSM N° 0181 24/07/21 10:36:24 2.5.1202 El paciente se ha movido CSM N° 0181 24/07/21 10:41:05 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (69.3 W). CSM N° 0181 24/07/21 10:41:09 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0181 24/07/21 10:41:14 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (77.6 %). CSM N° 0181 24/07/21 10:41:16 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:41:21 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (77.2 %). CSM N° 0181 24/07/21 10:41:22 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:41:27 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (76.8 %). CSM N° 0181 24/07/21 10:41:28 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:41:31 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (74.8 %). CSM N° 0181 24/07/21 10:41:31 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, CSM N° 0181 24/07/21 10:41:35 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (73.9 %). CSM N° 0181 24/07/21 10:41:36 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, CSM N° 0181 24/07/21 10:41:42 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (79.7 W). CSM N° 0181 24/07/21 10:41:44 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, CSM N° 0181 24/07/21 10:41:47 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (79.7 W). CSM N° 0181 24/07/21 10:41:48 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, CSM N° 0181 24/07/21 10:41:53 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (75.1 %). CSM N° 0181 24/07/21 10:41:54 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:41:57 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (79.5 W). CSM N° 0181 24/07/21 10:41:58 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:42:01 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (79.0 W). CSM N° 0181 24/07/21 10:42:02 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:42:05 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (78.6 W). CSM N° 0181 24/07/21 10:42:06 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:42:09 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (77.8 W). CSM N° 0181 24/07/21 10:42:16 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, CSM N° 0181 24/07/21 10:42:21 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (80.2 W). CSM N° 0181 24/07/21 10:42:21 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, CSM N° 0181 24/07/21 10:42:25 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (74.4 %). CSM N° 0181 24/07/21 10:42:26 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:42:31 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (71.5 %). CSM N° 0181 24/07/21 10:42:32 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0181 24/07/21 10:42:36 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (70.4 %). CSM N° 0181 24/07/21 10:42:36 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0181 24/07/21 10:42:39 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (73.1 %). CSM N° 0181 24/07/21 10:42:40 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0181 24/07/21 10:42:43 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (77.2 W). CSM N° 0181 24/07/21 10:42:44 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:42:49 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (78.3 W). CSM N° 0181 24/07/21 10:42:51 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:42:53 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (78.4 W). CSM N° 0181 24/07/21 10:42:54 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:42:57 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (78.4 W). CSM N° 0181 24/07/21 10:43:03 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:43:06 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (80.2 W). CSM N° 0181 24/07/21 10:43:07 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:43:12 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (80.4 W). CSM N° 0181 24/07/21 10:43:12 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:43:16 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (80.0 W). CSM N° 0181 24/07/21 10:43:16 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0181 24/07/21 10:43:22 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (80.1 W). CSM N° 0181 24/07/21 10:43:30 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:43:33 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (81.9 W). CSM N° 0181 24/07/21 10:43:34 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:43:39 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.6 %). CSM N° 0181 24/07/21 10:43:49 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (82.5 W). CSM N° 0181 24/07/21 10:43:50 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:43:54 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (81.0 W). CSM N° 0181 24/07/21 10:43:55 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:43:58 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (80.8 W). CSM N° 0181 24/07/21 10:43:59 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:44:02 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (79.8 W). CSM N° 0181 24/07/21 10:44:03 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:44:06 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (79.1 W). CSM N° 0181 24/07/21 10:44:07 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:44:10 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (74.5 %). CSM N° 0181 24/07/21 10:44:11 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0181 24/07/21 10:44:16 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (75.7 %). CSM N° 0181 24/07/21 10:44:17 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0181 24/07/21 10:45:45 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (72.0 %). CSM N° 0181 24/07/21 10:45:47 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0181 24/07/21 10:45:50 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (75.2 %). CSM N° 0181 24/07/21 10:45:51 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0181 24/07/21 10:45:57 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (80.3 W). CSM N° 0181 24/07/21 10:46:02 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:46:05 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (81.3 W). CSM N° 0181 24/07/21 10:46:06 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:46:09 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (81.1 W). CSM N° 0181 24/07/21 10:46:10 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0181 24/07/21 10:46:13 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (81.1 W). CSM N° 0181 24/07/21 10:46:14 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0181 24/07/21 10:46:18 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (82.1 W). CSM N° 0181 24/07/21 10:46:19 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0181 24/07/21 10:46:22 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (81.3 W). CSM N° 0181 24/07/21 10:46:22 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:46:26 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (82.0 W). CSM N° 0181 24/07/21 10:46:27 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:46:31 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (81.9 W). CSM N° 0181 24/07/21 10:46:34 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:47:08 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (82.7 W). CSM N° 0181 24/07/21 10:47:09 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 12, 13, 14, 15, CSM N° 0181 24/07/21 10:47:18 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (82.9 %). CSM N° 0181 24/07/21 10:47:25 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.1 W). CSM N° 0181 24/07/21 10:47:27 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 8, 12, 14, CSM N° 0181 24/07/21 10:47:30 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.5 W). CSM N° 0181 24/07/21 10:47:32 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 12, 14, CSM N° 0181 24/07/21 10:48:09 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (85.3 W). CSM N° 0181 24/07/21 10:48:12 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 12, 13, 14, CSM N° 0181 24/07/21 10:48:19 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.2 %). CSM N° 0181 24/07/21 10:48:23 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.5 W). CSM N° 0181 24/07/21 10:48:27 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.1 W). CSM N° 0181 24/07/21 10:48:34 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (84.9 W). CSM N° 0181 24/07/21 10:48:35 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 8, 12, 14, CSM N° 0181 24/07/21 10:48:40 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.1 %). CSM N° 0181 24/07/21 10:48:41 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 11, 12, CSM N° 0181 24/07/21 10:48:46 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.5 %). CSM N° 0181 24/07/21 10:48:48 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 11, 12, CSM N° 0181 24/07/21 10:48:48 2.8.1202 El paciente se ha movido CSM N° 0181 24/07/21 10:48:48 2.8.1202 El paciente se ha movido CSM N° 0181 24/07/21 10:49:00 2.8.1202 El paciente se ha movido CSM N° 0181 24/07/21 10:49:00 2.8.1202 El paciente se ha movido CSM N° 0181 24/07/21 10:49:22 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (77.4 W). CSM N° 0181 24/07/21 10:49:26 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0181 24/07/21 10:49:31 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (79.1 %). CSM N° 0181 24/07/21 10:49:35 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0181 24/07/21 10:49:37 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (79.0 %). CSM N° 0181 24/07/21 10:49:38 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0181 24/07/21 10:49:40 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (79.0 %). CSM N° 0181 24/07/21 10:49:48 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0181 24/07/21 10:49:55 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (77.7 W). CSM N° 0181 24/07/21 10:49:56 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0181 24/07/21 10:49:59 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (78.9 W). CSM N° 0181 24/07/21 10:50:00 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 9, 10, 11, CSM N° 0181 24/07/21 10:50:05 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (81.0 %). CSM N° 0181 24/07/21 10:50:06 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 11, 12, CSM N° 0181 24/07/21 10:50:11 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (82.4 %). CSM N° 0181 24/07/21 10:50:11 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 12, CSM N° 0181 24/07/21 10:50:17 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.3 W). CSM N° 0181 24/07/21 10:50:30 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (87.3 W). CSM N° 0181 24/07/21 10:50:34 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (88.2 W). CSM N° 0181 24/07/21 10:50:40 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (89.2 W). CSM N° 0181 24/07/21 10:50:53 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (90.0 W). CSM N° 0181 24/07/21 10:51:00 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.5 %). CSM N° 0181 24/07/21 10:51:04 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 10, CSM N° 0181 24/07/21 10:51:11 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.4 %). CSM N° 0181 24/07/21 10:51:13 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0181 24/07/21 10:51:17 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.7 %). CSM N° 0181 24/07/21 10:51:20 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0181 24/07/21 10:51:24 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.9 %). CSM N° 0181 24/07/21 10:51:29 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.4 %). CSM N° 0181 24/07/21 10:51:35 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.6 %). CSM N° 0181 24/07/21 10:51:41 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.8 %). CSM N° 0181 24/07/21 10:51:49 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.7 %). CSM N° 0181 24/07/21 10:51:50 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, CSM N° 0181 24/07/21 10:51:55 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.9 W). CSM N° 0181 24/07/21 10:52:05 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, 9, CSM N° 0181 24/07/21 10:54:31 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (27.4 mm) tiene que estar dentro del intervalo de 17.4 mm a 25.4 mm CSM N° 0181 24/07/21 10:54:44 1.8.1302 Error de posición de la pared rectal La posición de la pared rectal medida (27.2 mm) tiene que estar dentro del intervalo de 18.9 mm a 26.9 mm CSM N° 0181 24/07/21 10:55:05 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (24.0 mm) tiene que estar dentro del intervalo de 21.9 mm a 23.9 mm CSM N° 0181 24/07/21 10:55:50 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (75.8 W). CSM N° 0181 24/07/21 10:56:01 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 15, CSM N° 0181 24/07/21 10:56:04 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (81.5 W). CSM N° 0181 24/07/21 10:56:49 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 9, 10, 11, 12, CSM N° 0181 24/07/21 10:56:52 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (81.9 W). CSM N° 0181 24/07/21 10:56:54 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 11, 12, CSM N° 0181 24/07/21 10:57:01 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (87.5 W). CSM N° 0181 24/07/21 10:57:02 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 12, 13, 14, 15, CSM N° 0181 24/07/21 10:57:08 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.4 %). CSM N° 0181 24/07/21 10:57:19 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.3 %). CSM N° 0181 24/07/21 10:57:33 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.3 %). CSM N° 0181 24/07/21 10:57:45 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.1 %). CSM N° 0181 24/07/21 10:58:05 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.5 %). CSM N° 0181 24/07/21 10:58:34 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (84.3 %). CSM N° 0181 24/07/21 10:58:35 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 11, CSM N° 0181 24/07/21 10:58:42 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.8 %). CSM N° 0181 24/07/21 10:58:51 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.0 %). CSM N° 0181 24/07/21 10:59:02 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (88.6 %). CSM N° 0181 24/07/21 10:59:06 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.2 %). CSM N° 0181 24/07/21 10:59:08 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0181 24/07/21 10:59:13 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.5 %). CSM N° 0181 24/07/21 10:59:19 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.8 %). CSM N° 0181 24/07/21 10:59:28 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.5 %). CSM N° 0181 24/07/21 10:59:34 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (88.1 W). CSM N° 0181 24/07/21 10:59:35 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, 9, CSM N° 0181 24/07/21 11:01:27 2.7.1202 El paciente se ha movido CSM N° 0181 24/07/21 11:01:51 1.9.5104 Tratamiento cancelado en la fase de localización transversal CSM N° 0181 24/07/21 11:02:33 2.5.1202 El paciente se ha movido CSM N° 0181 24/07/21 11:02:45 2.5.1202 El paciente se ha movido CSM N° 0181 24/07/21 11:06:03 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (77.5 W). CSM N° 0181 24/07/21 11:06:06 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, CSM N° 0181 24/07/21 11:06:24 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.6 W). CSM N° 0181 24/07/21 11:06:28 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 9, CSM N° 0181 24/07/21 11:08:52 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.9%). CSM N° 0181 24/07/21 11:08:59 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, CSM N° 0181 24/07/21 11:09:03 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.9%). CSM N° 0181 24/07/21 11:09:05 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, CSM N° 0181 24/07/21 11:09:48 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.4%). CSM N° 0181 24/07/21 11:10:02 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, CSM N° 0181 24/07/21 11:10:07 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (111.1 W). CSM N° 0181 24/07/21 11:10:12 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 7, 8, CSM N° 0181 24/07/21 11:10:18 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). CSM N° 0181 24/07/21 11:10:19 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, CSM N° 0181 24/07/21 11:10:23 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (111.8 W). CSM N° 0181 24/07/21 11:10:31 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, CSM N° 0181 24/07/21 11:12:29 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.1%). CSM N° 0181 24/07/21 11:12:41 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, CSM N° 0181 24/07/21 11:14:20 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). CSM N° 0181 24/07/21 11:14:23 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, CSM N° 0181 24/07/21 11:14:29 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (114.0 W). CSM N° 0181 24/07/21 11:14:32 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, CSM N° 0181 24/07/21 11:14:37 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.7%). CSM N° 0181 24/07/21 11:14:40 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, CSM N° 0181 24/07/21 11:14:43 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (113.3 W). CSM N° 0181 24/07/21 11:14:44 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, CSM N° 0181 24/07/21 11:16:24 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (111.9 W). CSM N° 0181 24/07/21 11:16:29 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, CSM N° 0181 24/07/21 11:16:33 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.6%). CSM N° 0181 24/07/21 11:16:35 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, CSM N° 0181 24/07/21 11:16:39 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.9%). CSM N° 0181 24/07/21 11:16:41 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, CSM N° 0181 24/07/21 11:16:44 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (113.8 W). CSM N° 0181 24/07/21 11:16:46 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, CSM N° 0181 24/07/21 11:16:51 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.3%). CSM N° 0181 24/07/21 11:16:52 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0181 24/07/21 11:20:39 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (113.2 W). CSM N° 0181 24/07/21 11:20:43 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 11, CSM N° 0181 24/07/21 11:20:46 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (115.5 W). CSM N° 0181 24/07/21 11:20:47 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 11, CSM N° 0181 24/07/21 11:21:01 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.3%). CSM N° 0181 24/07/21 11:21:02 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0181 24/07/21 11:21:46 1.9.5105 Tratamiento cancelado durante la fase de disparo CSM N° 0182 24/07/21 12:56:58 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.8 %). CSM N° 0182 24/07/21 12:57:02 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 12, 13, 14, 15, CSM N° 0182 24/07/21 12:57:04 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.8 %). CSM N° 0182 24/07/21 12:57:08 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (85.9 %). CSM N° 0182 24/07/21 12:57:14 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.1 %). CSM N° 0182 24/07/21 12:57:22 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.1 %). CSM N° 0182 24/07/21 12:57:29 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (75.5 W). CSM N° 0182 24/07/21 12:57:30 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, 16, CSM N° 0182 24/07/21 12:57:34 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (86.8 W). CSM N° 0182 24/07/21 12:57:40 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.2 %). CSM N° 0182 24/07/21 12:57:44 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (87.2 %). CSM N° 0182 24/07/21 12:57:49 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (77.5 W). CSM N° 0182 24/07/21 12:57:52 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0182 24/07/21 12:57:54 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (77.7 W). CSM N° 0182 24/07/21 12:57:56 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0182 24/07/21 12:57:59 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (74.2 %). CSM N° 0182 24/07/21 12:58:00 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, CSM N° 0182 24/07/21 12:58:03 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (69.6 %). CSM N° 0182 24/07/21 12:58:04 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 14, CSM N° 0182 24/07/21 12:58:08 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (69.0 %). CSM N° 0182 24/07/21 12:58:08 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, CSM N° 0182 24/07/21 12:58:12 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (72.1 W). CSM N° 0182 24/07/21 12:58:13 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 13, CSM N° 0182 24/07/21 12:58:16 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (76.0 %). CSM N° 0182 24/07/21 12:58:17 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 11, CSM N° 0182 24/07/21 12:58:20 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (83.7 %). CSM N° 0182 24/07/21 12:58:21 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 12, CSM N° 0182 24/07/21 12:58:23 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (86.5 %). CSM N° 0182 24/07/21 12:58:24 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, CSM N° 0182 24/07/21 12:58:27 2.8.1003 probe number UV167-----------, Potencia medida insuficiente: menos del 90 % de lo solicitado (89.9 %). CSM N° 0182 24/07/21 12:59:58 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.9%). CSM N° 0182 24/07/21 13:00:05 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, CSM N° 0182 24/07/21 13:00:08 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (112.7 W). CSM N° 0182 24/07/21 13:00:09 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, CSM N° 0182 24/07/21 13:00:13 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.9%). CSM N° 0182 24/07/21 13:00:14 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, CSM N° 0182 24/07/21 13:02:02 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (115.0 W). CSM N° 0182 24/07/21 13:02:06 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 11, CSM N° 0182 24/07/21 13:04:15 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.4%). CSM N° 0182 24/07/21 13:04:18 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0182 24/07/21 13:04:23 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.9%). CSM N° 0182 24/07/21 13:04:24 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0182 24/07/21 13:04:27 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (115.5 W). CSM N° 0182 24/07/21 13:04:32 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, CSM N° 0182 24/07/21 13:04:35 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (118.5 W). CSM N° 0182 24/07/21 13:04:37 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, CSM N° 0182 24/07/21 13:04:40 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (119.2 W). CSM N° 0182 24/07/21 13:04:42 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, CSM N° 0182 24/07/21 13:04:45 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (117.3 W). CSM N° 0182 24/07/21 13:04:47 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, CSM N° 0182 24/07/21 13:04:50 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.9%). CSM N° 0182 24/07/21 13:04:54 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). CSM N° 0182 24/07/21 13:06:24 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.4%). CSM N° 0182 24/07/21 13:06:32 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.2%). CSM N° 0182 24/07/21 13:06:36 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.3%). CSM N° 0182 24/07/21 13:06:40 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (118.3 W). CSM N° 0182 24/07/21 13:06:44 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0182 24/07/21 13:06:47 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (119.4 W). CSM N° 0182 24/07/21 13:06:48 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0182 24/07/21 13:06:52 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (121.6 W). CSM N° 0182 24/07/21 13:06:53 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 11, CSM N° 0182 24/07/21 13:06:56 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (119.6 W). CSM N° 0182 24/07/21 13:06:58 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0182 24/07/21 13:07:01 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.9%). CSM N° 0182 24/07/21 13:07:02 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, 10, CSM N° 0182 24/07/21 13:07:05 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.3%). CSM N° 0182 24/07/21 13:07:07 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, CSM N° 0182 24/07/21 13:07:11 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.0%). CSM N° 0182 24/07/21 13:07:13 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, CSM N° 0182 24/07/21 13:07:18 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.5%). CSM N° 0182 24/07/21 13:07:19 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, CSM N° 0182 24/07/21 13:07:55 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (111.4 W). CSM N° 0182 24/07/21 13:08:12 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (116.9 W). CSM N° 0182 24/07/21 13:08:16 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 11, CSM N° 0182 24/07/21 13:08:53 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.6%). CSM N° 0182 24/07/21 13:09:04 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.7%). CSM N° 0182 24/07/21 13:09:08 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0182 24/07/21 13:09:11 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.1%). CSM N° 0182 24/07/21 13:09:16 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.0%). CSM N° 0182 24/07/21 13:09:19 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, CSM N° 0182 24/07/21 13:09:22 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (118.7 W). CSM N° 0182 24/07/21 13:09:23 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0182 24/07/21 13:09:27 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (122.2 W). CSM N° 0182 24/07/21 13:09:32 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 11, CSM N° 0182 24/07/21 13:12:02 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.9%). CSM N° 0182 24/07/21 13:12:10 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.5%). CSM N° 0182 24/07/21 13:12:11 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 8, CSM N° 0182 24/07/21 13:12:14 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.7%). CSM N° 0182 24/07/21 13:12:19 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.8%). CSM N° 0182 24/07/21 13:12:20 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, 10, CSM N° 0182 24/07/21 13:12:23 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (119.8 W). CSM N° 0182 24/07/21 13:12:25 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0182 24/07/21 13:12:34 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (118.0%). CSM N° 0182 24/07/21 13:12:36 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, CSM N° 0182 24/07/21 13:12:39 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.6%). CSM N° 0182 24/07/21 13:12:41 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, CSM N° 0182 24/07/21 13:12:45 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.4%). CSM N° 0182 24/07/21 13:12:47 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0182 24/07/21 13:13:45 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (118.2%). CSM N° 0182 24/07/21 13:13:48 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, CSM N° 0182 24/07/21 13:13:51 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (121.3%). CSM N° 0182 24/07/21 13:13:52 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, CSM N° 0182 24/07/21 13:14:45 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.8%). CSM N° 0182 24/07/21 13:14:48 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, CSM N° 0182 24/07/21 13:14:51 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.2%). CSM N° 0182 24/07/21 13:14:55 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (119.1%). CSM N° 0182 24/07/21 13:14:57 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0182 24/07/21 13:15:00 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (122.3%). CSM N° 0182 24/07/21 13:15:01 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, CSM N° 0182 24/07/21 13:15:04 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.7%). CSM N° 0182 24/07/21 13:15:05 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, CSM N° 0182 24/07/21 13:15:10 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.7%). CSM N° 0182 24/07/21 13:15:12 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0182 24/07/21 13:16:04 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (121.9%). CSM N° 0182 24/07/21 13:16:06 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, CSM N° 0182 24/07/21 13:16:51 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.4%). CSM N° 0182 24/07/21 13:16:53 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 9, CSM N° 0182 24/07/21 13:16:55 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.1%). CSM N° 0182 24/07/21 13:16:57 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 10, CSM N° 0182 24/07/21 13:17:00 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (122.4%). CSM N° 0182 24/07/21 13:17:01 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, CSM N° 0182 24/07/21 13:18:59 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.2%). CSM N° 0182 24/07/21 13:19:08 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.4%). CSM N° 0182 24/07/21 13:19:10 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 9, CSM N° 0182 24/07/21 13:20:16 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.1%). CSM N° 0182 24/07/21 13:20:18 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, CSM N° 0182 24/07/21 13:20:34 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.6%). CSM N° 0182 24/07/21 13:20:36 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0182 24/07/21 13:20:42 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.4%). CSM N° 0182 24/07/21 13:20:43 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, CSM N° 0182 24/07/21 13:20:49 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.7%). CSM N° 0182 24/07/21 13:20:52 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, CSM N° ---- 24/07/21 13:37:50 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 24/07/21 13:38:30 2.2.1604 El escáner no puede pasar a modo operativo CSM N° ---- 24/07/21 14:08:10 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 24/07/21 14:10:21 2.2.1423 La posición real está más allá de la posición solicitada en antero-posterior axis CSM N° ---- 24/07/21 14:10:43 2.2.1423 La posición real está más allá de la posición solicitada en antero-posterior axis CSM N° ---- 24/07/21 14:14:28 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 24/07/21 14:14:52 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 24/07/21 14:14:54 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 24/07/21 14:15:28 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 24/07/21 14:15:29 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 24/07/21 14:23:43 2.3.1901 Código Focalpak incorrecto CSM N° 0183 24/07/21 14:27:33 2.5.1202 El paciente se ha movido CSM N° 0183 24/07/21 14:31:38 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (60.7 W). CSM N° 0183 24/07/21 14:31:44 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, 15, CSM N° 0183 24/07/21 14:31:47 2.8.1002 probe number UV167-----------, Potencia medida insuficiente: más de 16W por debajo de lo solicitado (74.4 W). CSM N° 0183 24/07/21 14:31:47 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 1, 2, 3, 4, 5, 6, 8, 9, 10, 11, 12, 13, 14, CSM N° 0183 24/07/21 14:32:32 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.2%). CSM N° 0183 24/07/21 14:32:35 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, CSM N° 0183 24/07/21 14:32:38 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.4%). CSM N° 0183 24/07/21 14:32:40 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, CSM N° 0183 24/07/21 14:32:45 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.2%). CSM N° 0183 24/07/21 14:32:46 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:33:23 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (116.6 W). CSM N° 0183 24/07/21 14:33:25 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:33:30 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (116.9%). CSM N° 0183 24/07/21 14:33:31 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:33:36 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.5%). CSM N° 0183 24/07/21 14:33:37 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:33:42 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.6 W). CSM N° 0183 24/07/21 14:33:43 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0183 24/07/21 14:33:55 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.6 W). CSM N° 0183 24/07/21 14:33:57 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0183 24/07/21 14:34:01 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.8 W). CSM N° 0183 24/07/21 14:34:03 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0183 24/07/21 14:34:08 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.7 W). CSM N° 0183 24/07/21 14:34:10 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0183 24/07/21 14:34:15 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.8 W). CSM N° 0183 24/07/21 14:34:16 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0183 24/07/21 14:34:19 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (115.9%). CSM N° 0183 24/07/21 14:34:20 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:34:24 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (116.3%). CSM N° 0183 24/07/21 14:34:25 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:34:30 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.2%). CSM N° 0183 24/07/21 14:34:32 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:34:35 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.5%). CSM N° 0183 24/07/21 14:34:36 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, CSM N° 0183 24/07/21 14:34:39 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.4%). CSM N° 0183 24/07/21 14:34:43 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). CSM N° 0183 24/07/21 14:34:48 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.9%). CSM N° 0183 24/07/21 14:34:49 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, CSM N° 0183 24/07/21 14:34:52 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). CSM N° 0183 24/07/21 14:34:56 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (117.6%). CSM N° 0183 24/07/21 14:34:57 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:35:00 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (119.6%). CSM N° 0183 24/07/21 14:35:03 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:35:07 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (119.9%). CSM N° 0183 24/07/21 14:35:08 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:35:11 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (117.7%). CSM N° 0183 24/07/21 14:35:13 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:35:16 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (118.0%). CSM N° 0183 24/07/21 14:35:17 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:35:20 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.8%). CSM N° 0183 24/07/21 14:35:21 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, CSM N° 0183 24/07/21 14:35:24 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.8%). CSM N° 0183 24/07/21 14:35:25 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, CSM N° 0183 24/07/21 14:35:28 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.8%). CSM N° 0183 24/07/21 14:35:29 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:35:32 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.1%). CSM N° 0183 24/07/21 14:35:33 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, CSM N° 0183 24/07/21 14:35:38 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.9%). CSM N° 0183 24/07/21 14:35:39 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:35:42 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.0%). CSM N° 0183 24/07/21 14:35:43 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, CSM N° 0183 24/07/21 14:35:46 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (118.7%). CSM N° 0183 24/07/21 14:35:48 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:36:36 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (116.3%). CSM N° 0183 24/07/21 14:36:37 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, CSM N° 0183 24/07/21 14:36:40 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (115.5%). CSM N° 0183 24/07/21 14:36:42 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:36:47 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.5%). CSM N° 0183 24/07/21 14:36:52 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.2%). CSM N° 0183 24/07/21 14:36:53 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, CSM N° 0183 24/07/21 14:36:56 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.9%). CSM N° 0183 24/07/21 14:37:00 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.5%). CSM N° 0183 24/07/21 14:37:06 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.0%). CSM N° 0183 24/07/21 14:37:07 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:37:12 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.2%). CSM N° 0183 24/07/21 14:37:14 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:37:17 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.5%). CSM N° 0183 24/07/21 14:37:18 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, CSM N° 0183 24/07/21 14:37:21 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (118.7%). CSM N° 0183 24/07/21 14:37:21 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:37:25 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (118.6%). CSM N° 0183 24/07/21 14:37:26 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:37:29 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.0%). CSM N° 0183 24/07/21 14:37:30 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 7, 8, 9, CSM N° 0183 24/07/21 14:37:35 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.3%). CSM N° 0183 24/07/21 14:37:36 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:37:41 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.3%). CSM N° 0183 24/07/21 14:37:42 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:37:47 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.9%). CSM N° 0183 24/07/21 14:37:49 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:37:52 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.9%). CSM N° 0183 24/07/21 14:37:53 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 7, 8, 9, CSM N° 0183 24/07/21 14:37:56 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.8%). CSM N° 0183 24/07/21 14:37:56 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, CSM N° 0183 24/07/21 14:37:59 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (119.4%). CSM N° 0183 24/07/21 14:38:00 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:38:03 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (119.6%). CSM N° 0183 24/07/21 14:38:04 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:38:07 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (122.3%). CSM N° 0183 24/07/21 14:38:08 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:38:12 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (122.4%). CSM N° 0183 24/07/21 14:38:15 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:38:18 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (119.6%). CSM N° 0183 24/07/21 14:38:19 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:38:22 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.5%). CSM N° 0183 24/07/21 14:38:23 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:38:26 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.5%). CSM N° 0183 24/07/21 14:38:27 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:38:32 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.2%). CSM N° 0183 24/07/21 14:38:33 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, CSM N° 0183 24/07/21 14:38:36 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.2%). CSM N° 0183 24/07/21 14:38:36 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:38:42 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.7%). CSM N° 0183 24/07/21 14:38:42 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:38:48 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.3%). CSM N° 0183 24/07/21 14:38:49 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:38:54 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.6%). CSM N° 0183 24/07/21 14:38:54 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:38:57 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (120.0%). CSM N° 0183 24/07/21 14:38:58 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:39:02 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (120.7%). CSM N° 0183 24/07/21 14:39:05 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0183 24/07/21 14:39:12 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.4%). CSM N° 0183 24/07/21 14:39:13 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 9, 10, CSM N° 0183 24/07/21 14:39:18 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.4%). CSM N° 0183 24/07/21 14:39:19 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:39:26 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.3%). CSM N° 0183 24/07/21 14:39:27 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:39:32 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.4%). CSM N° 0183 24/07/21 14:39:33 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:39:36 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.7%). CSM N° 0183 24/07/21 14:39:37 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:39:40 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.7%). CSM N° 0183 24/07/21 14:39:40 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:39:43 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.7%). CSM N° 0183 24/07/21 14:39:45 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:39:48 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (120.6%). CSM N° 0183 24/07/21 14:39:49 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:39:52 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (122.7%). CSM N° 0183 24/07/21 14:39:53 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:39:58 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (123.2%). CSM N° 0183 24/07/21 14:40:00 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:40:03 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (120.5%). CSM N° 0183 24/07/21 14:40:05 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0183 24/07/21 14:40:08 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.9%). CSM N° 0183 24/07/21 14:40:09 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:40:14 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.5%). CSM N° 0183 24/07/21 14:40:15 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:40:18 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.4%). CSM N° 0183 24/07/21 14:40:19 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:40:24 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.3%). CSM N° 0183 24/07/21 14:40:26 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:40:31 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.1%). CSM N° 0183 24/07/21 14:40:32 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:40:37 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.6%). CSM N° 0183 24/07/21 14:40:39 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:40:44 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.4%). CSM N° 0183 24/07/21 14:40:45 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:40:48 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (121.5%). CSM N° 0183 24/07/21 14:40:49 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0183 24/07/21 14:41:36 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (118.1%). CSM N° 0183 24/07/21 14:41:38 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:41:43 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (117.4%). CSM N° 0183 24/07/21 14:41:44 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:41:47 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.3%). CSM N° 0183 24/07/21 14:41:49 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:41:52 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.6%). CSM N° 0183 24/07/21 14:41:55 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, CSM N° 0183 24/07/21 14:42:00 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.8%). CSM N° 0183 24/07/21 14:42:01 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, CSM N° 0183 24/07/21 14:42:06 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.7%). CSM N° 0183 24/07/21 14:42:07 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:42:12 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.9%). CSM N° 0183 24/07/21 14:42:13 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:42:18 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.6%). CSM N° 0183 24/07/21 14:42:20 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 9, 10, CSM N° 0183 24/07/21 14:42:23 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.2%). CSM N° 0183 24/07/21 14:42:24 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, CSM N° 0183 24/07/21 14:42:27 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (120.2%). CSM N° 0183 24/07/21 14:42:28 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0183 24/07/21 14:42:32 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (120.2%). CSM N° 0183 24/07/21 14:42:33 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:42:36 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.9%). CSM N° 0183 24/07/21 14:42:39 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:42:44 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.6%). CSM N° 0183 24/07/21 14:42:45 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:42:51 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.3%). CSM N° 0183 24/07/21 14:42:52 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:42:59 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.0%). CSM N° 0183 24/07/21 14:43:00 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 10, CSM N° 0183 24/07/21 14:43:03 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.5%). CSM N° 0183 24/07/21 14:43:04 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:43:07 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.0%). CSM N° 0183 24/07/21 14:43:08 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:43:11 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.7%). CSM N° 0183 24/07/21 14:43:12 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:43:15 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (120.2%). CSM N° 0183 24/07/21 14:43:16 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0183 24/07/21 14:43:19 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (122.9%). CSM N° 0183 24/07/21 14:43:20 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:43:23 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (122.4%). CSM N° 0183 24/07/21 14:43:24 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:43:28 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (120.2%). CSM N° 0183 24/07/21 14:43:29 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:43:32 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.2%). CSM N° 0183 24/07/21 14:43:33 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:43:36 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.1%). CSM N° 0183 24/07/21 14:43:36 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:43:39 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.6%). CSM N° 0183 24/07/21 14:43:42 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:43:47 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (114.3%). CSM N° 0183 24/07/21 14:43:48 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:43:53 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.4%). CSM N° 0183 24/07/21 14:43:54 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:43:59 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.0%). CSM N° 0183 24/07/21 14:44:02 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.9%). CSM N° 0183 24/07/21 14:44:05 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:44:08 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (120.2%). CSM N° 0183 24/07/21 14:44:09 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0183 24/07/21 14:44:13 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (120.1%). CSM N° 0183 24/07/21 14:44:16 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:48:04 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). CSM N° 0183 24/07/21 14:48:06 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, CSM N° 0183 24/07/21 14:48:17 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.6 W). CSM N° 0183 24/07/21 14:48:18 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0183 24/07/21 14:48:23 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.7 W). CSM N° 0183 24/07/21 14:48:25 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0183 24/07/21 14:48:31 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.9 W). CSM N° 0183 24/07/21 14:48:33 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0183 24/07/21 14:48:36 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). CSM N° 0183 24/07/21 14:48:41 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.8%). CSM N° 0183 24/07/21 14:48:45 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (117.5%). CSM N° 0183 24/07/21 14:48:46 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:48:49 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (120.4%). CSM N° 0183 24/07/21 14:48:49 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:48:53 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (120.9%). CSM N° 0183 24/07/21 14:48:54 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:48:59 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (118.6%). CSM N° 0183 24/07/21 14:49:00 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:49:03 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.5%). CSM N° 0183 24/07/21 14:49:04 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:49:07 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.5%). CSM N° 0183 24/07/21 14:49:36 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (119.5%). CSM N° 0183 24/07/21 14:49:38 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:49:42 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (118.1%). CSM N° 0183 24/07/21 14:49:43 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:49:46 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (119.2%). CSM N° 0183 24/07/21 14:49:47 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:49:50 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.7%). CSM N° 0183 24/07/21 14:49:51 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:49:54 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.6%). CSM N° 0183 24/07/21 14:49:55 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:49:58 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.8%). CSM N° 0183 24/07/21 14:49:58 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:50:03 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.9%). CSM N° 0183 24/07/21 14:50:04 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:50:07 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.9%). CSM N° 0183 24/07/21 14:50:08 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 6, 7, 8, 9, CSM N° 0183 24/07/21 14:50:12 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.6%). CSM N° 0183 24/07/21 14:50:13 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:50:41 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.0%). CSM N° 0183 24/07/21 14:50:48 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.0%). CSM N° 0183 24/07/21 14:50:49 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, CSM N° 0183 24/07/21 14:50:54 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.6%). CSM N° 0183 24/07/21 14:50:55 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:50:58 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.7%). CSM N° 0183 24/07/21 14:50:59 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:51:02 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.7%). CSM N° 0183 24/07/21 14:51:03 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:51:05 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.5%). CSM N° 0183 24/07/21 14:51:06 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:51:09 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (120.3%). CSM N° 0183 24/07/21 14:51:10 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0183 24/07/21 14:51:13 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (120.6%). CSM N° 0183 24/07/21 14:51:14 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, 11, CSM N° 0183 24/07/21 14:51:18 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (124.3 W). CSM N° 0183 24/07/21 14:51:18 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:51:25 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (123.1%). CSM N° 0183 24/07/21 14:51:26 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, 7, 8, 9, CSM N° 0183 24/07/21 14:51:30 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (120.4%). CSM N° 0183 24/07/21 14:51:31 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:51:34 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (120.5%). CSM N° 0183 24/07/21 14:51:35 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 3, 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:51:39 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.9%). CSM N° 0183 24/07/21 14:51:40 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 4, 5, 6, 7, 8, 9, 10, CSM N° 0183 24/07/21 14:51:59 1.9.5104 Tratamiento cancelado en la fase de localización transversal EDAP N° ---- 26/07/21 14:27:35 2.2.1604 El escáner no puede pasar a modo operativo EDAP N° ---- 26/07/21 14:27:45 2.2.1604 El escáner no puede pasar a modo operativo EDAP N° ---- 26/07/21 14:27:46 2.2.1604 El escáner no puede pasar a modo operativo EDAP N° ---- 26/07/21 14:27:47 2.2.1604 El escáner no puede pasar a modo operativo EDAP N° ---- 26/07/21 14:27:48 2.2.1604 El escáner no puede pasar a modo operativo EDAP N° ---- 26/07/21 14:31:52 2.3.1901 Código Focalpak incorrecto EDAP N° 0145 26/07/21 14:33:25 2.5.1604 El escáner no puede pasar a modo operativo EDAP N° 0145 26/07/21 14:33:25 2.5.1607 La sonda del escáner de ultrasonidos ha sido desconectada. EDAP N° 0145 26/07/21 14:33:31 1.5.5102 Tratamiento cancelado en la fase de adquisición EDAP N° 0149 28/07/21 19:54:22 2.5.1202 El paciente se ha movido EDAP N° 0149 28/07/21 19:57:32 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (111.0 W). EDAP N° 0149 28/07/21 19:57:40 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (110.3 W). EDAP N° 0149 28/07/21 19:58:15 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.0%). EDAP N° 0149 28/07/21 19:58:23 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.5%). EDAP N° 0149 28/07/21 19:58:28 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (111.4 W). EDAP N° 0149 28/07/21 19:58:44 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (111.9 W). EDAP N° 0149 28/07/21 19:58:48 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.9%). EDAP N° 0149 28/07/21 19:58:53 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). EDAP N° 0149 28/07/21 19:58:57 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.0%). EDAP N° 0149 28/07/21 19:59:02 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.8%). EDAP N° 0149 28/07/21 19:59:06 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.1%). EDAP N° 0149 28/07/21 19:59:10 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.5%). EDAP N° 0149 28/07/21 19:59:15 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.9%). EDAP N° 0149 28/07/21 19:59:25 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.5%). EDAP N° 0149 28/07/21 19:59:30 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.8%). EDAP N° 0149 28/07/21 19:59:36 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.6%). EDAP N° 0149 28/07/21 19:59:42 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (112.5 W). EDAP N° 0149 28/07/21 19:59:51 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (113.0 W). EDAP N° 0149 28/07/21 20:00:01 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.3%). EDAP N° 0149 28/07/21 20:00:07 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.5%). EDAP N° 0149 28/07/21 20:00:16 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (112.8 W). EDAP N° 0149 28/07/21 20:00:21 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (112.3 W). EDAP N° 0149 28/07/21 20:00:25 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.9%). EDAP N° 0149 28/07/21 20:00:29 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.4%). EDAP N° 0149 28/07/21 20:00:33 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.4%). EDAP N° 0149 28/07/21 20:00:37 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.2%). EDAP N° 0149 28/07/21 20:00:40 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.8%). EDAP N° 0149 28/07/21 20:00:44 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.3%). EDAP N° 0149 28/07/21 20:00:48 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.6%). EDAP N° 0149 28/07/21 20:00:52 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.3%). EDAP N° 0149 28/07/21 20:00:56 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.9%). EDAP N° 0149 28/07/21 20:01:00 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.4%). EDAP N° 0149 28/07/21 20:01:04 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (113.4 W). EDAP N° 0149 28/07/21 20:01:10 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.4%). EDAP N° 0149 28/07/21 20:01:15 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.3%). EDAP N° 0149 28/07/21 20:01:22 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (113.3 W). EDAP N° 0149 28/07/21 20:01:27 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.4%). EDAP N° 0149 28/07/21 20:01:31 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.0%). EDAP N° 0149 28/07/21 20:01:35 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.2%). EDAP N° 0149 28/07/21 20:01:39 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.2%). EDAP N° 0149 28/07/21 20:01:42 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.4%). EDAP N° 0149 28/07/21 20:01:46 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.7%). EDAP N° 0149 28/07/21 20:01:50 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.4%). EDAP N° 0149 28/07/21 20:01:54 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.5%). EDAP N° 0149 28/07/21 20:01:59 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.7%). EDAP N° 0149 28/07/21 20:02:04 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.0%). EDAP N° 0149 28/07/21 20:02:08 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (113.4 W). EDAP N° 0149 28/07/21 20:02:12 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (114.4 W). EDAP N° 0149 28/07/21 20:02:13 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, EDAP N° 0149 28/07/21 20:02:18 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.2%). EDAP N° 0149 28/07/21 20:03:22 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.6 W). EDAP N° 0149 28/07/21 20:03:23 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 8, 9, EDAP N° 0149 28/07/21 20:03:38 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.0 W). EDAP N° 0149 28/07/21 20:04:01 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.1 W). EDAP N° 0149 28/07/21 20:04:04 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 5, 6, 7, EDAP N° 0149 28/07/21 20:04:17 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.1 W). EDAP N° 0149 28/07/21 20:04:20 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 5, 6, 7, EDAP N° 0149 28/07/21 20:04:33 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.1 W). EDAP N° 0149 28/07/21 20:04:36 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 5, 6, 7, EDAP N° 0149 28/07/21 20:04:47 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.0 W). EDAP N° 0149 28/07/21 20:05:19 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, EDAP N° 0149 28/07/21 20:05:38 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.0 W). EDAP N° 0149 28/07/21 20:05:43 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, EDAP N° 0149 28/07/21 20:05:55 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.2 W). EDAP N° 0149 28/07/21 20:06:00 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 5, 6, 7, EDAP N° 0149 28/07/21 20:06:13 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.2 W). EDAP N° 0149 28/07/21 20:06:16 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 5, 6, 7, EDAP N° 0149 28/07/21 20:06:30 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.2 W). EDAP N° 0149 28/07/21 20:06:31 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 5, 6, 7, EDAP N° 0149 28/07/21 20:06:45 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.2 W). EDAP N° 0149 28/07/21 20:06:46 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 5, 6, 7, EDAP N° 0149 28/07/21 20:06:59 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.1 W). EDAP N° 0149 28/07/21 20:07:03 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 3, 4, 5, 6, EDAP N° 0149 28/07/21 20:13:47 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.6 W). EDAP N° 0149 28/07/21 20:13:53 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.8 W). EDAP N° 0149 28/07/21 20:14:00 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.0 W). EDAP N° 0149 28/07/21 20:14:08 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.0 W). EDAP N° 0149 28/07/21 20:14:16 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.0 W). EDAP N° 0149 28/07/21 20:14:39 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (1.8 W). EDAP N° 0149 28/07/21 20:14:45 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.0 W). EDAP N° 0149 28/07/21 20:14:58 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.0 W). EDAP N° 0149 28/07/21 20:15:06 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (112.8%). EDAP N° 0149 28/07/21 20:15:12 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (112.7 W). EDAP N° 0149 28/07/21 20:15:21 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (113.1 W). EDAP N° 0149 28/07/21 20:15:29 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (113.8%). EDAP N° 0149 28/07/21 20:15:37 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.2 W). EDAP N° 0149 28/07/21 20:15:43 2.8.1015 probe number UV167-----------, Potencia medida excesiva: más de 1,6W por encima de lo solicitado (2.3 W). EDAP N° 0149 28/07/21 20:15:46 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.1%). EDAP N° 0149 28/07/21 20:15:50 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.1%). EDAP N° 0149 28/07/21 20:15:54 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.1%). EDAP N° 0149 28/07/21 20:15:58 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.3%). EDAP N° 0149 28/07/21 20:16:02 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.8%). EDAP N° 0149 28/07/21 20:16:05 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.7%). EDAP N° 0149 28/07/21 20:16:09 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). EDAP N° 0149 28/07/21 20:16:14 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). EDAP N° 0149 28/07/21 20:16:18 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.1%). EDAP N° 0149 28/07/21 20:16:22 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.3%). EDAP N° 0149 28/07/21 20:16:25 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.9%). EDAP N° 0149 28/07/21 20:16:29 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.7%). EDAP N° 0149 28/07/21 20:16:33 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.8%). EDAP N° 0149 28/07/21 20:16:37 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.8%). EDAP N° 0149 28/07/21 20:16:41 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.7%). EDAP N° 0149 28/07/21 20:16:44 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.8%). EDAP N° 0149 28/07/21 20:16:48 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.0%). EDAP N° 0149 28/07/21 20:16:53 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (114.9 W). EDAP N° 0149 28/07/21 20:16:54 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, EDAP N° 0149 28/07/21 20:16:58 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (115.0 W). EDAP N° 0149 28/07/21 20:16:59 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, EDAP N° 0149 28/07/21 20:17:04 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.3%). EDAP N° 0149 28/07/21 20:17:07 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). EDAP N° 0149 28/07/21 20:17:11 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). EDAP N° 0149 28/07/21 20:17:15 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). EDAP N° 0149 28/07/21 20:17:18 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.5%). EDAP N° 0149 28/07/21 20:17:22 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). EDAP N° 0149 28/07/21 20:17:26 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.8%). EDAP N° 0149 28/07/21 20:17:31 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.8%). EDAP N° 0149 28/07/21 20:17:35 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.3%). EDAP N° 0149 28/07/21 20:17:39 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). EDAP N° 0149 28/07/21 20:17:43 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.9%). EDAP N° 0149 28/07/21 20:17:47 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). EDAP N° 0149 28/07/21 20:17:51 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.1%). EDAP N° 0149 28/07/21 20:17:55 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.6%). EDAP N° 0149 28/07/21 20:17:58 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.7%). EDAP N° 0149 28/07/21 20:18:02 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.9%). EDAP N° 0149 28/07/21 20:18:06 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.9%). EDAP N° 0149 28/07/21 20:18:10 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.0%). EDAP N° 0149 28/07/21 20:18:15 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (110.9%). EDAP N° 0149 28/07/21 20:18:22 2.8.1004 probe number UV167-----------, Potencia medida excesiva: más del 110 % de lo solicitado (111.3%). EDAP N° 0149 28/07/21 20:18:27 2.8.1001 probe number UV167-----------, Potencia medida excesiva: más de 16W por encima de lo solicitado (115.1 W). EDAP N° 0149 28/07/21 20:18:28 2.8.1006 probe number UV167-----------, Error de potencia en demasiados canales: en canal(es) 2, 4, 5, 6, EDAP N° 0150 03/08/21 14:08:56 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (24.4 mm) tiene que estar dentro del intervalo de 21.7 mm a 23.7 mm EDAP N° 0150 03/08/21 14:09:42 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.1 mm) tiene que estar dentro del intervalo de 23.8 mm a 25.8 mm EDAP N° 0150 03/08/21 14:09:58 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.6 mm) tiene que estar dentro del intervalo de 23.8 mm a 25.8 mm EDAP N° 0150 03/08/21 14:10:21 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (24.1 mm) tiene que estar dentro del intervalo de 24.4 mm a 26.4 mm EDAP N° 0150 03/08/21 14:11:19 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (25.0 mm) tiene que estar dentro del intervalo de 22.8 mm a 24.8 mm EDAP N° 0150 03/08/21 14:11:39 1.9.5105 Tratamiento cancelado durante la fase de disparo EDAP N° 0152 03/08/21 15:35:17 2.8.1202 El paciente se ha movido CSM N° ---- 25/08/21 11:01:11 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:02:53 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:03:05 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:03:41 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:03:46 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:03:50 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:04:13 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:04:17 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:04:22 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:04:48 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:04:52 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:04:55 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:05:28 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:05:33 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:05:37 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:05:44 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:06:00 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:09:57 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:10:04 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:10:16 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:10:20 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:10:27 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:10:31 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:10:35 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:10:38 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:10:45 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:10:48 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:10:54 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:11:03 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:11:11 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:11:23 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:11:30 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:11:34 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:11:39 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:11:50 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:11:54 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:12:03 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:12:07 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:12:36 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:12:40 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:12:44 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:12:48 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:12:52 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:12:56 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:13:01 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:13:09 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:13:28 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:13:31 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:13:35 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:13:40 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:13:46 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:13:50 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:13:54 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:13:58 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:14:02 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:14:06 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:14:14 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:14:18 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:14:35 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:14:39 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:14:54 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:15:03 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:15:07 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:15:11 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:15:15 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:15:18 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:15:22 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:15:26 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:15:30 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:15:52 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:16:35 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:16:43 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:16:49 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:16:53 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:16:57 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:17:01 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:17:07 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:17:11 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:17:34 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:17:38 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:17:42 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:18:04 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:18:07 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:18:27 2.2.1604 El escáner no puede pasar a modo operativo CSM N° ---- 25/08/21 11:19:13 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:19:17 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:19:21 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:19:28 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:19:33 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:19:37 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:19:45 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:20:04 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:20:09 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:20:13 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:20:16 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:20:31 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:20:41 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:20:45 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:20:49 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:20:53 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:20:57 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:21:01 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:21:16 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:21:26 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:21:30 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:21:34 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:21:37 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:21:41 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:21:45 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:26:11 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:26:19 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:26:26 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:26:31 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:26:59 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:27:03 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:27:07 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:27:14 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:27:18 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:27:22 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:27:25 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:27:29 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:27:33 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:27:37 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:27:40 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:27:44 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:27:48 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:27:51 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:27:55 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:27:58 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:28:02 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:29:06 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:29:10 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:32:40 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:32:49 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:33:49 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:33:58 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:36:01 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:36:13 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:36:33 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:36:38 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:36:51 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:37:27 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:37:32 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:37:36 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:37:46 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:41:38 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:42:11 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 25/08/21 11:42:23 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 25/08/21 11:51:56 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 11:52:12 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 25/08/21 11:52:42 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 25/08/21 11:52:47 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 25/08/21 15:22:51 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 15:23:01 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 15:23:05 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 15:23:09 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 16:31:30 2.0.5003 24 V no está presente en la máquina. Verifique el botón de parada de emergencia no esté presionado CSM N° ---- 25/08/21 16:35:44 2.0.5003 24 V no está presente en la máquina. Verifique el botón de parada de emergencia no esté presionado Maintenance N° ----25/08/21 16:45:17 2.11.5002 Integrity error on MEP Error 24V Maintenance N° ----25/08/21 16:46:16 2.11.5002 Integrity error on MEP Error 24V Maintenance N° ----25/08/21 16:46:28 2.11.5002 Integrity error on MEP Error 24V CSM N° ---- 25/08/21 16:54:09 2.0.5003 24 V no está presente en la máquina. Verifique el botón de parada de emergencia no esté presionado CSM N° ---- 25/08/21 17:12:57 2.0.5003 24 V no está presente en la máquina. Verifique el botón de parada de emergencia no esté presionado CSM N° ---- 25/08/21 17:21:57 2.0.5003 24 V no está presente en la máquina. Verifique el botón de parada de emergencia no esté presionado CSM N° ---- 25/08/21 18:32:03 2.0.1009 probe number , Imposible abrir puerto COM del amplificador CSM N° ---- 25/08/21 18:40:04 2.0.1009 probe number , Imposible abrir puerto COM del amplificador CSM N° ---- 25/08/21 18:58:50 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 18:58:56 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 18:59:07 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 18:59:11 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 18:59:15 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 20:23:20 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 25/08/21 20:24:33 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 25/08/21 20:24:34 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 25/08/21 20:29:03 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 25/08/21 20:29:26 2.2.1601 No se puede comunicar con el escáner de ultrasonidos CSM N° ---- 25/08/21 20:29:27 2.2.1601 No se puede comunicar con el escáner de ultrasonidos EDAP N° 0153 25/08/21 20:39:41 2.5.1202 El paciente se ha movido EDAP N° 0153 25/08/21 20:49:44 1.9.5104 Tratamiento cancelado en la fase de localización transversal CSM N° ---- 25/08/21 20:56:55 1.3.5101 Tratamiento cancelado en la fase de edición de los datos del tratamiento CSM N° ---- 25/08/21 20:57:38 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 20:57:43 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 20:57:47 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 20:57:52 2.2.1502 No se puede comunicar con la sonda CSM N° ---- 25/08/21 20:58:09 2.2.1502 No se puede comunicar con la sonda Maintenance N° ----01/09/21 13:11:13 3.11.1805 File integrity corrupt on parameters set file. EDAP N° 0154 01/09/21 13:31:28 1.9.5104 Tratamiento cancelado en la fase de localización transversal EDAP N° 0154 01/09/21 13:35:59 1.9.5104 Tratamiento cancelado en la fase de localización transversal EDAP N° 0154 01/09/21 13:38:48 1.9.5104 Tratamiento cancelado en la fase de localización transversal EDAP N° 0154 01/09/21 13:41:05 1.9.5104 Tratamiento cancelado en la fase de localización transversal EDAP N° 0154 01/09/21 13:43:13 1.9.5104 Tratamiento cancelado en la fase de localización transversal EDAP N° 0154 01/09/21 13:44:54 1.9.5104 Tratamiento cancelado en la fase de localización transversal EDAP N° ---- 01/09/21 14:50:24 1.3.5101 Tratamiento cancelado en la fase de edición de los datos del tratamiento CSM N° ---- 07/09/21 17:35:11 1.3.5101 Tratamiento cancelado en la fase de edición de los datos del tratamiento CSM N° ---- 07/09/21 18:00:19 1.3.5101 Tratamiento cancelado en la fase de edición de los datos del tratamiento CSM N° 0184 09/09/21 10:15:52 2.5.1202 El paciente se ha movido CSM N° 0184 09/09/21 10:25:15 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.6 mm) tiene que estar dentro del intervalo de 25.0 mm a 27.0 mm CSM N° 0184 09/09/21 10:25:55 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (29.5 mm) tiene que estar dentro del intervalo de 25.0 mm a 27.0 mm CSM N° 0184 09/09/21 10:28:00 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.8 mm) tiene que estar dentro del intervalo de 20.4 mm a 22.4 mm CSM N° 0184 09/09/21 10:28:18 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.7 mm) tiene que estar dentro del intervalo de 20.4 mm a 22.4 mm CSM N° 0184 09/09/21 10:28:27 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.0 mm) tiene que estar dentro del intervalo de 20.4 mm a 22.4 mm CSM N° 0184 09/09/21 11:02:18 1.9.5105 Tratamiento cancelado durante la fase de disparo CSM N° 0185 09/09/21 14:53:24 2.5.1202 El paciente se ha movido CSM N° 0185 09/09/21 14:53:37 2.5.1102 Temperatura de la sonda excesiva : 20.0 CSM N° 0185 09/09/21 14:53:39 2.5.1102 Temperatura de la sonda excesiva : 19.9 CSM N° 0185 09/09/21 15:05:18 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (24.2 mm) tiene que estar dentro del intervalo de 25.0 mm a 27.0 mm CSM N° 0185 09/09/21 15:21:31 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.6 mm) tiene que estar dentro del intervalo de 24.1 mm a 26.1 mm CSM N° 0185 09/09/21 15:21:53 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (26.8 mm) tiene que estar dentro del intervalo de 24.1 mm a 26.1 mm CSM N° 0185 09/09/21 15:22:10 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.8 mm) tiene que estar dentro del intervalo de 24.1 mm a 26.1 mm CSM N° 0185 09/09/21 15:24:11 1.9.5105 Tratamiento cancelado durante la fase de disparo CSM N° 0186 14/09/21 09:54:02 2.5.1202 El paciente se ha movido CSM N° 0186 14/09/21 10:02:23 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.8 mm) tiene que estar dentro del intervalo de 20.6 mm a 22.6 mm CSM N° 0186 14/09/21 10:03:44 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (24.6 mm) tiene que estar dentro del intervalo de 24.8 mm a 26.8 mm CSM N° 0186 14/09/21 10:04:20 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (27.7 mm) tiene que estar dentro del intervalo de 24.8 mm a 26.8 mm CSM N° 0186 14/09/21 10:06:10 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (24.6 mm) tiene que estar dentro del intervalo de 25.0 mm a 27.0 mm CSM N° 0186 14/09/21 10:22:03 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0186 14/09/21 10:22:12 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0186 14/09/21 10:22:26 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0186 14/09/21 10:27:14 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (24.8 mm) tiene que estar dentro del intervalo de 25.0 mm a 27.0 mm CSM N° 0186 14/09/21 10:39:30 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0186 14/09/21 11:06:01 2.8.1421 El movimiento de la sonda es imposible. El motor ha alcanzado el paro mecánico Eje transversal CSM N° 0187 22/09/21 11:14:02 2.5.1202 El paciente se ha movido CSM N° 0187 22/09/21 11:19:51 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (20.7 mm) tiene que estar dentro del intervalo de 21.0 mm a 23.0 mm CSM N° 0187 22/09/21 11:20:14 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.2 mm) tiene que estar dentro del intervalo de 23.5 mm a 25.5 mm CSM N° 0187 22/09/21 11:21:10 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (24.8 mm) tiene que estar dentro del intervalo de 25.0 mm a 27.0 mm CSM N° 0187 22/09/21 11:21:40 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (25.2 mm) tiene que estar dentro del intervalo de 23.1 mm a 25.1 mm CSM N° 0187 22/09/21 11:28:46 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (24.1 mm) tiene que estar dentro del intervalo de 24.2 mm a 26.2 mm CSM N° 0187 22/09/21 11:47:13 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.2 mm) tiene que estar dentro del intervalo de 20.1 mm a 22.1 mm CSM N° 0187 22/09/21 11:48:51 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (23.5 mm) tiene que estar dentro del intervalo de 24.5 mm a 26.5 mm CSM N° 0187 22/09/21 11:53:12 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.9 mm) tiene que estar dentro del intervalo de 20.6 mm a 22.6 mm CSM N° 0187 22/09/21 11:54:08 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.6 mm) tiene que estar dentro del intervalo de 20.6 mm a 22.6 mm CSM N° 0187 22/09/21 11:55:21 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.5 mm) tiene que estar dentro del intervalo de 22.6 mm a 24.6 mm CSM N° 0187 22/09/21 11:56:38 1.8.1303 Error de posición de la pared rectal La posición de la pared rectal medida (22.7 mm) tiene que estar dentro del intervalo de 20.5 mm a 22.5 mm Maintenance N° ----24/09/21 09:12:28 3.11.1805 File integrity corrupt on parameters set file. EDAP N° 0160 24/09/21 09:39:11 1.9.5105 Tratamiento cancelado durante la fase de disparo EDAP N° 0160 24/09/21 09:43:07 2.8.1202 El paciente se ha movido EDAP N° 0160 24/09/21 09:43:07 2.8.1202 El paciente se ha movido EDAP N° 0160 24/09/21 09:43:56 2.8.1202 El paciente se ha movido